Analysis on Longitudinal Dose according to Change of Field Width

선속 폭(Field Width) 변화에 따른 종축선량 분석

  • Jung, Won-Seok (Department of Radiation Oncology, Kyunghee University Hospital) ;
  • Back, Jong-Geal (Yonsei Cancer Center, Yonsei University Health System) ;
  • Shin, Ryung-Mi (Department of Radiation Oncology, Kyunghee University Hospital) ;
  • Oh, Byung-Cheon (Department of Radiation Oncology, Kyunghee University Hospital) ;
  • Jo, Jun-Young (Department of Radiation Oncology, Kyunghee University Hospital) ;
  • Kim, Gi-Chul (Department of Radiation Oncology, Kyunghee University Hospital) ;
  • Choi, Tae-Gu (Department of Radiation Oncology, Kyunghee University Hospital)
  • 정원석 (경희대학교병원 방사선종양학과) ;
  • 백종걸 (연세의료원 방사선종양학과) ;
  • 신령미 (경희대학교병원 방사선종양학과) ;
  • 오병천 (경희대학교병원 방사선종양학과) ;
  • 조준영 (경희대학교병원 방사선종양학과) ;
  • 김기철 (경희대학교병원 방사선종양학과) ;
  • 최태규 (경희대학교병원 방사선종양학과)
  • Received : 2011.06.11
  • Accepted : 2011.08.02
  • Published : 2011.09.30

Abstract

Purpose: To analyze the accuracy of tumor volume dose following field width change, to check the difference of dose change by using self-made moving car, and to evaluate practical delivery tumor dose when tomotherapy in the treatment of organ influenced by breathing. Materials and Methods: By using self-made moving car, the difference of longitudinal movement (0.0 cm, 1.0 cm, 1.5 cm, 2.0 cm) was applied and compared calculated dose with measured dose according to change of field width (1.05 cm, 2.50 cm, 5.02 cm) and apprehended margin of error. Then done comparative analysis in degree of photosensitivity of DQA film measured by using Gafchromic EBT film. Dose profile and Gamma histogram were used to measure degree of photosensitivity of DQA film. Results: When field width were 1.05 cm, 2.50 cm, 5.02 cm, margin of error of dose delivery coefficient was -2.00%, -0.39%, -2.55%. In dose profile of Gafchromic EBT film's analysis, the movement of moving car had greater motion toward longitudinal direction and as field width was narrower, big error increased considerably at high dose part compared to calculated dose. The more field width was narrowed, gamma index had a large considerable influence of moving at gamma histogram. Conclusion: We could check the difference of longitudinal dose of moving organ. In order to small field width and minimize organ moving due to breathing, it is thought to be needed to develop breathing control unit and fixation tool.

목 적: 호흡에 의하여 영향을 받는 장기의 토모 치료 시 선속 폭(Field Width) 변화에 따른 종양 용적선량의 선량 전달 정확성을 분석하고 자체 제작한 Moving car를 사용하여 선량 변화의 차이를 확인하고 실질적인 종양 선량의 전달 선량을 평가해 보고자 한다. 대상 및 방법: 자체 제작한 Moving car를 사용하여 종축 움직임 차이(0.0 cm, 1.0 cm, 1.5 cm, 2.0 cm)를 적용하고 선속 폭(Field Width) 변화(1.05 cm, 2.50 cm, 5.02 cm)에 따른 측정된 선량의 값을 계산된 선량 값과 비교하여 오차범위를 파악하였다. Gafchromic EBT 필름을 이용하여 측정한 DQA (Delivery Quality Assurance) 필름의 감광정도를 선량윤곽(Dose Profile)과 Gamma Histogram을 이용하여 비교 분석하였다. 결 과: 선속 폭(Field Width) 1.05 cm, 2.50 cm, 5.02 cm일 때 오차범위가 각각 -2.00%, -0.39%, -2.55%의 선량 전달율을 나타내었다. Gafchromic EBT 필름 분석의 선량 윤곽(Dose profile)에서는 Moving car의 움직임이 있는 상태에서 종축방향으로의 움직임이 클수록, 선속 폭(Field Width)이 좁아질수록 고선량 부분에서 계산된 선량과의 오차가 크게 나타났다. Gamma histogram에서는 gamma index가 선속 폭(Field Width)이 좁을수록 움직임의 영향이 상당히 증가하는 것을 알 수 있었다. 결 론: 자체 제작한 Moving car를 이용하여 움직이는 장기의 종축선량 차이를 확인할 수 있었으며 토모 치료 시 선속 폭(Field Width)을 작게 하고 호흡에 의한 장기의 움직임을 최소화하기 위하여 호흡조절 장치와 고정기구의 개발이 필요하리라 사료된다.