In this study, we tried to analyze the influence of ICM(Iodinated Contrast Media) in MR imaging compare to GBCA(Gadolinium Based Contrast Agent), and as this result we discussed whether resonable or not the protocol which is MRI scan after enhanced CT scan without proper time interval in clinical field. For this research, we assembled two phantoms. which one was iodine and another one was gadolinium. We did test two phantoms in conventional MRI scan which is T1, T2, T2 FLAIR and 3D angio. After that, quantitative analysis was progressed. The results of study were as follow : SSI(Saline's Signal Intensity) was shown as each sequences 175, 1231, 333, 37 [a.u] at iodine. and 1297, 123, 757, 232 [a.u] was recorded at gadolinium. BDEPS(the Biggest Difference of EPS) was shown as each sequences 1297, 123, 757, 232 [a.u] at iodine and 793, 6, 1495, 365 [a.u] was recorded at gadolinium. At this time, EPS(Enhancement Percentage to Saline) was shown 641.1, -90.0, 127.3, 527% at iodine and 685.1, 99.4, 365.7, 1077.4% was recorded at gadolinium. BP(BDEPS's point) was shown 900, 900, 477, 900 mmol at iodine and 4, 0.2, 0.2, 40 mmol was recorded at gadolinium. CPSS(Change Point of SI to SSI) was shown 63, 423, 63, 29 mmol at iodine and each [50, 30], [4, 0.2], [4, 1], 0.2 mmol was recorded at gadolinium. According to this research, we could not only discover the fact that was iodine could effect on MR signal, but also the pattern is different as various sequences compare to gadolinium. Therefore, we expect useful diagnostic MR image in clinical field with this quantitative data for deciding protocol regarding MRI and CT scan order.