A study of Curved Dosimeter for Flattening Filter Free Beam Quality Assurance Evaluation using Curved Dosimeter in Radiotherapy (Flattening Filter Free Beam의 정도관리를 위한 곡면선량계 가능성 연구)

  • Han, Moojae;Shin, Yohan;Jung, Jaehoon;Heo, Seunguk;Kim, Kyotae;Heo, Yeji;Cho, Heunglae;Park, Sungkwang
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.1
    • /
    • pp.119-124
    • /
    • 2019
  • Radiation therapy using flattening filter free beam can prevent beam attenuation caused by flattening filter and can improve treatment efficiency. However, accurate dose control is not established for nonuniform iso dose distributions. In this study, curved dosimeter based on photoconductive material $HgI_2$ was fabricated and its reproducibility and linearity were evaluated at 6 MV photon energy to verify its performance. In order to show the usefulness of the curved measurement, the signals measured on the flat substrate and the curved substrate were compared in the flattening filter free beam using the acrylic filter. As a result, the reproducibility of the unit cell dosimeter was evaluated as SE 0.613%, and the linearity was evaluated as R-sq 0.9999. The usability evaluation of the array curve dosimeter demonstrated its usefulness by indicating a curvature error rate of 11.073%p and a reduced error rate.

A Study on the Photoneutron Dose Estimation in Flattening Filter Mode and Flattening Filter Free Mode for Medical Linear Accelerator (의료용 선형가속기의 Flattening Filter Mode와 Flattening Filter Free Mode 간에 광중성자 선량 평가)

  • Yang, Oh Nam;Lim, Cheong Hwan
    • Journal of radiological science and technology
    • /
    • v.40 no.2
    • /
    • pp.297-302
    • /
    • 2017
  • In this study, the generation of photoneutrons between the 10 MV FF mode and the FFF mode was evaluated and the amount of photoneutrons generated by the 10 MV and 15 MV energy changes in the FFF mode was evaluated. The generated neutrons were evaluated at 13 measurement points and the KTEPC was used to collect the generated neutrons. 10 MV FF mode was measured at 10 MV FF mode and FFF mode at all measurement points. In the superior direction, 0.455mSv and 0.152mSv were the largest, and more than 33% optical neutron was generated in FF. 10 MV in FFF mode, 15 MV in 15 MV, and 0.402 mSv in the direction of Superior, and 6.9% in the direction.

The evaluation of properties for radiation therapy techniques with flattening filter-free beam and usefulness of time and economy to a patient with the radiation therapy (Flattening filter-free beam을 이용한 방사선 치료 기법의 특성 및 환자의 시간적.경제적 유용성 평가)

  • Goo, Jang Hyeon;Won, Hui Su;Hong, Joo Wan;Chang, Nam Jun;Park, Jin Hong
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.26 no.2
    • /
    • pp.363-368
    • /
    • 2014
  • Purpose : The aim of this study was to appraise properties for radiation therapy techniques and effectiveness of time and economy to a patient in the case of applying flattening filter-free (3F) and flattening filter (2F) beam to the radiation therapy. Materials and Methods : Alderson rando phantom was scanned for computed tomography image. Treatment plans for intensity modulated radiation therapy (IMRT), volumetric modulated arc therapy (VMAT) and stereotactic body radiation therapy (SBRT) with 3F and 2F beam were designed for prostate cancer. To evaluate the differences between the 3F and 2F beam, total monitor units (MUs), beam on time (BOT) and gantry rotation time (GRT) were used and measured with $TrueBeam^{TM}$ STx and Surveillance And Measurement (SAM) 940 detector was used for photoneutron emitted by using 3F and 2F. To assess temporal and economical aspect for a patient, total treatment periods and medical fees were estimated. Results : In using 3F beam, total MUs in IMRT plan increased the highest up to 34.0% and in the test of BOT, GRT and photoneutron, the values in SBRT plan decreased the lowest 39.8, 38.6 and 48.1%, respectively. In the temporal and economical aspect, there were no differences between 3F and 2F beam in all of plans and the results showed that 10 days and 169,560 won was lowest in SBRT plan. Conclusion : According as the results, total MUs increased by using 3F beam than 2F beam but BOT, GRT and photoneutron decreased. From above the results, using 3F beam can decrease intra-fraction setup error and risk of radiation-induced secondary malignancy. But, using 3F beam did not make the benefits of temporal and economical aspect for a patient with the radiation therapy.

A Study on IMRT (Intensity Modulated Radiation Therapy) Delivery Technique and FFF (Flattening Filter Free) Beam to Increase Skin Dose to Irregularly Shaped Skin Surface. (IMRT(Intensity Modulated Radiation Therapy)전달 기법과 FFF(Flattening Filter Free) 빔을 이용한 요철 부위 피부 선량 증가 방법에 대한 고찰)

  • Woo Heon;Son Sang Jun;Je Young Wan
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.34
    • /
    • pp.7-12
    • /
    • 2022
  • Purpose: When it is difficult to secure the skin dose when treating Irregularly Shaped Skin Surface such as the nose where it is difficult to apply a bolus, increase the skin dose with a treatment plan that combines the IMRT (Intensity Modulated Radiation Therapy) delivery technique and FFF (Flattening Filter Free), It was tried to find out whether or not through the phantom experiment. Materials & Methods: Based on the 6MV-FF (Flattening Filter) and VMAT (Volumetric-Modulated Arc Therapy) treatment plans, which are the most commonly used treatment plans for head and neck cancer, A comparison group was created by combining VMAT and IMRT, FF and FFF, and the presence or absence of 5 mm bolus application. A virtual target was created on the Rando Phantom's nose, and a virtual bolus of 5 mm was applied assuming full contact on the Rando Phantom's nose. Five measurement points were determined based on the phantom's nose, and the absorbed dose was measured by irradiating each treatment plan 3 times per treatment plan according to the treatment technique and whether or not the bolus was applied. Result: The difference in skin dose in FF vs FFF increased in the case of FFF in VMAT bolus off, and there was no difference in case of IMRT bolus off. In VMAT bolus 5 mm and IMRT bolus 5 mm, it was confirmed that the skin dose was rather decreased in FFF. The difference in skin dose between VMAT and IMRT increased only in the case of FFF bolus off, and there was no statistical difference in the rest. For the difference in skin dose between bolus off vs bolus 5 mm, it was confirmed that the skin dose increased at bolus 5 mm, except for the case of using IMRT FFF. The treatment plan combining IMRT and FFF did not find any statistically significant difference as a result of analyzing the measured values of the treatment plan skin dose applied with a 5 mm bolus using the commonly used VMAT and FF. Therefore, it is thought that by using IMRT_FFF, it is possible to deliver a skin dose similar to that of applying a 5 mm bolus to VMAT_FF, which can be useful for patients who need a high skin dose but have difficulty applying a bolus. Conclusion: For patients who find it difficult to apply bolus, an increase in skin dose can be expected with a treatment plan that properly combines IMRT and FFF compared to VMAT and FF.

Dosimetric Comparison of 6 MV Flattening Filter Free and 6 MV Stereotactic Radiosurgery Beam Using 4 mm Conical Collimator for Trigeminal Neuralgia Radiosurgery

  • Mhatre, Vaibhav R;Chadha, Pranav;Kumar, Abhaya P;Talapatra, Kaustav
    • Journal of Radiation Protection and Research
    • /
    • v.43 no.3
    • /
    • pp.107-113
    • /
    • 2018
  • Background: The purpose of our study was to compare the dosimetric advantages of Flattening filter free (FFF) beams for trigeminal neuralgia patients using 4 mm conical collimators over previously treated patients with 6 MV SRS beam. Materials and Methods: A retrospective study was conducted for 5 TN patients who had been previously treated at our institution using frame-based, LINAC-based stereotactic radiosurgery (SRS) on Novalis Tx using 6 MV SRS beam were replanned on 6X FFF beams on Edge Linear accelerator with same beam angles and dose constraints using 4 mm conical collimator. The total number of monitor units along with the beam on time was compared for both Edge and Novalis Tx by redelivering the plans in QA mode of LINAC to compare the delivery efficiency. Plan quality was evaluated by homogeneity index (HI) and Paddick gradient index (GI) for each plan. We also analyzed the doses to brainstem and organ at risks (OARs). Results and Discussion: A 28% beam-on time reduction was achieved using 6X FFF when compared with 6X SRS beam of Novalis Tx. A sharp dose fall off with gradient index value of $3.4{\pm}0.27$ for 4 mm Varian conical collimator while $4.17{\pm}0.20$ with BrainLab cone. Among the 5 patients treated with a 4 mm cone, average maximum brainstem dose was 10.24 Gy for Edge using 6X FFF and 14.28 Gy for Novalis Tx using 6X SRS beam. Conclusion: The use of FFF beams improves delivery efficiency and conical collimator reduces dose to OAR's for TN radiosurgery. Further investigation is warranted with larger sample patient data.

STUDY ON MONITORING UNIT EFFICIENCY OF FLATTENING-FILTER FREE PHOTON BEAM IN ASSOCIATION WITH TUMOR SIZE AND LOCATION

  • Kim, Dae Il;Kim, Jung-In;Yoo, Sook Hyun;Park, Jong Min
    • Journal of Radiation Protection and Research
    • /
    • v.38 no.4
    • /
    • pp.194-201
    • /
    • 2013
  • To investigate monitoring unit (MU) efficiency and plan quality of volumetric modulated arc therapy (VMAT) using flattening-filter free (FFF) photon beam in association with target size and location. A virtual patient was generated in Eclipse$^{TM}$ (ver. A10, Varian Medical Systems, Palo Alto, USA) treatment planning system. The length of major and minor axis in axial view was 50 cm and 30 cm, respectively. Cylindrical-shaped targets were generated inside that patient at the center (symmetric target) and in the periphery (asymmetric target, 7.5 cm away from the center of the patient to the right direction) of the virtual patient. The longitudinal length was 10 cm and the diameters were 2, 5, 10 and 15 cm. Total 8 targets were generated. RapidArc$^{TM}$ plans using TrueBeam STx$^{TM}$ were generated for each target. Two full arcs were used and the axis of rotation of the gantry was set to be at the center of the virtual patient. Total MU, homogeneity index (HI), target mean dose, the value of gradient measure and body mean dose were calculated. In the case of symmetric targets, averaged total MU of FFF plan was 23% and 19% higher than that of flattening filter (FF) plan when using 6 MV and 10 MV photons, respectively. The difference of HI, target mean dose, gradient measure and body mean dose between FF and FFF was less than 0.04, 2.6%, 0.1 cm and 2.2%, respectively. For the asymmetric targets, total MU of FFF plan was 21% and 32% was higher than that of FF when using 6 MV and 10 MV photons, respectively. The homogeneity of the target was always worse when using FFF than using FF. The maximum difference of HI was 0.22. The target mean dose of FFF was 3.2% and 4.1% higher than that of FF for the 6 MV and 10 MV, respectively. The difference of gradient measure was less than 0.1 cm. The body mean dose was higher when using FFF than FF about 4.2% and 2.8% for the 6 MV and 10 MV, respectively. No significant differences between VMAT plans of FFF beam and FF beam were observed in terms of quality of treatment plan. The HI was higher when using FFF 10 MV photons for the asymmetric targets. The MU was increased noticeably when using FFF photon beams.

Evaluation of Photoneutron During Radiation Therapy when Using Flattening Filter and Tracking Jaw with High Energy X-ray (고 에너지 X선 방사선치료 시 Flattening Filter와 Tracking Jaw 사용에 따른 광중성자 발생 평가)

  • Park, Euntae;Jin, Seongjin;Park, Cheolwoo
    • Journal of the Korean Society of Radiology
    • /
    • v.10 no.2
    • /
    • pp.125-131
    • /
    • 2016
  • Radiation therapy is usually using linear accelerator and used X-ray energy is also getting higher. Recently linear accelerators has been developed 3F mode and tracking jaw technology and that was applied for patient therapy. This study aims at measuring photoneutrons depending on the use of 3F and tracking jaw system when radiation is irradiated using a linear accelerator. The generation of photoneutrons of 3F system was 70% smaller than 2F system and that of tracking jaw system was 83% higher than static jaw system. Photoneutron value is relatively low. However, it must be minimized for Photoneutron exposure during radiation therapy.

An Empirical Approach to Dosimetric Effect of Carbon Fiber Couch for Flattening Filter Free Beam of Elekta LINAC

  • Ahn, Sohyun;Park, Kwangwoo;Kim, Jinsung;Lee, Ho;Yoon, Jeongmin;Lee, Eungman;Park, Sohyun;Park, Jeongeun;Kim, Juhye;Keum, Ki Chang
    • Progress in Medical Physics
    • /
    • v.27 no.4
    • /
    • pp.220-223
    • /
    • 2016
  • Generally, it is recommended that the dosimetric effect of carbon fiber couch should be considered especially for an intensity-modulated therapy with a large portion of monitor units from posterior angles. Even a flattening filter free (FFF) beam has been used for stereotactic body radiation therapy (SBRT), the effect of carbon fiber couch for FFF beam is not well known. This work is an effort to evaluate the dosimetric effect of carbon fiber couch for flattened and FFF beam of Elekta linac empirically. The absorbed doses were measured with Farmer type chamber and water-equivalent phantoms with and without couch. And differences of the absorbed doses between with and without couch defined as "couch effect". By comparing calculated dose in treatment planning system (TPS) with measured dose, the optimal density of couch was evaluated. Finally, differences on patient's skin dose and target dose by couch were evaluated in TPS. As a result, the couch effect for 6 and 10 MV flattened beam were -2.71% and -2.32%, respectively. These values were agreed with provided data by vendor within 0.5%. The couch effect for 6 and 10 MV FFF beam were -3.75% and -2.80%, respectively. The patient's skin dose was increased as 18.6% and target dose was decreased as 0.87%, respectively. It was realized that the couch effect of FFF beam was more severe than that of flattened beam. Patient's skin dose and target dose were changed by the couch effect.

Use of Flattening Filter Free Photon Beams for Off-axis Targets in Conformal Arc Stereotactic Body Radiation Therapy

  • Smith, Ashley;Kim, Siyong;Serago, Christopher;Hintenlang, Kathleen;Ko, Stephen;Vallow, Laura;Peterson, Jennifer;Hintenlang, David;Heckman, Michael;Buskirk, Steven
    • Progress in Medical Physics
    • /
    • v.25 no.4
    • /
    • pp.288-297
    • /
    • 2014
  • Dynamic conformal arc therapy (DCAT) and flattening-filter-free (FFF) beams are commonly adopted for efficient conformal dose delivery in stereotactic body radiation therapy (SBRT). Off-axis geometry (OAG) may be necessary to obtain full gantry rotation without collision, which has been shown to be beneficial for peripheral targets using flattened beams. In this study dose distributions in OAG using FFF were evaluated and the effect of mechanical rotation induced uncertainty was investigated. For the lateral target, OAG evaluation, sphere targets (2, 4, and 6 cm diameter) were placed at three locations (central axis, 3 cm off-axis, and 6 cm off-axis) in a representative patient CT set. For each target, DCAT plans under the same objective were obtained for 6X, 6FFF, 10X, and 10FFF. The parameters used to evaluate the quality of the plans were homogeneity index (HI), conformality indices (CI), and beam on time (BOT). Next, the mechanical rotation induced uncertainty was evaluated using five SBRT patient plans that were randomly selected from a group of patients with laterally located tumors. For each of the five cases, a plan was generated using OAG and CAG with the same prescription and coverage. Each was replanned to account for one degree collimator/couch rotation errors during delivery. Prescription isodose coverage, CI, and lung dose were evaluated. HI and CI values for the lateral target, OAG evaluation were similar for flattened and unflattened beams; however, 6FFF provided slightly better values than 10FFF in OAG. For all plans the HI and CI were acceptable with the maximum difference between flattened and unflattend beams being 0.1. FFF beams showed better conformality than flattened beams for low doses and small targets. Variation due to rotational error for isodose coverage, CI, and lung dose was generally smaller for CAG compared to OAG, with some of these comparisons reaching statistical significance. However, the variations in dose distributions for either treatment technique were small and may not be clinically significant. FFF beams showed acceptable dose distributions in OAG. Although 10FFF provides more dramatic BOT reduction, it generally provides less favorable dosimetric indices compared to 6FFF in OAG. Mechanical uncertainty in collimator and couch rotation had an increased effect for OAG compared to CAG; however, the variations in dose distributions for either treatment technique were minimal.

A Study of volumetric modulated arc therapy for stereotactic body radiation therapy in case of multi-target liver cancer using flattening filter free beam (다중표적 간암의 정위적체부방사선치료에서 무편평화여과기선질 용적변조회전치료의 유용성 평가)

  • Yeom, Misuk;Yoon, Inha;Hong, Donggi;Back, Geummun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.27 no.1
    • /
    • pp.31-43
    • /
    • 2015
  • Purpose : Stereotactic body radiation therapy (SBRT) has proved its efficacy in several patient populations with primary and metastatic limited tumors. Because SBRT prescription is high dose level than Conventional radiation therapy. SBRT plan is necessary for effective Organ at risk (OAR) protection and sufficient Planning target volume (PTV) dose coverage. In particular, multi-target cases may result excessive doses to OAR and hot spot due to dose overlap. This study evaluate usefulness of Volumetric modulated arc therapy (VMAT) in dosimetric and technical considerations using Flattening filter free (FFF) beam. Materials and Methods : The treatment plans for five patients, being treated on TrueBeam STx(Varian$^{TM}$, USA) with VMAT using 10MV FFF beam and Standard conformal radiotherapy (CRT) using 15MV Flattening filter (FF) beam. PTV, liver, duodenum, bowel, spinal cord, esophagus, stomach dose were evaluated using the dose volume histogram(DVH). Conformity index(CI), homogeneity index(HI), Paddick's index(PCI) for the PTV was assessed. Total Monitor unit (MU) and beam on time was assessed. Results : Average value of CI, HI and PCI for PTV was $1.381{\pm}0.028$, $1.096{\pm}0.016$, $0.944{\pm}0.473$ in VMAT and $1.381{\pm}0.042$, $1.136{\pm}0.042$, $1.534{\pm}0.465$ in CRT respectively. OAR dose in CRT plans evaluated 1.8 times higher than VMAT. Total MU in VMAT evaluated 1.3 times increase than CRT. Average beam on time was 6.8 minute in VMAT and 21.3 minute in CRT. Conclusion : VMAT for SBRT in multi-target liver cancer using FFF beam is effective treatment techniqe in dosimetric and technical considerations. VMAT decrease intra-fraction error due to treatment time shortening using high dose rate of FFF beam.

  • PDF