• 제목/요약/키워드: Dose Area Product (DAP)

검색결과 47건 처리시간 0.032초

Review of National Diagnostic Reference Levels for Interventional Procedures

  • Lee, Min Young;Kwon, Jae;Ryu, Gang Woo;Kim, Ki Hoon;Nam, Hyung Woo;Kim, Kwang Pyo
    • 한국의학물리학회지:의학물리
    • /
    • 제30권4호
    • /
    • pp.75-88
    • /
    • 2019
  • Diagnostic reference level (DRL) is employed to optimize the radiation doses of patients. The objective of this study is to review the DRLs for interventional procedures in Korea and abroad. Literature review was performed to investigate radiation dose index and measurement methodology commonly used in DRL determination. Dose area product (DAP) and fluoroscopy time within each major procedure category were systematically abstracted and analyzed. A wide variation was found in the radiation dose. The DAP values and fluoroscopy times ranged 0.01-3,081 Gy·㎠ and 2-16,878 seconds for all the interventional procedures, 8.5-1,679 Gy·㎠ and 32-5,775 seconds for the transcatheter arterial chemoembolization (TACE), and 0.1-686 Gy·㎠ and 16-6,636 seconds for the transfemoral cerebral angiography (TFCA), respectively. The DRL values of the DAP and fluoroscopy time were 238 Gy·㎠ and 1,224 seconds for the TACE and 189 Gy·㎠ and 686 seconds for the TFCA, respectively. Generally, the DRLs of Korea were lower than those of other developed countries, except for the percutaneous transluminal angioplasty with stent in arteries of the lower extremity (LE PTA and stent), aneurysm coil embolization, and Hickman insertion procedures. The wide variation in the radiation doses of the different procedures suggests that more attention must be paid to reduce unnecessary radiation exposure from medical imaging. Furthermore, periodic nationwide survey of medical radiation exposures is necessary to optimize the patient dose for radiation protection, which will ultimately contribute to patient dose reduction and radiological safety.

인터벤션 시술 시 면적선량계를 이용한 환자 방사선 선량 평가 (Evaluation of Patient Radiation Doses Using DAP Meter in Interventional Radiology Procedures)

  • 강병삼;윤용수
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제40권1호
    • /
    • pp.27-34
    • /
    • 2017
  • 해외 각국에서 발표된 중재적 방사선 영역에서의 환자선량을 조사하였으며, 전국 23개 주요 병원에서 국내에서 많이 시행되는 13개의 주요 중재적 방사선 시술에 대한 환자선량을 DAP meter를 이용하여 측정하였으며, 시술별 피폭선량에 관한 8,415 건의 국내 자료를 확보하고, 각 주요 시술별로 참고 선량치를 제시하였다. 연구결과는 경동맥화학색전술 $237.7Gy{\cdot}cm^2$, 투석용 동정맥루 인터벤션시술 $17.3Gy{\cdot}cm^2$, 하지 혈관질환의 인터벤션시술 $114.1Gy{\cdot}cm^2$, 뇌혈관조영술 $188.5Gy{\cdot}cm^2$, 뇌동맥류 코일색전술 $383.5Gy{\cdot}cm^2$, 경피경간 담즙 배액술 $64.6Gy{\cdot}cm^2$, 담도 스텐트설치술 $64.6Gy{\cdot}cm^2$, 요로 폐색에 대한 신루설치술 $22.4Gy{\cdot}cm^2$, 다목적 중심정맥 카테터 삽입시술 $4.3Gy{\cdot}cm^2$, 항암제 주입 목적의 매몰형 중심정맥 카테터 삽입시술 $2.8Gy{\cdot}cm^2$, 혈액 투석 목적의 중심정맥 카테터 삽입시술 $4.4Gy{\cdot}cm^2$, 카테터를 이용한 배액시술 $17.1Gy{\cdot}cm^2$ 그리고 장기혈관색전술 $357.9Gy{\cdot}cm^2$이다. 본 연구를 통해 얻은 선량참고치는 방사선 인터벤션 시술에서 환자선량을 줄일 수 있는 최소한의 가이드라인으로 활용될 수 있을 것으로 생각되며, 참고선량치에 대한 연구는 장비의 발달과 시술방법, 재료의 발전으로 변화될 수 있음을 감안하고 선진국에서는 5년마다 시행되고 있음을 참고할 때, 향후 면적선량을 유효선량으로 변환할 수 있는 국내시술에 적합한 한국형 변환계수가 연구되어야 할 것으로 사료된다.

디지털 심혈관조영장치의 기하학적 특성에 따른 선량 감소 (Reduction of Radiation Dose according to Geometric Parameters from Digital Coronary Angiography)

  • 강영한;조평곤
    • 한국방사선학회논문지
    • /
    • 제7권4호
    • /
    • pp.277-284
    • /
    • 2013
  • 심혈관 촬영 시 선량감소를 위해 실무자가 조절 가능한 기하학적 특성을 살펴보고, 각 특성에 따른 유효선량을 비교해 보았다. 인체 모형 팬텀을 이용하여 투시촬영과 영화촬영을 시행하였고, 선량의 측정은 심혈관 장치의 DAP meter를 이용하여 DAP 값을 유효선량으로 환산하였다. 먼저 영화촬영은 투시촬영에 비해 기하학적 특성 전반에 걸쳐서 선량이 약 6-7배 높았다. FPS mode에 따른 선량은 FPS를 낮게 설정할수록 선량이 70%까지 감소하였다. 선관 각도에 따른 선량은 LAO $45^{\circ}$+Caudal $30^{\circ}$이 가장 높게 측정되었고, 조사야 조절장치를 많이 적용할수록 선량은 감소하였다. X-선관과 영상증배관의 거리가 10 cm 멀어질수록 투시촬영과 영화촬영에서 각각 선량이 25-35%까지 증가하였다. FOV가 확대될수록 선량이 1.21-2배 증가하였고, 테이블과 영상증배관 거리가 10cm 멀어질수록 선량은 1.11-1.25배 까지 증가하였다. 따라서 본 연구에서 실험한 기하학적 특성인 FPS mode, 선관 각도, 조준기, 선관과 영상증배관 거리, 테이블과과 영상증배관 거리, FOV 영상 확대 등은 중재적 시술 시 실무자가 수시로 조절 가능한 인자이며, 각 특성에 따라 선량 감소 효과를 기대할 수 있다.

관전류 변화에 따른 면적선량과 영상 농도 평가 (Evaluation of Area Dose Product and Image Density according to the Variable Tube Current)

  • 윤영우;제재용
    • 한국방사선학회논문지
    • /
    • 제12권5호
    • /
    • pp.645-650
    • /
    • 2018
  • 본 연구는 면적선량에 따른 영상의 농도를 측정하여 피폭선량에 대한 정도관리 필요성을 제시하고자 하였다. 관전압을 80 kVp로 고정하고 관전류를 1, 25, 50, 80, 100 mAs로 조사한 결과 면적선량은 25 mAs에서 50 mAs로 증가하면 1.88배의 선량이 증가하고 50 mAs에서 100 mAs로 증가하면 2.05배 증가하였다. 하지만 필름으로 획득한 영상의 농도는 25 mAs에서 50 mAs로 증가하면 48% 증가하고, 50 mAs에서 100 mAs로 증가하면 29% 증가하였다. 또한 DR 영상의 농도는 25 mAs에서 50 mAs로 증가하면 12% 증가하고, 50 mAs에서 100 mAs로 증가하면 30% 증가하는 것으로 나타났다. 즉, 디지털 영상촬영 장비는 적정 촬영 조건에서 선량 증가에 따른 영상의 농도차이가 필름 영상보다는 적게 나타났다. 본 연구의 결과에서 디지털 영상 촬영 장비를 사용하는 의료기관에서는 방사선 선량에 대한 정도관리를 통하여 현재보다 촬영 부위별 피폭선량을 조금이나마 더 줄일 수 있을 것으로 판단되어진다.

조리개형 X선 콜리메이터의 개발에 따른 유용성 평가 (Usability Evaluation by Development of IRIS Type X-ray Collimator)

  • 강인석;박재윤;임청환;최재호;정홍량
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제41권3호
    • /
    • pp.249-254
    • /
    • 2018
  • In this study, we evaluated the DAP(Dose Area Product) reduction effect of the newly developed IRIS collimator by measuring the DAP of the Rectangle collimator and the IRIS collimator depending on the field, SID(Source to Image recpetor Distance) change, and AEC mode use. The results were as follows. The IRIS collimator decreased DAP by 34.91, 29.33, and 29.04%, respectively, compared to the Rectangle collimator when the field was increased to $8{\times}8$, $12{\times}12$, $16{\times}16inch$. And also, when the SID was increased to 100, 120 and 140 cm, the IRIS collimator decreased DAP by 10.73, 33.68 and 46.22%, respectively, compared to the Rectangle collimator. In AEC mode and none-AEC mode, DAP in IRIS collimator was reduced by 32.71 and 21.69%, respectively, compared with the Rectangle type. The IRIS collimator can reduce DAP by 29.62% on average compared to Rectangle type, which is statistically significant. These results suggest that the newly developed IRIS collimator can be used in medical field to alleviate radiation exposure.

관상동맥 조영술 및 경피적 관상동맥 중재술에 대한 진단참고준위에 관한 연구; 경상도지역중심 (A Study of Diagnostic Reference Levels for Coronary Angiography and Percutaneous Coronary Intervention in Gyeongsang Area)

  • 임시왕;김정수;조평곤
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제46권2호
    • /
    • pp.123-129
    • /
    • 2023
  • Interventional cardiology procedures can involve relatively high radiation doses compared to conventional radiography. During CAG, CAG + PCI and PCI the same area is exposed to radiation for a long period. In this study, radiation exposure data of 421 examinations in Gyeongsang area were collected, and the DRLs and ADs in actual medical practice for three types of interventional cardiology procedures in Korea were established. In CAG 286 case, 75th percentile DRLs and ADs of the total DAP were 55.89 Gy·cm2 and 37.47 Gy·cm2 , respectively. In CAG + PCI 92 case, those values were 222.84 Gy·cm2 and 117.51 Gy·cm2 respectively. In PCI 43 case, those values were 198.73 Gy·cm2 and 120.13 Gy·cm2 respectively. In this study, for the first time, the diagnostic reference level of interventional cardiology procedures in Gyeongsang area were established. Using the diagnostic reference level of interventional cardiology procedures derived from this study, it will help to identify and improve the level of exposure dose in the region and country.

PCXMC 소프트웨어를 이용한 소아에서의 CBCT 환자선량 평가 (Dose estimation of cone-beam computed tomography in children using personal computer-based Monte Carlo software)

  • 김은경
    • 대한치과의사협회지
    • /
    • 제58권7호
    • /
    • pp.388-397
    • /
    • 2020
  • Objective: The purpose of the study was to calculate the effective and absorbed organ doses of cone-beam computed tomography (CBCT) in pediatric patient using personal computer-based Monte Carlo (PCXMC) software and to compare them with those measured using thermoluminescent dosimeters (TLDs) and anthropomorphic phantom. Materials and Methods: Alphard VEGA CBCT scanner was used for this study. A large field of view (FOV) (20.0 cm × 17.9 cm) was selected because it is a commonly used FOV for orthodontic analyses in pediatric patients. Ionization chamber of dose-area product (DAP) meter was located at the tube side of CBCT scanner. With the clinical exposure settings for a 10-year-old patient, DAP value was measured at the scout and main projection of CBCT. Effective and absorbed organ doses of CBCT at scout and main projection were calculated using PCXMC and PCXMCRotation software respectively. Effective dose and absorbed organ doses were compared with those obtained by TLDs and a 10-year-old child anthropomorphic phantom at the same exposure settings. Results: The effective dose of CBCT calculated by PCXMC software was 292.6 μSv, and that measured using TLD and anthropomorphic phantom was 292.5 μSv. The absorbed doses at the organs largely contributing to effective dose showed the small differences between two methods within the range from -18% to 20%. Conclusion: PCXMC software might be used as an alternative to the TLD measurement method for the effective and absorbed organ dose estimation in CBCT of large FOV in pediatric patients.

  • PDF

인터벤션 시 방사선조사 조건에 따른 선량감소 : 면적선량과 영상화질 변화를 중심으로 (Dose Reduction According to the Exposure Condition in Intervention Procedure : Focus on the Change of Dose Area and Image Quality)

  • 황준호;정구민;김현수;강병삼;이경배
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제40권3호
    • /
    • pp.393-400
    • /
    • 2017
  • 본 연구는 NEMA Phantom을 사용하여 관전류량 변화에 따른 면적선량과 영상화질의 분석을 통해 선량을 감소할 수 있는 방법을 제시하고자 하였다. 인터벤션의 특성상 중요한 화질평가 항목인 공간 분해능, 저 대조도 분해능과 함께 신호 대 잡음비(Signal to Noise Ratio; SNR), 대조도 대 잡음비(Contrast to Noise Ratio; CNR)를 평가기준으로 하였고, 관전압은 80 kVp로 고정하고, 관전류량을 20, 30, 40, 50 mAs로 변화시켜 면적선량과 영상화질을 비교분석하였다. 그 결과 관전압을 고정한 상태에서 면적선량은 조건이 증가함에 따라 $1,066mGycm^2$에서 $6,160mGycm^2$으로 약 6배까지 증가하였고, 공간분해능과 저 대조도 분해능은 20 mAs, 30 mAs에서 평가기준보다 높게 관찰되었으나 40 mAs는 공간 분해능, 50 mAs에서는 공간 분해능과 저 대조도 분해능이 평가기준 이하로 관찰되었다. 또한 SNR과 CNR은 30 mAs까지는 증가하다 40 mAs에서 다소 증가하기는 하였으나 이전과 큰 차이가 없었고, 50 mAs에서는 감소하는 경향을 보였다. 결론적으로 관전류의 과잉노출로 피폭선량이 증가하고, 공간 분해능, 저 대조도 분해능, SNR, CNR 모든 영역에서 영상화질이 저하된다는 것을 알 수 있었고, 적절한 관전류량의 설정으로 인터벤션에서 피폭선량을 줄이고 영상화질을 충분히 개선할 수 있다는 것을 알 수 있었다.

심장 전기생리학 검사 시 조건 변화에 따른 환자 피폭 선량 평가 (Evaluation of Patient Exposure Dose during Cardiac Electrophysiology Study under Various Conditions)

  • 고성빈;안성민
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제46권6호
    • /
    • pp.501-508
    • /
    • 2023
  • This study used a adult absorption dose phantom (CIRS model 701-G, USA) made of human equivalent material and the vascular imaging equipment Allura Xper FD 20 (Philips, Netherlands). Optically stimulated luminescent dosimeters (OSLD) were inserted into the anatomical positions corresponding to each organ, and the exposure dose was measured. Dose area product (DAP) and air kerma (AK) measured by the dose meter in the equipment were compared. Continuous imaging was performed at two angles for a total of 20 minutes, with a frame per seconds of 3.75 and 7.5 fps and an FOV of 42 cm, 37 cm, and 31 cm, respectively, under the conditions of fluoflavor I, II, and III, each selected for 5 repetitions. This study was found that selecting a lower fps was the most effective way to reduce patient exposure dose, and adjusting the fluoflavor was a good alternative method for reducing patient exposure dose at high fps. Therefore the method of condition change with the greatest dose reduction effect is to set the minimum FPS and can reduce patient exposure dose according to geometric conditions and fluoflavor characteristics.

디지털 유방촬영장치에서 다엽 조리개 적용으로 인한 화질과 선량의 변화 (Variation of Image Quality and Dose by Applying Multi-Leaf Collimator for Digital Mammography)

  • 권순무;김부순;박형준;강영한
    • 한국방사선학회논문지
    • /
    • 제9권7호
    • /
    • pp.535-540
    • /
    • 2015
  • X선 조사야(X-ray beam field)는 X선관 앞에 부착된 선속제한기구(beam limiting device)인 조리개(collimator)에 의해 조절된다. 조리개는 1차 X선을 조절하여 환자의 피폭선량을 감소시키고 산란선량을 감소시켜 영상의 화질을 향상시키는 중요한 기능을 한다. 그러나 유방촬영장치의 조리개에 대해서 따로 규정되어 있지 않으며 상업화된 장치에서는 조사야가 사각형으로 조절되는 조리개를 사용하고 있다. 최근 디지털 유방촬영장치의 보급이 확대되고 있으나 film/screen방식에서부터 적용되던 사각형으로 조사야가 조절되는 조리개가 사용되고 있다. 따라서 본 연구에서는 디지털 유방촬영에서 조사야 조절 사용에 대한 실태조사 후 유방의 크기와 모양에 따라 조사야를 조절할 수 있는 다엽조리개를 제작 후 적용했을 때 영상의 화질변화와 평균유선선량, 면적선량 값의 변화를 알아보았다. 사각형으로 조사야가 조절되는 조리개와 비교하여 다엽 조리개를 적용한 영상에서 통계적으로 유의한 차이가 없었으며 DAP값 또한 50.72% 감소되었다. 따라서 디지털 유방촬영장치에서 조사야를 최적화하기 위한 자동조사야조절장치 개발이 환자 피폭선량 감소에 도움이 될 것으로 판단된다.