• Title/Summary/Keyword: Brachytherapy Source

Search Result 81, Processing Time 0.028 seconds

Radiation Dose Calculation using MIRD TYPE PHANTOM in the Surrounding Organs during Brachytherapy of Breast Cancer (유방암의 근접치료 시 수학적 모의피폭체를 이용한 인접장기의 선량평가)

  • Park, Eun-tae;Kim, Jung-hoon;Im, In-chul
    • Journal of the Korean Society of Radiology
    • /
    • v.10 no.4
    • /
    • pp.271-278
    • /
    • 2016
  • This study was fulfilled to evaluate the absorbed dose of breast and adjacent organs using MIRD type phantom in brachytherapy of breast cancer. The absorbed dose was analyzed assuming left or right breast is source organ which is $^{103}Pd$ or $^{192}Ir$. As a result, $^{192}Ir$ dose is higher than $^{103}Pd$ in source organ and also in contralateral breast. Particularly, significant adjacent organs are lung, liver, heart and contralateral breast in brachytherapy of breast cancer.

Determination of dosimetric dependence for effective atomic number of LDR brachytherapy seed capsule by Monte Carlo simulation

  • Berkay Camgoz;Dilara Tarim
    • Nuclear Engineering and Technology
    • /
    • v.55 no.8
    • /
    • pp.2734-2741
    • /
    • 2023
  • Brachytherapy is a special case of radiotherapy. It should be arranged according to some principles in medical radiation applications and radiation physics. The primary principle is to use as low as reasonably achievable dose in all ionizing radiation applications for diagnostic and therapeutic treatments. Dosimetric distributions are dependent on radioactive source properties and radiation-matter interactions in an absorber medium such as phantom or tissue. In this consideration, the geometrical structure and material of the seed capsule, which surrounds a radioactive material, are directly responsible for isodose profiles and dosimetric functions. In this study, the radiometric properties of capsule material were investigated on dose distribution in a water phantom by changing its nuclear properties using the EGSnrc Monte Carlo (MC) simulation code. Effective atomic numbers of hypothetic mixtures were calculated by using different elements with several fractions for capsule material. Model 6711 brachytherapy seed was modeled by EGSnrc/Dosrcnrc Code and dosimetric functions were calculated. As a result, dosimetric parameters of hypothetic sources have been acquired in large-scale atomic number. Dosimetric deviations between the data of hypothetic seeds and the original one were analyzed. Unit dose (Gy/Particle) distributions belonging to different types of material in seed capsule have remarkably differed from the original capsule's data. Capsule type is major variable to manage the expected dose profile and isodose distribution around a seed. This study shows us systematically varied scale of material type (cross section or effective atomic number dependent) offers selective material usage in production of seed capsules for the expected isodose profile of a specific source.

The Activity Check of Brachytherapy Isotope (근접치료동위원소의 Activity Check)

  • Kim, Gun-Oh;Lee, Byung-Koo;Kwon, Young-Ho
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.16 no.1
    • /
    • pp.21-27
    • /
    • 2004
  • An isotope Ir-192, which is used in brachytherapy depends on import in whole quantities. There are a few ways for its activity. measurement using Welltype chamber or the way to rely on authentic decay table of manufacturer. In-air dosimetry using Farmer Chamber, etc. In this paper, let me introduce the way using Farmer chamber which is easier and simple. With the Farmer chamber and source calibration jig, take a measurement the activity of an isotope Ir-192 and compare the value with the value from decay table of manufacturer and check the activity of source. The result of measurement, compared the value from decay table, by ${\pm}2.1\%$. (which belongs to recommendable value for AAPM ${\pm}5\%$ as difference of error range) It is possible to use on clinical medicine. With the increase in use of brachytherapy, the increase of import is essential. And an accurate activity check of source is compulsory. For the activity check of source, it was possible to use Farmer chamber and source calibration jig without additional purchase of Well type chamber.

  • PDF

Evaluation of Absorbed Dose According to the Gold Nanoparticle Density in Prostate Cancer Brachytherapy (전립선암의 근접치료 시 금 나노입자 밀도에 따른 흡수선량평가)

  • Lee, Deuk-Hee;Kim, Jung-Hoon
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.2
    • /
    • pp.247-252
    • /
    • 2019
  • This study was evaluated absorbed dose according to the gold nanoparticle density in prostate brachytherapy which was constantly occurred in Korean men. Absorbed dose evaluation was using MCNPX program which was applied Monte Carlo simulation. Source were applied $^{192}Ir$ which was temporary insertion source and $^{103}Pd$ which was permanently insertion source. And gold nanoparticle density was applied 0 mg, 7 mg, 18 mg and 30 mg. The prostate absorbed dose was increased in proportion to the density 2.95E-14 Gy/e to 4.42E-14 Gy/e in $^{192}Ir$ and showed the same tendency in $^{103}Pd$. And surrounding organ absorbed dose was inversely proportional to the density. Therefore using nanoparticle in brachytherapy was increased therapeutic ratio.

EVALUATION OF BRACHYTHERAPY FACILITY SHIELDING STATUS IN KOREA OBTAINED FROM RADIATION SAFETY REPORTS

  • Keum, Mi Hyun;Park, Sung Ho;Ahn, Seung Do;Cho, Woon-Kap
    • Nuclear Engineering and Technology
    • /
    • v.45 no.5
    • /
    • pp.695-700
    • /
    • 2013
  • Thirty-eight radiation safety reports for brachytherapy equipment were evaluated to determine the current status of brachytherapy units in Korea and to assess how radiation oncology departments in Korea complete radiation safety reports. The following data was collected: radiation safety report publication year, brachytherapy unit manufacturer, type and activity of the source that was used, affiliation of the drafter, exposure rate constant, the treatment time used to calculate workload and the HVL values used to calculate shielding design goal values. A significant number of the reports (47.4%) included the personal information of the drafter. The treatment time estimates varied widely from 12 to 2,400 min/week. There was acceptable variation in the exposure rate constant values (ranging between 0.469 and 0.592 ($R{\cdot}m^2/Ci{\cdot}hr$), as well as in the HVLs of concrete, steel and lead for Iridium-192 sources that were used to calculate shielding design goal values. There is a need for standard guidelines for completing radiation safety reports that realistically reflect the current clinical situation of radiation oncology departments in Korea. The present study may be useful for formulating these guidelines.

Measurement of Ir-192 Source Activity for High Dose Rate Brachytherapy (고 선량률 근접치료시 사용되는 Ir-192 선원의 방사능 평가)

  • 최동락;허승재;안용찬;임도훈;김대용;우홍균
    • Progress in Medical Physics
    • /
    • v.8 no.1
    • /
    • pp.25-29
    • /
    • 1997
  • Ir-192 source activity for high dose rate brachytherapy is measured using Farmertype ionization chamber. The source-to-chamber distance is 10 cm and the measured charge unit is converted to activity unit. The measured values are compared to the values provided from vendor. Because of time dependency of Ir-192 source activity, the activities are regularly checked and compared to calculated values. As the accuracy of Ir-192 source activity is depend on the mechanical measurement setup, we estimated the precision of remote controlled source dwell position using home-made device and film scanner. The difference between measured and predicted dwell position is within 1 mm. As a result, the errors of source activity are 0.7${\pm}$1.5 % for measured and vendor-provided values and 0.l${\pm}$1.2% for measured and time-dependent calculated vlaues. In conclusion, our measured activity has been comparable to the values provided from vendor and our brachytherapy unit has been very accurate until now. Regular quality control of brachytherapy is essential for successful treatment which depends on the accuracy of source position and activity.

  • PDF

Calibration of $^{192}Ir$ HDR Brachytherapy Source in Air and in a Cylindrical Phantom

  • Djarwani S. Soejoko;I, Arief-Riva'
    • Proceedings of the Korean Society of Medical Physics Conference
    • /
    • 2002.09a
    • /
    • pp.23-27
    • /
    • 2002
  • Two $^{192}$ Ir HDR brachytherapy sources were calibrated with a Farmer ionization chamber in air method and in a PMMA cylindrical phantom. The calibration air method used ionization chamber with buildup cap, and 8 variation distances for center-to-center of the source to chamber. In the optimum distance the measured activity, especially for the high activity source, deviation was 0.3% from the activity provided by manufacturer. Calibration with a PMMA cylindrical phantom was less sensitive, and suitable for quick check method with accuracy less than 10%.

  • PDF

Real Time On-line Quality Assurance System for HDR Brachytherapy (고선량률 근접 방사선 치료를 위한 실시간 온-라인 정도 관리(QA) 시스템 개발)

  • Lee Su Jin;Lee Re Na;Yi Byang Yang;Lim Sang Waak;Choi Jin Ho
    • Progress in Medical Physics
    • /
    • v.15 no.3
    • /
    • pp.156-160
    • /
    • 2004
  • An essential quality assurance (QA) procedure in high dose rate (HDR) remote after-loading brachytherapy is that of the verification of the Ir-192 HDR source positioning accuracy. A number of methods using mechanical rulers or autoradiograph and video cameras have been reported to check the positional error of the Ir-192 source. In this study, the feasibility of a CMOS (Complementary Metal Oxide Semiconductor) PC camera, with a fluorescent screen, was investigated. The agreement between the planned and measured dwell position was better than 1 mm and dwell times better than 0.4 sec. Our results indicate that the CMOS PC camera system could be used as a QA tool for the on-line determination of the source position and dwell time.

  • PDF

Treatment Planning Software for High Dose Rate Remote Afterloading Brachytherapy of Uterine Cervical Cancer (Personal computer를 이용한 자궁경부암의 고선량을 강내치료 계획)

  • Huh, Seung-Jae;Kang, Wee-Saing
    • Radiation Oncology Journal
    • /
    • v.4 no.2
    • /
    • pp.183-186
    • /
    • 1986
  • In brachytherapy of uterine cervical cancer using the high dose rate remote afterloading system, it is of prime importance to determine the position of the radiation sources and to estimate the irradiation time. However, calculation with manual method is so time consuming and laborious, that authors designed a software as an aid to intracavitary radiotherapy Planning using the personal computer to obtain the precision of treatment without being too complicated for routine use. Optimal source arrangement in combination with dose rate at each specific points and irradiation time can be easily determined using this software in several minutes.

  • PDF