Real Time On-line Quality Assurance System for HDR Brachytherapy

고선량률 근접 방사선 치료를 위한 실시간 온-라인 정도 관리(QA) 시스템 개발

  • 이수진 (이화여자대학교 의과대학 목동병원 방사선종양학과) ;
  • 이레나 (이화여자대학교 의과대학 목동병원 방사선종양학과) ;
  • 이병용 (울산대학교 의과대학 아산병원 방사선종양학과) ;
  • 임상욱 (울산대학교 의과대학 아산병원 방사선종양학과) ;
  • 최진호 (가천 의과대학 길병원 방사선종양학과)
  • Published : 2004.09.01

Abstract

An essential quality assurance (QA) procedure in high dose rate (HDR) remote after-loading brachytherapy is that of the verification of the Ir-192 HDR source positioning accuracy. A number of methods using mechanical rulers or autoradiograph and video cameras have been reported to check the positional error of the Ir-192 source. In this study, the feasibility of a CMOS (Complementary Metal Oxide Semiconductor) PC camera, with a fluorescent screen, was investigated. The agreement between the planned and measured dwell position was better than 1 mm and dwell times better than 0.4 sec. Our results indicate that the CMOS PC camera system could be used as a QA tool for the on-line determination of the source position and dwell time.

고선량률 후 장착 근접치료 과정에서의 필수적인 정도 관리(QA) 과정은 Ir-192 선원 위치의 정확성을 검증하는 것이다. 눈금자나 방사선 사진(autoradiograph), 비디오 모니터(video monitor) 등을 이용한 여러 방법들이 Ir-192 선원의 위치적인 오차를 점검하기 위해 사용됨이 보고되고 있다. 본 논문에서는, fluorescent screen 과 CMOS (Complementary Metal Oxide Semiconductor) PC 카메라를 이용하는 방법을 새로운 고선량률 근접치료의 정도 관리 (QA) 도구로 사용할 수 있는 가능성에 대해 연구하였다. 계획된 dwell position 과 그 측정값은 1 mm 이내에서 일치함을 보였고, dwell time의 정확성은 0.4초 이내였다. 이번 연구의 결과는 CMOS PC 카메라를 사용하여 새롭게 제작된 정도 관리(QA) 시스템이 선원의 위치와 dwell time점검이 실시간으로 가능한 정도 관리(QA) 도구로 사용될 수 있음을 보여준다.