• Title/Summary/Keyword: 유방X선촬영

Search Result 30, Processing Time 0.025 seconds

Variation of Image Quality and Dose by Applying Multi-Leaf Collimator for Digital Mammography (디지털 유방촬영장치에서 다엽 조리개 적용으로 인한 화질과 선량의 변화)

  • Kwon, Soon Mu;Kim, Boo Soon;Park, Hyung Jun;Kang, Yeong Han
    • Journal of the Korean Society of Radiology
    • /
    • v.9 no.7
    • /
    • pp.535-540
    • /
    • 2015
  • Collimator has important functions with control primary X-ray that decrease radiation exposure dose for patients and reduce scatter ray and make better quality of image. But there are no regulations for X-ray mammography device of collimator, so widely used device adopt rectangularly controlled collimator. Though digital X-ray mammography device expand supply recently, rectangularly controlled collimator of film/screen mode still used. After searching for real condition of beam field with digital mammography, we made a multi-leaf collimator which is able to adjust the beam field in accordance with size and shape of breast, and we measuring up the transitions of image quality, average glandular dose(AGD) and, Dose area product(DAP). There are no significant differences between rectangularly controlled collimator and multi-leaf collimator, and DAP value decreased by 50.72%. As conclusion, there needs to expand the use of multi-leaf collimator for optimum adoption of beam field in digital mammography, and also need to develop an automatic regulation of beam field for reduce of exposure dose to patients.

The Evaluation of the Thick Polycrystalline HgO and PbO Films Derived by Particle Sedimentation Method for the Mammographic Application (입자침전법을 이용한 다결정 산화수은과 산화납 필름의 방사선 유방촬영 장치 적용성 평가)

  • Noh, Si-Cheol;Park, Ji-Koon;Choi, Il-Hong;Jung, Hyoung-Jin;Kang, Sang-Sik;Jung, Bong-Jae
    • Journal of the Korean Society of Radiology
    • /
    • v.8 no.7
    • /
    • pp.429-433
    • /
    • 2014
  • In this study, the morphology and the x-ray quantum efficient of mercury oxide (HgO) and lead oxide (PbO) sensors derived by particle sedimentation method were discussed. In the pursuit of this purpose, we investigated the electrical characteristics and the x-ray quantum efficiency of various thicknesses of HgO and PbO films in mammographic x-ray energy. We have therefore developed a particle-in-binder sedimentation method of fabricating large area polycrystalline films onto transparent glass substrates coated with indium tin oxide. We are currently optimizing the growth method to improve the quantum efficiency with the ultimate goal of obtaining as quantum efficiency close to that of single crystal performance. Our future efforts will concentrate on optimization of large area film growth techniques specifically for deposition on a-Si:H flat panel readout arrays.

A study on Quality Control of Mammography Equipment in the Gwang-Ju (일부지역 유방 촬영 장비의 정도관리 실태조사 연구)

  • Shin, Ji-Yun;Lee, Sang-Bock;Jin, Gye-Hwan;Lee, Sam-Yeol;Lee, Jun-Haeng
    • Journal of the Korean Society of Radiology
    • /
    • v.2 no.4
    • /
    • pp.11-25
    • /
    • 2008
  • As breast cancer shows the highest incidence among women's cancers, the use of mammography is also increasing as a screening test. Mammography should produce high-quality images for accurate diagnosis. For this, it is necessary to manage the performance and image quality of mammography. Thus, in order to investigate quality control, the present study conducted a questionnaire survey of 37 hospitals registered as breast cancer medical examination centers in Gwang-Ju, concerning their quality control of mammography. In the results of surveying the characteristics of apparatus use for mammography, many respondents did not know about the equipment that they were using. Of the hospitals, 19 (49%) were using film, 19 (49%) CR, and 1 DR. In the results of asking how to do quality control, only 38% answered that they inspected according to the manual. In addition, all the surveyed hospitals had specialized agencies do quality control for them. As to the reason for using agencies, 65% mentioned limitations in personnel, time, distance, etc, and 44% mentioned the absence of machines and materials. These results suggest that quality control is being performed perfunctorily, as well as passively and indifferently as it relies on specialized agencies. Accordingly, it is necessary to provide regular education for enhancing people's perception of quality control and to perform quality control adequately in the presence of Radiologist.

  • PDF

Evaluation of Linearity Air Kerma Applied to the IEC 60601-2-45 Standard in Mammographic X-ray (유방촬영용 X선 진단장치에서 IEC 60601-2-45 표준규격을 적용한 직선성 평가)

  • Hong, Dong-Hee;Jung, Hong-Ryang;Lim, Cheong-Hwan;Han, Beom-Hee;Han, Sang-Hyun;Lee, Sang-Ho;Mo, Eun-Hee;Kim, Ki-Jung;Lee, Mi-Hwa;You, In-Gyu
    • Journal of Digital Convergence
    • /
    • v.12 no.3
    • /
    • pp.339-344
    • /
    • 2014
  • The quality control items of mammography devices in South Korea do not include the linearity, which is required by international standards. The linearity is a requirement for the adjustment of radiation dose and radiation quality. This study tested the linearity, which was suitable for the IEC 60601-2-45 standard, of the 5 mammography devices. All showed adequate results. Consistent measurement management is required for more developed quality control in the future.

Testing and Analysis of Tube Voltage and Tube Current in The Radiation Generator for Mammography (유방촬영용 방사선발생장치의 관전압과 관전류 시험 분석)

  • Jung, Hong-Ryang;Hong, Dong-Hee;Han, Beom-Hui
    • Journal of radiological science and technology
    • /
    • v.37 no.1
    • /
    • pp.1-6
    • /
    • 2014
  • Breast shooting performance management and quality control of the generator is applied to the amount of current IEC(International Electrotechnical Commission) 60601-2-45 tube voltage and tube current are based on standards that were proposed in the analysis of the test results were as follows. Tube voltage according to the value of the standard deviation by year of manufacture from 2001 to 2010 as a 42-3.15 showed the most significant, according to the year of manufacture by tube amperage value of the standard deviation to 6.38 in the pre-2000 showed the most significant, manufactured after 2011 the standard deviation of the devices, the PAE(Percent Average Error) was relatively low. This latest generation device was manufactured in the breast of the tube voltage and tube diagnosed shooting the correct amount of current to maintain the performance that can be seen. The results of this study as the basis for radiography diagnosed breast caused by using the device's performance and maintain quality control, so the current Food and Drug Administration "about the safety of diagnostic radiation generator rule" specified in the test cycle during three years of self-inspection radiation on a radiation generating device ensure safety and performance of the device using a coherent X-ray(constancy) by two ultimately able to keep the radiation dose to the public to reduce the expected effect is expected.

Comparison of Average Glandular Dose in Screen-Film and Digital Mammography Using Breast Tissue-Equivalent Phantom (유방조직등가 팬텀을 이용한 Screen-Film과 Digital Mammography에서의 평균 유선선량)

  • Shin, Gwi-Soon;Kim, Jung-Min;Kim, You-Hyun;Choi, Jong-Hak;Kim, Chang-Kyun
    • Journal of radiological science and technology
    • /
    • v.30 no.1
    • /
    • pp.13-23
    • /
    • 2007
  • In recent years, mammography system is changed rapidly from conventional screen-film system to digital system for application to screening and diagnosis. Digital mammography system provides several advantages over screen-film mammography system. According to the information provided by the manufacturer, digital mammography system offers radiation dose reduction in comparison with screen-film mammography system, because of digital detector, particularly direct digital detector has higher x-ray absorption efficiency than screen-film combination or imaging plate(IP). We measured average glandular doses(AGD) in screen-film mammography(SFM) system with slow screen-film combination, computed mammography(CM) system, indirect digital mammography(IDM) system and direct digital mammography(DDM) system using brest tissue-equivalent phantom(glandularity 30%, 50% and 70%). The results were shown as follows : AGD values for DDM system were highest than those for other systems. Although automatic exposure control(AEC) mode was selected, the curve of the AGD values against thickness or glandularity increased significantly for the SFM system with the uniform target/filter(Mo/Mo) combination. Therefore, the AGD values for the high energy examinations were highest in the SFM system, and those for the low energy examinations were highest in the DDM system. But the curve of the AGD values against thickness and glandularity increased gently for CM system with the automatic selection of the target/filter combination (from Mo/Mo to Mo/Rh or from Mo/Rh to Rh/Rh), and the AGD values were lowest. Consequently, the parameters in mammography for each exposure besides detection efficiency play an important role in oder to estimate a patient radiation dose.

  • PDF

Analysis of Radiation Dose for Lens, Thyroid Gland, Breast, and Gonad on Upper Gastrointestinal Series (위장조영검사에서 수정체, 갑상선, 유방, 생식선에 대한 피폭선량 분석)

  • Lim, Byung-Hak;Chon, Kwon Su
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.6
    • /
    • pp.889-894
    • /
    • 2019
  • Upper gastrointestinal series is an examination that uses X-rays. It is important to defend against exposure to radiation during upper gastrointestinal examination because the organs, such as thyroid gland, lens, breasts, and gonads, with relatively high biological sensitivity to radiation are distributed on the examination area. We have made a whole body phantom that can change the depth of organs. radiation dose of eye, thyroid gland, breast and gonads were measured by the same procedure as the actual upper gastrointestinal examination. When performed only fluoroscopy the mean dose reduction of lens, thyroid gland, breast and gonads was 62.2%. The mean dose reduction of lens, thyroid gland, breast and gonads was 59.0% when both fluoroscopy and spot shoot were performed. Therefore, when performed upper gastrointestinal examination it was confirmed that shielding of the lens, thyroid gland, breast and gonads was effective in decreasing the exposure dose. The manufactured human phantom can be used in measuring radiation dose for deep organ because it can adjust the height corresponding to the organs located in the human body.

Evaluation of Image Quality using Monte Carlo Simulation in Digital Mammography System (디지털유방영상시스템에서 몬테카를로 시뮬레이션을 이용한 영상평가)

  • Kim, Changsoo;Kang, Se-Sik;Kim, Jung-Hoon;Lee, Jin-Soo
    • The Journal of the Korea Contents Association
    • /
    • v.14 no.6
    • /
    • pp.247-254
    • /
    • 2014
  • For the purpose of early diagnosis of the breast cancer, the attention on the screening mammography has been increasing-with supply of digital mammography day by day. Image quality is decided by target materials and filter combinations. Optimized selection by a glandular density and a thickness is needed, because these combinations change x-ray spectrum and effect to image quality. The purpose of this study is to find out optimized target and filter combinations through the evaluation of quantitative image quality and to suggest means which minimize patient dose through MCNPX. In results, spatial frequency resolution evaluation which is quantitative image quality evaluation method, MTF, NPS, DQE shows that we have to select Mo/Mo combinations or Mo/Rh combinations when compressed breast is thin. but in case of that when compressed breast is thick, we have to select Rh/Rh combinations or W /Rh combinations. In a comprehensive evaluation, W!Rh combinations which are not used in thin breasts in practice was superior to all image quality evaluation. This result is somewhat different-with clinical examination results. Secondary end point was organ dose evaluation, radiation dose of opposite breast was approximately 47 ~73% effectiveness when selecting standard breast. In conculsion, the most important point is that we have to select the optimal combinations-with considering dose evaluation and various thickness.

A Study on Absorbed Dose in the Breast Tissue using Geant4 simulation for Mammography (유방촬영에서 Geant4 시뮬레이션를 이용한 유방조직내 흡수선량에 관한 연구)

  • Lee, Sang-Ho;Lee, Jong-Seok;Han, Sang-Hyun
    • Journal of radiological science and technology
    • /
    • v.35 no.4
    • /
    • pp.345-352
    • /
    • 2012
  • As the breast cancer rate is increasing fast in Korean women, people pay more attention to mammography and number of mammography have been increasing dramatically over the last few years. Mammography is the only means to diagnose breast cancer early, but harms caused by radiation exposure shouldn't be overlooked. Therefore, it is important to calculate the radiation dose being absorbed into the breast tissue during the process of mammography for a protective measure against radiation exposure. Because it is impossible to directly measure the radiation dose being absorbed into the human body, statistical calculation methods are commonly used, and most of them are supposed to simulate the interaction between radiation and matter by describing the human body internal structure with anthropomorphic phantoms. However, a simulation using Geant4 Code of Monte Carlo Method, which is well-known as most accurate in calculating the absorbed dose inside the human body, helps calculate exact dose by recreating the anatomical human body structure as it is through the DICOM file of CT. To calculate the absorbed dose in the breast tissue, therefore, this study carried out a simulation using Geant4 Code, and by using the DICOM converted file provided by Geant4, this study changed the human body structure expressed on the CT image data into geometry needed for this simulation. Besides, this study attempted to verify if the dose calculation of Geant4 interlocking with the DICOM file is useful, by comparing the calculated dose provided by this simulation and the measured dose provided by the PTW ion chamber. As a result, under the condition of 28kVp/190mAs, the Difference(%) between the measured dose and the calculated dose was found to be 0.08 %~0.33 %, and at 28 kVp/70 mAs, the Difference(%) of dose was 0.01 %~0.16 %, both of which showed results within 2%, the effective difference range. Therefore, this study found out that calculation of the absorbed dose using Geant4 Simulation is useful in measuring the absorbed dose in the breast tissue for mammography.

Evaluation of Scattered Rays of Jelly Type Shielding Body by L-spine AP using X-ray (L-Spine X-선 촬영에서의 Jelly type 차폐체의 산란선 차폐평가)

  • Jang, Hui-Min;Kim, Do-Gwon;Kim, Hyeong-Bin;Yoon, Joon
    • Journal of the Korean Society of Radiology
    • /
    • v.14 no.7
    • /
    • pp.907-913
    • /
    • 2020
  • There have been continuous controversies on medical X-ray protection and numerous researchers have been trying to prevent unnecessary exposure to radiation. As X-ray passes through the patient and obtains an image, it creates scattered ray due to interactions such as photoelectric effect and Compton scattering with the subject. As a result, both medical radiation staff and patient are exposed to unnecessary radiation on areas other than the target area. In response, this study will be assuming a body of a female, radiating X-ray on the phantom under the conditions of lumbar spine AP test, and measuring scattered ray around breasts and thyroid glands. Then, The experiment results were as follows. After application of non-shielding material, the average of scattered ray was 0.88 mR in thyroid measurement, 3.34 mR, Lt Axillary 3.54 mR, and Rt Axillary 3.03 mR in mamonary measurement but, After application of shielding material, the average of scattered ray was 0.16 mR in thyroid measurement, 0.60 mR, Lt Axillary 0.64 mR, and Rt Axillary 0.54 mR in mamonary measurement showing average scattered ray protection effect of about 82%. This study suggested the manufacturing method of a Jelly-type shielding material, identified the possibilities of researches on mixing various substances with radiology field, and verified the usability of the Jelly-type shielding material as a substitute for existing protection tools.