• Title/Summary/Keyword: 도플러 초음파 영상

Search Result 38, Processing Time 0.031 seconds

Influence to the Doppler Images by the Defects of SAE in the Probe of Medical Ultrasonic Scanners (초음파 프로브에서 인접 단위 소자군(SAE) 결함이 도플러 영상에 미치는 영향)

  • Lee, Kyung-Sung
    • Journal of radiological science and technology
    • /
    • v.38 no.1
    • /
    • pp.7-15
    • /
    • 2015
  • A ultrasonic probe is very important in medical ultrasonic image, but the frequency of probe defects are often. Therefore practical tools for probe based ultrasonic QA should be developed. Advanced research on the effects of the probe defects on the quality of ultrasonic images is required. This study has investigated the effects of the defects in the probe elements influence Doppler images in the medical ultrasonic scanners. Especially the defects in a set of adjacent elements(SAE) electrically disconnected influence Doppler images were tested. The results show Doppler brightness and velocity became rapidly reduced as the defected elements is located centrally, as the defected elements is activated. The more the defected elements increased, the more Doppler brightness and velocity increased. As a set of the element disconnected moved, it appeared Doppler velocity starting to decrease and then was followed by brightness. The strength is not consistent with the velocity in the number and location of the defected elements. The defects in the probe elements influence Doppler velocity when the defected elements got out of the elements activated at Doppler mode.

The Impacts of Piezoelectric Elements' Defects On Color & Power Doppler Images (초음파 프로브에서 소자결함이 컬러 및 파워 도플러 영상에 미치는 영향)

  • Lee, Kyung-Sung
    • Journal of radiological science and technology
    • /
    • v.38 no.4
    • /
    • pp.443-449
    • /
    • 2015
  • An ultrasound probe has a big impact on Doppler images even though it has very high risk of frequent function-breakdowns occurring in medical ultrasound scanners. This study experimentally analyses the impacts of an ultrasonic probe's defected elements on power & color Doppler images. The results show that, the bigger the size of defected probe elements is, and the closer a group of action elements is to the center, the more the brightness of images and the velocity of Doppler diminish. When elements' defects increase in color & power Doppler images, false images are formed to be mistaken for blood-vessel plaque in neighboring regions. Accordingly, whenever element defects are suspected, we need check-up process in B-mode. From this respective, it is advisable to have primary interest in a probe and carry out continuous probe QA for ultrasonography.

Extraction of Hypertension Blood flow of Brachial Artery from Color Doppler Ultrasonography by Using 4-directional Contour Tracking Algorithm and Enhanced FCM Method (4 방향 윤곽선 추적 알고리즘과 개선된 FCM 방법을 이용한 색조 도플러 초음파 영상에서 상완 동맥의 고혈압 혈류 추출)

  • Yu, Seong-won;Jung, Young-hun;Shim, Sung-bo;Kim, Hye-ran;Kim, Min-ji;Kim, Kwang Beak
    • Proceedings of the Korean Institute of Information and Commucation Sciences Conference
    • /
    • 2017.05a
    • /
    • pp.71-73
    • /
    • 2017
  • 본 논문에서는 4 방향 윤곽선 추적 기법과 히스토그램 분석 기법을 기반으로 한 개선된 FCM 알고리즘을 적용하여 색조 도플러 초음파 영상에서 상완 동맥의 혈류를 추출하고 분석하는 방법을 제안한다. 제안된 방법에서는 상완 동맥의 혈류를 정확히 추출하기 위해 전처리 과정으로 색조 도플러 초음파 영상 이외의 환자 정보가 있는 영역을 제거한 후, ROI 영역을 추출한다. 추출된 ROI 영역에서 영상의 최대 명암도를 임계치로 설정한 이진화 기법을 적용하여 ROI 영역을 이진화한다. 이진화된 ROI 영역에서 4 방향 윤곽선 추적 기법을 적용하여 상완 동맥이 존재하는 사다리꼴 형태의 영역을 추출한다. 색 정보를 분석한 히스토그램을 이용하여 특징점의 개수를 계산하고 계산된 특징점의 개수를 FCM 알고리즘의 초기 클러스터의 개수로 설정한 후, 추출된 사다리꼴 형태의 영역에 적용하여 양자화 한다. 양자화된 영역 중에서 빨간색으로 분류된 영역을 고혈압 영역으로 추출한다. 제안된 추출 방법을 20개의 색조 도플러 초음파 영상을 대상으로 실험한 결과, 20개의 색조 도플러 초음파 영상에서 18개의 색조 도플러 초음파 영상이 정확히 추출되었다.

  • PDF

Influence to the Doppler Images by the Defects of Piezoelectric Elements of the Probe of Medical Ultrasonic Scanners (Focusing on the Impact of an Increase in the Defects of Piezoelectric Elements) (초음파 프로브 소자 결함이 도플러 영상에 미치는 영향 (소자 결함 증가에 따른 영향을 중심으로))

  • Lee, Kyung-Sung
    • Journal of radiological science and technology
    • /
    • v.37 no.2
    • /
    • pp.117-124
    • /
    • 2014
  • This study has investigated the effects of the defects in the probe elements influence Doppler images in the medical ultrasonic scanners. This work was implemented that the quality of Doppler images depended on the extent and location of the probe element defected. The probe performance was rated in terms of the number of piezoelectric elements lost and this was studied in the experiment by electrical disconnection to the elements. The results showed that Doppler velocity became rapidly reduced as the defected elements encountered with the element group activated at the Doppler mode, not as the flow velocity. The effect of the probe defect is decreased when the defects occurred at the element group activated for Doppler mode, as was increased the number of the elements. It was observed that the higher the flow velocity of Doppler flow phantom is, the wider the spectrum of Doppler is. And the Doppler velocity soared up and the dispersion of image brightness is increased when the defected elements got out of the elements activated at Doppler mode. The result showed that TADVP(time-average-Doppler-velocity-profile) is decreased with the increase of the probe element defect, especilly in the region of high frequency. It is expected that the research of various defects of probe elements are needed, and this study can be practical tools for probe based ultrasonic QA in the future.

Velocities Analysis of Hypertension Blood Flow of Brachial Artery on Color Doppler Ultrasonography using IHb Color Information (IHb 색상 정보를 이용한 색조 도플러 초음파 영상에서 상완 동맥의 고혈압 혈류 속도 분석)

  • Oh, Heung-Min;Shim, Sung-Bo;Kim, Kwang Beak
    • Proceedings of the Korean Institute of Information and Commucation Sciences Conference
    • /
    • 2018.05a
    • /
    • pp.366-368
    • /
    • 2018
  • 본 논문에서는 이러한 문제점을 개선하기 위해 상완 동맥 영역에 대한 RGB 채널을 HSV 채널로 변환한다. 변환된 HSV 채널에 대해 고혈압 영역의 특징을 강조하게 하기 위해 밝기 값을 나타내는 V값을 조절한다. 조절된 HSV 채널을 다시 RGB 채널로 변환한 후, Fuzzy C_Means 기반 무게중심과 Possibilistc C_Means 기반 무게 중심을 기반으로 새로운 무게 중심을 구하여 픽셀들을 클러스터링하여 상완동맥 영역의 고혈압 영역을 추출한다. 추출된 상완 동맥의 고혈압 영역에 대해 헤모글로빈 색소 정보를 나타내는 IHb 값을 이용하여 상완 동맥의 고혈압 영역에서 유사한 헤모글로빈 색소 정보를 가지는 영역을 분할한다. 분할된 영역들을 혈류의 속도를 나타내는 색상표와 대조하여 고혈압의 진행에 대해 분석하는 방법을 제안한다. 제안된 방법을 색조 도플러 초음파 영상을 대상으로 실험한 결과, 제안된 방법이 고혈압의 진행에 대한 분석 결과와 색조 도플러 초음파 영상 장비에 나타난 고혈압 진행 결과와 거의 일치하는 것을 확인할 수 있었다.

  • PDF

A New Method for Extending Doppler Mean Frequency in Ultrasonic Imaging Systems (초음파 영상 시스템에서 새로운 도플러 평균주파수 확장 방법)

  • Kwon, Sung-Jae
    • Journal of the Korea Academia-Industrial cooperation Society
    • /
    • v.8 no.5
    • /
    • pp.1047-1056
    • /
    • 2007
  • Basically, an ultrasonic imaging system has two fundamental imaging modes available. One is the B-mode imaging modality which provides an image of reflection coefficient, and the other is the Doppler color flow mode that maps blood flow inside the human heart and blood vessels. This paper presents a new method of detecting and compensating for aliasing that occurs when the Doppler frequency exceeds one-half of the pulse-repetition frequency (PRF). Its validity is shown by computer simulation. The new method not only extends the measurable Doppler frequency, but also helps to reduce the effect of noise. The results show that the aliasing can be compensated for correctly fur signal-to-noise ratios down to 20 dB.

  • PDF

Extraction of Blood Velocity Using FCM and Fuzzy Decision Trees in Doppler Ultrasound Images of Brachial Artery (상완동맥 색조 도플러 초음파 영상에서 FCM과 퍼지 의사 결정 트리를 이용한 혈류 속도 추출)

  • Kim, Kwang Baek;Jung, Young Jin;Nam, Youn Man;Lee, Jae Yeol
    • Proceedings of the Korean Society of Computer Information Conference
    • /
    • 2019.07a
    • /
    • pp.19-22
    • /
    • 2019
  • 상완동맥은 어깨에서부터 팔꿈치까지 내려오는 상완골의 내측부에 존재하며 혈압을 측정할 때 사용되는 혈관이다. 이 혈관은 골절로 인해 찢어지거나, 또는 혈액순환에 문제가 생겨 혈관이 막히는 경우가 발생한다. 이러한 경우 혈관의 상태를 확인하기 위하여 색조 도플러 초음파 검사를 사용하지만, 사용자에 따라 영상을 통한 판단 기준이 다르다는 문제점이 발생한다. 따라서 본 논문에서는 FCM과 Fuzzy Decision Tree를 이용한 영상 처리를 통해 일관성 있는 판단기준을 세우기 위한 혈류의 속도를 제안한다. 색조 도플러 초음파 영상에서의 상완 동맥을 추출하여 기울기를 이용한 FCM 알고리즘을 통해 소속도를 추출한 뒤 퍼지 룰에 적용하여 의사 결정 트리로 등급을 분류하고 결과적으로 혈류 속도를 추출한다. 색조 도플러 초음파 영상에서 환자의 개인 정보를 보호하기 위해 개인 정보 영역을 제거하여 ROI 영역을 추출하고 ROI 영역을 이진화를 통하여 상완동맥이 있는 영역을 추출한다. 이진화 된 ROI 영역에서 혈관 영상의 혈류 방향으로의 무게중심을 설정하고 각각의 픽셀과 무게중심 선과의 거리를 이용하여 소속도를 추출한 후 FCM을 사용하여 최적의 기울기를 선정한다. FCM을 통해 추출한 최종 소속도를 이용하여 퍼지 룰에 적용한 뒤 계산된 T-norm과 소속도의 분산을 이용하여 의사 결정 트리를 형성 트리의 단말 노드들은 각 픽셀을 분류한다. 분류되어진 데이터들의 노드별 소속도 평균을 구한 뒤 디퍼지화를 통해 COG(Center of Gravity)를 계산한다. 마지막으로 그 값을 이용하여 혈류 속도에 영향을 미치는 정도를 계산한 뒤 최종 혈류의 속도를 제안한다.

  • PDF

Extraction of Blood Flow of Brachial Artery on Color Doppler Ultrasonography by Using 4-Directional Contour Tracking and K-Means Algorithm (4 방향 윤곽선 추적과 K-Means 알고리즘을 이용한 색조 도플러 초음파 영상에서 상환 동맥의 혈류 영역 추출)

  • Park, Joonsung;Kim, Kwang Baek
    • Journal of the Korea Institute of Information and Communication Engineering
    • /
    • v.24 no.11
    • /
    • pp.1411-1416
    • /
    • 2020
  • In this paper, we propose a method of extraction analysis of blood flow area on color doppler ultrasonography by using 4-directional contour tracking and K-Means algorithm. In the proposed method, ROI is extracted and a binarization method with maximum contrast as a threshold is applied to the extracted ROI. 4-directional contour algorithm is applied to extract the trapezoid shaped region which has blood flow area of brachial artery from the binarized ROI. K-Means based quantization is then applied to accurately extract the blood flow area of brachial artery from the trapezoid shaped region. In experiment, the proposed method successfully extracts the target area in 28 out of 30 cases (93.3%) with field expert's verification. And comparison analysis of proposed K-Means based blood flow area extraction on 30 color doppler ultrasonography and brachial artery blood flow ultrasonography provided by a specialist yielded a result of 94.27% accuracy on average.

An implementation of the continuous wave doppler system for blood flow measurement using the ultrasound (효율적인 혈류 속도 측정을 위한 연속 초음파 도플러 장치의 구현)

  • 박형재;김영길
    • Proceedings of the Korean Institute of Information and Commucation Sciences Conference
    • /
    • 2001.05a
    • /
    • pp.516-519
    • /
    • 2001
  • To diagnose a patient's blood vessel disease, apoplexy, hypertension, arteriosclerosis, the blood velocity is very important. Determining the blood velocity methods using ultrasound are Continuous Doppler System and Pulse Doppler System. In using the Pulse Doppler System, we can obtain the position of blood velocity. But it is more complex hardware than Continuous Doppler System and it has low SNR(signal-noise ratio). So in this study, to obtain a believable information we use the Continuous Pulse Doppler System. Thus system have analog part and digital part. In analog part is composed of ultrasound generating part, the amplifying part to amplify the received signal from ultrasound sensor, the demodulation part to detect blood velocity and the filtering part to remove the noise. In digital part is composed of the A/D conversion part, digital signal processing part, and the communication part to communicate the PC. In this study to implement efficient ultrasound blood velocity measurement system, we can get the patient's blood velocity information in realtime. Thus, It is a useful in the accurate diagnosis with C.T(computered tomography), M.R.I(magnetic resonance imaging).

  • PDF

Usefulness of Twinkling Artifacts in Color Doppler Ultrasonography (컬러 도플러 초음파에서 Twinkling artifacts의 유용성)

  • Sim, Hyun-Sun;Kwon, Kyung-Tae
    • The Journal of the Korea Contents Association
    • /
    • v.16 no.10
    • /
    • pp.291-298
    • /
    • 2016
  • The aim of the study was to investigate the diagnostic value of the color Doppler twinkling artifact in disease of urinary system. The intensity of twinkling artifact(TA) with color Doppler was classified into 3 levels, 0(non-TA) to 3(distinct TA). In the in vitro study, acorn jelly with various materials on top was examined using color Doppler at B-mode sonography in a water bath for TA. 31 patients with diagnosis of urinary calculi(renal stones 16, urinary stones 15) based on B-mode sonography were studied in vivo for TA. The materials with rough surfaces such as salt, screw and cubics at B-mode sonography with color Doppler contributed to causing TA. At B-mode sonography without color Doppler 37% of renal stones and 60% of ureter stones were detected. but at B-mode sonography with color Doppler TA was demonstrated for all cases. Superficial roughness of materials affected occurrence of TA at B-mode sonography with color Doppler. Therefore, TA at B-mode sonography without color Doppler could play a role in confident diagnosis of the disease of urinary system.