Abstract
This study has investigated the effects of the defects in the probe elements influence Doppler images in the medical ultrasonic scanners. This work was implemented that the quality of Doppler images depended on the extent and location of the probe element defected. The probe performance was rated in terms of the number of piezoelectric elements lost and this was studied in the experiment by electrical disconnection to the elements. The results showed that Doppler velocity became rapidly reduced as the defected elements encountered with the element group activated at the Doppler mode, not as the flow velocity. The effect of the probe defect is decreased when the defects occurred at the element group activated for Doppler mode, as was increased the number of the elements. It was observed that the higher the flow velocity of Doppler flow phantom is, the wider the spectrum of Doppler is. And the Doppler velocity soared up and the dispersion of image brightness is increased when the defected elements got out of the elements activated at Doppler mode. The result showed that TADVP(time-average-Doppler-velocity-profile) is decreased with the increase of the probe element defect, especilly in the region of high frequency. It is expected that the research of various defects of probe elements are needed, and this study can be practical tools for probe based ultrasonic QA in the future.
Keywords
Ultrasonic probe; Piezoelectric element; Doppler image; TADVP;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Moore G W, Gessert J and Schafer M : Perspectives On Image Performance Assessment II : The Impact of Compromised Elements Within An Ultrasound Transducer On Image and Doppler Quality. American Association of Physicists in Medicine (AAPM) 45th Annual Meeting, San Diego, California, USA, 10-14 Aug 2003
2 Price R : Routine Quality Assurance of Ultrasound Imaging Systems, The Institute of Physical Sciences in Medicine (IPSM), Report No.71: York, UK, 1995
3 Kang G S, Choi M J, Park S S, Paeng D G, Gibson N M and Coleman A J : Deterioration of the quality of ultrasonic images caused by probe defects. Proc. Acoust. Soc. Kor. Conference 2005, 24, 165-168, 2005
4 Choi M J, Kang K S, Lee K S, Yang J H, Paeng D G , Kim C C, Rhim S, Gibson N and Coleman A : Low Contrast Sensitivity of B-mode Images against Ultrasonic Probe Defects, The World Federation for Ultrasound in Medicine & Biology, KOEX, Seoul, Korea, 28 May-1 June 2006
5 서대철, 권태원 역 : 최신 혈관초음파검사, 고려의학, 2007
6 Burns P N : The physical principles of Doppler and spectral analysis. J. Clinical Ultrasound, 15, 567-590, 1987   DOI   ScienceOn
7 Chivers R C and Parry R J : Ultrasonic velocity and attenuation in mammalian tissue. J Acoust Soc Am, 63, 940-953, 1978   DOI   ScienceOn
8 Hoskins P R, Sherriff S B and Evans J A (ed) : Testing of Doppler Ultrasound Equipment, The Institute of Physical Sciences in Medicine (IPSM), Report No.70: York, UK, 1994
9 Merritt C R B : Physics of Ultrasound: In Diagnostic Ultrasound. 3rd ed. Mosby: St. Louis, 3-34, 2005
10 Taylor K J and Holland S : Doppler US. Part I. Basic principles, instrumentation and pitfalls. Radiology, 174, 297-307, 1990   DOI
11 Fish P : Physics and instrumentation of diagnostic medical ultrasound. John Wiley & Suns, 17-19, 57-64 and 172, 1990
12 Alkinson P and Woodcock J P : Doppler ultrasound and its use in clinical measurement, Academic Press: New York, 1982
13 Boote E J : Doppler US Techniques: Concepts of Blood Flow Detection and Flow Dynamics. Radiographics, 23, 1315-1327, 2003   DOI   ScienceOn
14 Browne J E, Watson A J, Gibson N M, Dudley N J and Elliott A T : Objective measurements of image quality. Ultrasound in Med. & Biol., 30, 229-237, 2004   DOI   ScienceOn
15 Dudley N J, Griffith K, Houldsworth G, Holloway M and Dunn M A : A review of two alternative ultrasound quality assurance programmes. European Journal of Ultrasound, 12, 233-245, 2001   DOI   ScienceOn