Influence to the Doppler Images by the Defects of SAE in the Probe of Medical Ultrasonic Scanners (초음파 프로브에서 인접 단위 소자군(SAE) 결함이 도플러 영상에 미치는 영향)

  • Lee, Kyung-Sung
    • Journal of radiological science and technology
    • /
    • v.38 no.1
    • /
    • pp.7-15
    • /
    • 2015
  • A ultrasonic probe is very important in medical ultrasonic image, but the frequency of probe defects are often. Therefore practical tools for probe based ultrasonic QA should be developed. Advanced research on the effects of the probe defects on the quality of ultrasonic images is required. This study has investigated the effects of the defects in the probe elements influence Doppler images in the medical ultrasonic scanners. Especially the defects in a set of adjacent elements(SAE) electrically disconnected influence Doppler images were tested. The results show Doppler brightness and velocity became rapidly reduced as the defected elements is located centrally, as the defected elements is activated. The more the defected elements increased, the more Doppler brightness and velocity increased. As a set of the element disconnected moved, it appeared Doppler velocity starting to decrease and then was followed by brightness. The strength is not consistent with the velocity in the number and location of the defected elements. The defects in the probe elements influence Doppler velocity when the defected elements got out of the elements activated at Doppler mode.

The Impacts of Piezoelectric Elements' Defects On Color & Power Doppler Images (초음파 프로브에서 소자결함이 컬러 및 파워 도플러 영상에 미치는 영향)

  • Lee, Kyung-Sung
    • Journal of radiological science and technology
    • /
    • v.38 no.4
    • /
    • pp.443-449
    • /
    • 2015
  • An ultrasound probe has a big impact on Doppler images even though it has very high risk of frequent function-breakdowns occurring in medical ultrasound scanners. This study experimentally analyses the impacts of an ultrasonic probe's defected elements on power & color Doppler images. The results show that, the bigger the size of defected probe elements is, and the closer a group of action elements is to the center, the more the brightness of images and the velocity of Doppler diminish. When elements' defects increase in color & power Doppler images, false images are formed to be mistaken for blood-vessel plaque in neighboring regions. Accordingly, whenever element defects are suspected, we need check-up process in B-mode. From this respective, it is advisable to have primary interest in a probe and carry out continuous probe QA for ultrasonography.

Speech synthesis using acoustic Doppler signal (초음파 도플러 신호를 이용한 음성 합성)

  • Lee, Ki-Seung
    • The Journal of the Acoustical Society of Korea
    • /
    • v.35 no.2
    • /
    • pp.134-142
    • /
    • 2016
  • In this paper, a method synthesizing speech signal using the 40 kHz ultrasonic signals reflected from the articulatory muscles was introduced and performance was evaluated. When the ultrasound signals are radiated to articulating face, the Doppler effects caused by movements of lips, jaw, and chin observed. The signals that have different frequencies from that of the transmitted signals are found in the received signals. These ADS (Acoustic-Doppler Signals) were used for estimating of the speech parameters in this study. Prior to synthesizing speech signal, a quantitative correlation analysis between ADS and speech signals was carried out on each frequency bin. According to the results, the feasibility of the ADS-based speech synthesis was validated. ADS-to-speech transformation was achieved by the joint Gaussian mixture model-based conversion rules. The experimental results from the 5 subjects showed that filter bank energy and LPC (Linear Predictive Coefficient) cepstrum coefficients are the optimal features for ADS, and speech, respectively. In the subjective evaluation where synthesized speech signals were obtained using the excitation sources extracted from original speech signals, it was confirmed that the ADS-to-speech conversion method yielded 72.2 % average recognition rates.

Representation of spatial variation and particle mean size for suspended sediment using acoustic backscatter in controlled channel experiments (수리실험을 통한 통제된 유사 조건에서의 횡방향 초음파도플러유속계 초음파산란도 활용 부유사농도 분포 및 평균 입경 분석)

  • Son, Geunsoo;Kim, Jongmin;Kang, Woochul
    • Proceedings of the Korea Water Resources Association Conference
    • /
    • 2022.05a
    • /
    • pp.252-252
    • /
    • 2022
  • 최근 기존의 전통적인 부유사채집기를 활용한 유사량 측정방법의 대안으로 지속적인 유량측정을 목적으로 활용되고 있는 횡방향 초음파도플러유속계의 초음파산란도를 활용하여 부유사농도를 측정에 대한 연구가 수행되고 있다. 하지만, 하천에서 횡방향 초음파도플러유속계를 활용한 유사량 측정 연구는 현장에서 측정된 자료를 기반으로 분석이 수행되기 때문에 통제된 유사 조건에서의 유사의 농도 및 입도분포에 대한 사전 정보를 바탕으로 정밀한 분석의 어려움이 있다. 이에 본 연구에서는 안동하천실험센터 순환수로에서 균일한 입도를 가지고 있는 황토를 활용하여 주입량에 따른 초음파산란도의 변화와 동시에 측정된 LISST(laser in-situ scattering and transmissometry)의 부유사 농도와 입도 자료를 활용하여 유사공급에 따른 3Mhz의 단일 주파수를 사용하는 횡방향 초음파도플러유속계의 초음파산란도를 활용한 부유사농도 및 평균입경의 측정에 대한 연구를 수행하였다. 측정결과, 유사공급에 따라서 초음파산란도를 활용하여 부유사농도가 증가하는 경향을 나타내고 있었으며, 횡방향 초음파도플러유속계로부터 측정된 셀별 초음파산란도를 활용하여 부유사농도의 공간적인 분포의 분석이 가능함을 확인할 수 있었다. 그리고 유사감쇄계수를 활용하여 LISST로부터 측정된 평균입경과의 관계식을 개발하여 유사공급에 따른 평균입경의 변화를 측정할 수 있을 것으로 나타났다. 추후, 통제된 조건에서의 다양한 입도분포에 따른 실험을 통해 횡방향 초음파도플러유속계의 초음파산란도를 활용한 부유사농도 측정 정확도를 개선과 유사감쇄계수를 통해 평균입경을 측정하는데 활용이 될 것으로 판단된다.

  • PDF

Doppler Ultrasound guided modified Allen's Test in Volkmann's Ischemic Contracture Patient with Ulnar Artery Constriction (도플러 초음파 감시하 알렌 검사로 진단한 상지 볼크만 구축 환자의 척골 동맥 협착 - 1례 보고 -)

  • Kim, Chul Hong;Ha, Dong Ho;Kim, Bo Kun;You, Sung Gon
    • The Journal of Korean Orthopaedic Ultrasound Society
    • /
    • v.4 no.2
    • /
    • pp.84-87
    • /
    • 2011
  • We had applied Doppler ultrasound guided modified Allen's test for a Volkmann's ischemic contracture patient who had ulnar artery constriction which was not detected with acoustic Doppler and CT-angiogram preoperatively. We report this case of Volkmann's ischemic contracture with brief review of literatures for awareness of the Doppler-ultrasound guided modified Allen's test.

  • PDF

Study of Doppler Fluid Effects of Carbonated Water in a Bi-directional Flow Phantom Model (양방향 흐름 팬텀 모델에서 탄산수의 도플러 유체 효과 연구)

  • Ji-Hye Kim;Yeong-Cheol Heo
    • Journal of the Korean Society of Radiology
    • /
    • v.18 no.2
    • /
    • pp.83-91
    • /
    • 2024
  • The purpose of this study was to determine the doppler fluid effects of carbonated water (CBW) in a bi-directional flow phantom model. A bi-directional flow phantom model was chosen to realize arterial and venous flow, and the structure of the inner and outer tanks allowed for fluid circulation and also made the size of the phantom small. Carbonated water (CBW), salt fluid (SAF), sugar fluid (SUF), and distilled water (DW) were used as fluids, and ultrasound scans were performed at depths of 1.5 cm and 3.0 cm from the surface of the tank, using B-mode and color Doppler effects. All fluids tested showed color Doppler effects, but CBW had the highest doppler shift and the least variation with depth. In conclusion, we determined that CBW was the most suitable fluid to be used as a doppler fluid and confirmed that the bubbles dissolved in CBW act as doppler scatterers, just like red blood cells inside human blood. Therefore, it is possible that CBW can be used as a blood-mimicking fluid in doppler ultrasound phantoms through further research, and this study will provide basic data.

Power Doppler Sonography in Children with Acute Pyelonephritis (소아 급성 신우신염에서 출력 도플러 초음파검사)

  • Lee Seung-Joo;Lee Sun-Wha
    • Childhood Kidney Diseases
    • /
    • v.6 no.2
    • /
    • pp.218-226
    • /
    • 2002
  • Purpose: The purpose of this study is to assess the ability of power Doppler sonography in the detection of acute pyelonephritis. Materials and Methods: We performed gray scale sonography, power Doppler sonography, and $^{Tc-99m}DMSA$ scintigraphy of the kidney in 80 patients with symptoms suggesting upper urinary tract infection. All imaging studies were obtained within 4 days. On $^{Tc-99m}DMSA$ scintigraphy, decreased radioactivity or photopenic lesions were considered indicative of acute pyelonephritis. Triangular areas of decreased perfusion identified on power Doppler sonography were considered as parenchymal lesions of acute pyelonephritis. The results of gray scale sonography and power Doppler sonography were retrospectively analysed and compared with those of $^{Tc-99m}DMSA$ scintigraphy which was given as the standards of reference. Results: For 40(85%) of the 47 patients with scintigraphy-proven acute pyelonephritis, power Doppler sonography diagnosed this condition on the correct side. The acute pyelonephritis which was not revealed by power Doppler sonography was observed in seven patients. Also, in three patients, false-positive indication of pyelonephritis was given by power Doppler sonography. Gray scale sonography showed positive findings in 23(49%) of 47 patients with positive findings on scintigraphy. Conclusion: Power Doppler sonography seems to be less sensitive than $^{Tc-99m}DMSA$ DMSA scintigraphy but significantly more sensitive than gray scale sonography for the detection of acute pyelonephritis in children. Power Doppler sonography shows promise as a noninvasive means of assessing renal cortical perfusion in children with clinically suspected acute pyelonephritis.

  • PDF

Development of Ultrasound Sector B-Scanner(III)-Pulsed Ultrasonic Doppler System- (초음파 섹터 B-스캐너의 개발(III)-초음파 펄스 도플러 장치-)

  • 백광렬;안영복
    • Journal of Biomedical Engineering Research
    • /
    • v.7 no.2
    • /
    • pp.139-146
    • /
    • 1986
  • Pulsed ultrasonic Doppler system is a useful diagnostic instrument to measure blood-flow-velocity, velocity profile, and volume-blood-flow. This system is more powerful compare with 2-dimensional B-scan tissue image. A system has been deve- loped and ii being evaluated using TMS 32010 DSP. We use this DSP for the purpose of real-time spectrum analyzer to obtain spectrogram in singlegate pulsed Doppler system and for the serial comb filter to cancel clutter and zero crossing counter to estimate Doppler mean frequency in multigate pulsed Doppler system. The Doppler shift of the backscattered signals is sensed in a phase detector. This Doppler signal corresponds to the mean velocity over a some region in space defined by the ultrasonic beam dimensions, transmitted pulse duration, and transducer ban(iwidth. Multi- gate pulsed Doppler system enable the transcutaneous and simultaneous assessment of the velocities in a number of adjacent sample volumes as a continuous function of time. A multigate pulsed Doppler system processing the information originating from presented.

  • PDF

Influence to the Doppler Images by the Defects of Piezoelectric Elements of the Probe of Medical Ultrasonic Scanners (Focusing on the Impact of an Increase in the Defects of Piezoelectric Elements) (초음파 프로브 소자 결함이 도플러 영상에 미치는 영향 (소자 결함 증가에 따른 영향을 중심으로))

  • Lee, Kyung-Sung
    • Journal of radiological science and technology
    • /
    • v.37 no.2
    • /
    • pp.117-124
    • /
    • 2014
  • This study has investigated the effects of the defects in the probe elements influence Doppler images in the medical ultrasonic scanners. This work was implemented that the quality of Doppler images depended on the extent and location of the probe element defected. The probe performance was rated in terms of the number of piezoelectric elements lost and this was studied in the experiment by electrical disconnection to the elements. The results showed that Doppler velocity became rapidly reduced as the defected elements encountered with the element group activated at the Doppler mode, not as the flow velocity. The effect of the probe defect is decreased when the defects occurred at the element group activated for Doppler mode, as was increased the number of the elements. It was observed that the higher the flow velocity of Doppler flow phantom is, the wider the spectrum of Doppler is. And the Doppler velocity soared up and the dispersion of image brightness is increased when the defected elements got out of the elements activated at Doppler mode. The result showed that TADVP(time-average-Doppler-velocity-profile) is decreased with the increase of the probe element defect, especilly in the region of high frequency. It is expected that the research of various defects of probe elements are needed, and this study can be practical tools for probe based ultrasonic QA in the future.

An implementation of the continuous wave doppler system for blood flow measurement using the ultrasound (효율적인 혈류 속도 측정을 위한 연속 초음파 도플러 장치의 구현)

  • 박형재;김영길
    • Proceedings of the Korean Institute of Information and Commucation Sciences Conference
    • /
    • 2001.05a
    • /
    • pp.516-519
    • /
    • 2001
  • To diagnose a patient's blood vessel disease, apoplexy, hypertension, arteriosclerosis, the blood velocity is very important. Determining the blood velocity methods using ultrasound are Continuous Doppler System and Pulse Doppler System. In using the Pulse Doppler System, we can obtain the position of blood velocity. But it is more complex hardware than Continuous Doppler System and it has low SNR(signal-noise ratio). So in this study, to obtain a believable information we use the Continuous Pulse Doppler System. Thus system have analog part and digital part. In analog part is composed of ultrasound generating part, the amplifying part to amplify the received signal from ultrasound sensor, the demodulation part to detect blood velocity and the filtering part to remove the noise. In digital part is composed of the A/D conversion part, digital signal processing part, and the communication part to communicate the PC. In this study to implement efficient ultrasound blood velocity measurement system, we can get the patient's blood velocity information in realtime. Thus, It is a useful in the accurate diagnosis with C.T(computered tomography), M.R.I(magnetic resonance imaging).

  • PDF