Tooth implant is located in oral cavity and affects neck, skull base, and facail image. These magnetic inhomogeneities are usually frequency encoding direction which cause artifacts due to change of signal strength and geometric distortion. First, to evaluate signal to noise ratio (SNR) of magnetic resonance image caused by tooth implant this study uses meat phantom which is similar to human body and is consisted with fat, muscle, and water to measure signal to noise ratio. Second, signal to noise ratio by using custom-made fixed phantom is measured, and then signal to noise ratio size of different tooth implant types is compared and analyzed. The measured signal to noise ratio values of Brushite, HSA, Metal, and RBM for meat phantom were 2.76, 2.22, 1.88, and 1.57 on T1 SE, 1.88, 1.78, 1.65, and 1.79 on T2 FLAIR, 2.28, 2.25, 2.88, and 2.05 on T2 FSE, and 2.74, 1.94, 1.67, and 1.48 on T2 GRE. The measured signal to noise ratio values of Brushite, HSA, Metal, and RBM for fixed water phantom were 1.2, 1.06, 1.12, and 1.22 on DWI, 1.93, 1.87, 1.93, and 2.06 T1 SE, 1.83, 1.76, 1.82, and 1.92 on T2 FLAIR, 1.85, 1.79, 7.86, and 1.97 on T2 FSE, and 1.97, 1.93, 1.99, and 2.06 on T2 GRE. By considering through the results, patients and dentists need to consider some impacts from testing many aspects although their main purpose of having tooth implants is a dental restoration. Moreover, depending on the tooth implant characteristics of individual patients this study results can be used as baseline data when choosing test protocol.