Abstract
In this paper, we introduced a near-infrared multi-channel fluorescence imaging system and analyzed the effects of measurements variables such as exposure time, working distance and intensity of excitation light. Fluorescence signal is increased as exposure time becomes longer, excitation light intensity increases or working distance becomes smaller. Furthermore, the proper composition of optical filters and precise packaging of the imaging modules prevent the increase of background signal. Thus, we confirmed an increase in SBR. Based on the result of this research, we proposed a method to use a multi-channel fluorescence imaging system.
Keywords
Fluorescence; Multi-channel imaging; Fluorescence Imaging; Real-time imaging;
Citations & Related Records
Times Cited By KSCI : 4  (Citation Analysis)
연도 인용수 순위
1 A. W. Yang, S. U. Cho, M. Y. Jeong, and H. S. Choi, "NIR Fluorescence Imaging Systems with Optical Packaging Technology", J. Microelectron. Packag. Soc., 21(4), 25 (2014).   DOI
2 A. M. De Grand, and J. V. Frangioni, "An operational nearinfrared fluorescence imaging system prototype for large animal surgery", Technol. Cancer. Res. Treat., 2(6), 553 (2003).   DOI
3 T. Kim, S. U. Cho, C. S. Park, H. Lee, D. Kim, and M. Y. Jeong, "A study on fluorescence imaging system characteristics depending on tilting of band pass filter", J. Microelectron. Packag. Soc., 23(2), 85 (2016).   DOI
4 T. Kim, S. U. Cho, C. S. Park, H. G. Lee, D. I. Kim, and M. Y. Jeong, "A study on fluorescence imaging system packaging and optical intensity characteristics", J. Microelectron. Packag. Soc., 23(3), 37 (2016).   DOI
5 T. Kim, C. O'Brien, H. S. Choi, and M. Y. Jeong, "Fluorescence molecular imaging systems for intraoperative imageguided surgery", Applied Spectroscopy Reviews, 53(2-4), 349 (2018).   DOI
6 J. G. Jeong, "General perspectives for molecular nuclear imaging", The Korean Society of Nuclear Medicine, 38(2), 111 (2004).
7 J. Choi, S. U. Cho, D. Kim, H. Lee, H. S. Choi, and M. Y. Jeong, "A Study on Characteristics Analysis of Multichannel Filter Module for Near-infrared Fluorescence Imaging", J. Microelectron. Packag. Soc., 23(1), 29 (2016).   DOI
8 J. H. Lee, G. Park, G. H. Hong, J. Choi, and H. S. Choi, "Design considerations for targeted optical contrast agents", Quant. Imaging. Med. Surg., 2(4), 266 (2012).
9 J. V. Frangioni, "In vivo near-infrared fluorescence imaging", Curr. Opin. Chem. Biol., 7(5), 626 (2003).   DOI