Evaluation of Resolution Characteristics by Using Chart Device Angle

차트 각도를 이용한 해상력 특성 평가

  • Min, Jung-Whan (Department of Radiological technology, Shingu University) ;
  • Jeong, Hoi-Woun (Department of Radiological Science, Baekseok Culture University)
  • 민정환 (신구대학교 방사선학과) ;
  • 정회원 (백석문화대학교 방사선학과)
  • Received : 2021.06.15
  • Accepted : 2021.08.02
  • Published : 2021.08.31

Abstract

This study aim was quantitative assessment of MTFs of spectrum of the square wave chart images and Coltman chart images for 0°, 1.7°, 2.2°, 2.9°, 4.1° by using chart method. In general device was AccuRay-650 (DK Medical System, Korea) used, indirect flat panel detector(FPD) Aero (Konica, Japan) used and MATLAB R2019a (MathWorks, USA) used. The result of comparison for each angle of MTF the edge image was highest quantitatively value for MTF finding of showed the best value of 0.1 based on the frequency of 3.5 mm-1, value of 0.1 based on the square wave was frequency of 3.0 mm-1 and value of 0.1 based on the Coltman transform was frequency of 2.4 mm-1. In this study it was significant that the methodology of the international Electro-technical Commission was applied mutandis by using the Fujita method within 2~3°.

Keywords

Acknowledgement

This study was supported by from the Shingu University Industry-Academic Cooperation Foundation Grant 2021.

References

  1. Min JW, Jeong HW, Kim KW, et al. Study on the resolution characteristics by using magnetic resonance imaging 3.0T. Journal of Radiological Science and Technology. 2020;43(4):251-57. https://doi.org/10.17946/JRST.2020.43.4.251
  2. Jeong HW, Min JW, Kim JM, et al. Performance characteristic of a CsI(Tl) flat panel detector radiography system. Journal of Radiological Science and Technology. 2012;35(2):109-17.
  3. Miyati T, Fujita H, Kasuga T, et al. Measurements of MTF and SNR(f) using a subtraction method im MRI. Physics in Medicine and Biology. 2002;47:2961-72. https://doi.org/10.1088/0031-9155/47/16/311
  4. Min JW, Jeong HW, Kim KW, et al. Evaluation of image quality for various electronic portal imaging devices in radiation therapy. Journal of Radiological Science and Technology. 2015;38(4):451-61. https://doi.org/10.17946/JRST.2015.38.4.16
  5. Kim KW, Jeong HW, Min JW, et al. Measurement of image quality according to the time of computed radiography system. Journal of Radiological Science and Technology. 2015;38(4):365-74. https://doi.org/10.17946/JRST.2015.38.4.05
  6. Kim KW, Jeong HW, Min JW, et al. Evaluation of the performance characteristic for mammography by using edge device. Journal of Radiological Science and Technology. 2016;39(3):415-20. https://doi.org/10.17946/JRST.2016.39.3.16
  7. Kim KW, Jeong HW, Min JW, et al. Evaluation of the modulation transfer function for computed tomography by using american association physics medicine phantom. Journal of Radiological Science and Technology. 2016;39(2):193-8. https://doi.org/10.17946/JRST.2016.39.2.09
  8. Sawant A, Antonuk LE, El-Mohri Y. Slit design for efficient and accurate MTF measurement at megavoltage X-ray energies. Medical Physics. 2007;34:1535-45. https://doi.org/10.1118/1.2717405
  9. Kim JM, Jeong HW, Min JW, et al. Correction method of slit modulation transfer function on digital medical imaging system. Journal of Radiological Science and Technology. 2006;29(3):133-9.
  10. Min JW, Jeong HW. Comparison of modulation transfer function in measurements by using edge device angle in indirect digital radiography. Journal of Radiological Science and Technology. 2019;42(4):259-63. https://doi.org/10.17946/JRST.2019.42.4.259
  11. Min JW, Jeong HW. Evaluation of angle optimization on edge test device setting in modulation transfer function. Journal of Radiological Science and Technology. 2020;43(1):15-21. https://doi.org/10.17946/JRST.2020.43.1.15
  12. Ichikawa K, et al. New method of measuring presampling MTF by fourier transform of square wave chart image. Japanese Journal of Radiological Technology. 2002;58(9):1261-7. https://doi.org/10.6009/jjrt.KJ00001364806
  13. Kunitomo H. Modulation transfer function-presampled MTF measurements by fourier analysis using square wave chart. Japanese Journal of Radiological Technology. 2009;65(11):1561-7. https://doi.org/10.6009/jjrt.65.1561
  14. Greer PB, van Doorn T. Evaluation of an algorithm for the assessment of the MTF using an edge method. Medical Physics. 2000;27:2048. https://doi.org/10.1118/1.1288682
  15. Fujita H, Tasai DY, Itoh T, et al. A simple method for determining the modulation transfer function in digital radiography. IEEE Trans Med Imaging. 1992;11(1):34-9. https://doi.org/10.1109/42.126908
  16. Samei E, Flynn MJ, Reimann DA, et al. A method for measuring the presampled MTF of digital radiographic systems using an edge test device. Medical Physics. 1998;25:102. https://doi.org/10.1118/1.598165
  17. IEC (International Electro-technical Commission) 62220-1. Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency. Geneva; 2003.
  18. Min JW, Jeong HW, Kim SC. Evaluation of noise power spectrum characteristics by using magnetic resonance imaging 3.0T. Journal of Radiological Science and Technology. 2021;44(1):279-88.