Abstract

In this study, for the measurement of 3H(tritium) radioactivity concentration, a study was conducted on whether the type of cocktail and the material of the vial had an effect on the measurement before liquid scintillation counter measurement on HTO-type samples that had undergone physical and chemical pretreatment. As a result of the study, the efficiency according to the type of cocktail was higher in Ultima Gold LLT than Ultima Flo-AP cocktail with polyethylene (1.49%), glass (5.10%), and teflon (6.58%), respectively. Regarding the effect according to the type of vial, the efficiency and SQP(E) of both Ultima Gold LLT and Ultima Flo-AP showed the highest values in the order of teflon, polyethylene, and glass.

Keywords

References

  1. Yeom YS. An evaluation on uncertainty in LSC calibration and verification [dissertation]. Gwangju: University of Chosun; 2011.
  2. Han SJ, Lee KJ, Yeom JM, Shin DY. A study and analysis on tritium radioactivity and environmental behavior in domestic NPPs. Journal of Radiation Protection and Research. 2015;40(4):267-76. https://doi.org/10.14407/jrp.2015.40.4.267
  3. KRISS. Liquid scintillation method. 1997.
  4. HASL-300 3H-04-RC. Tritium in water-Liquid scintillation counting. 1997.
  5. Wallac Company. 1220 quantulus liquid scintillation counter, instrument manual. 1995.
  6. Packard Instrument Company. Laboratory manual for liquid scintillation counting. 1988.
  7. PerkinElmer. Quantulus 1220 liquid scintillation counter instrument manual. 2004.
  8. L'Annunziata MF. Handbook of radioactivity analysis. 2nd ed. Academic Press; 2003.
  9. Lee GJ, Han SJ, Cho YJ. Development on the measurement procedures of the C-14 activity of the nuclear power plants worker's urine. Chosun University; 2008.
  10. Han SJ, Lee KJ, Lee SJ, Kim HG, Park ES. Effect evaluation by activity and geometry difference in calibration on LSC. Journal of Radiation Protection and Research. 2008;33(1):21-6.
  11. KINS. Manual of standard procedures for environmental sampling and radioactivity analysis. Korea Institute of Nuclear Safety, KINS/G-009.
  12. KINS. Development of procedure manual for the radionuclides analysis. 2013.
  13. A Canberra Company. Liquid scintillation analysis. Science and Technology; 1998.
  14. PerkinElmer. Cocktails for liquid scintillation counting. 1997.
  15. Jo JG, Lee SB, Nam JY, Noh EJ, Beak HW, Lee YJ, et al. Evaluation of radioactive substance and measurement of harmfulness in drinking water. Journal of Radiological Science and Technology. 2021;44(3):247-52. https://doi.org/10.17946/JRST.2021.44.3.247
  16. Jang ES, Gim YS, Lee SY. Marinelli beaker measurement and self absorption correction and application for various environmental samples in montecarlo simulation. Journal of Radiological Science and Technology. 2017;40(4):605-11. https://doi.org/10.17946/JRST.2017.40.4.10
  17. KETEP. A research on development and correction of standard source of liquid scintillation counter for analyzing environment sample. R-2005-0-005; 2007.