Background reduction by Cu/Pb shielding and efficiency study of NaI(TI) detector

  • Ramadhan, Revink A. (Department of Physics, College of Science, University of Duhok) ;
  • Abdullah, Khairi MS. (Department of Water Resources, College of Engineering, University of Duhok)
  • Received : 2017.10.07
  • Accepted : 2017.12.28
  • Published : 2018.04.25

Abstract

The background spectrum of a $3^{{\prime}{\prime}}{\times}3^{{\prime}{\prime}}$ NaI(Tl) well-type scintillation SILENA detector was measured without shielding, in 6 cm thick lead shielding, and with 2 mm thick electrolytic copper covering the detector inside the lead shielding. The relative remaining background of the lead shield lined with copper was found to be ideal for low-level environmental radioactive spectroscopy. The background total count rate in the (20-2160 KeV) was reduced 28.7 times by the lead and 29 times by the Cu + Pb shielding. The effective reduction of background (1.04) by the copper mainly appeared in the energy range from X-ray up to 500 KeV, while for the total energy range the ratio is 1.01 relative to the lead only. In addition, a strong relation between the full-energy peak absolute efficiency and the detector well height was found using gamma-ray isotropic radiation point sources placed inside the detector well. The full-energy peak efficiency at a midpoint of the well (at 2.5 cm) is three times greater than that on the detector surface. The energy calibrations and the resolution of any single energy line are independent of the locations of the gamma source inside or outside of the well.

Keywords

References

  1. I. Akkurt, K. Gunoglu, S.S. Arda, Detection Efficiency of NaI ( Tl ) Detector in 511 - 1332 keV Energy Range, Sci. Technol. Nucl. Install. 2014 (2014) 6-11.
  2. D.S. Mrd, I.S. Bikit, I.V. Anicin, J.M. Slivka, J.J. Hansman, N.M. Zikic-Todorovic, E.Z. Varga, S.M. Curcic, J.M. Puzovi, Background reduction using Fe and Pb shielding, J. Res. Phys. 30 (2004) 49-54.
  3. J. Hansman, Design and construction of a shield for the $900^{{\prime}{\prime}}\;{\times}\;900^{{\prime}{\prime}}$ NaI(Tl) welltype detector, Nucl. Technol. Radiat. Prot. 29 (2014) 165-169. https://doi.org/10.2298/NTRP1402165H
  4. A. Jehouani, R. Ichaoui, M. Boulkheir, Study of the NaI(Tl) efficiency by Monte Carlo method, Appl. Radiat. Isot. 53 (2000) 887-891. https://doi.org/10.1016/S0969-8043(00)00254-2
  5. S. Sen, A. Chandra, D. Pramanik, M.S. Sarkar, Study of Intrinsic Photo peak Efficiency of NaI ( Tl ) Detectors at 662 KeV, Proceeding DAE Symo, Nucl. Phys. 58 (2013) 934-935.
  6. M.S. Badawi, M. Abd-Elzaher, A.A. Thabet, A.M. El-Khatib, An empirical formula to calculate the full energy peak efficiency of scintillation detectors, Appl. Radiat. Isot. 74 (2013) 46-49. https://doi.org/10.1016/j.apradiso.2012.12.011
  7. J. Hansman, D. Mrdja, J. Slivka, M. Krmar, I. Bikit, Efficiency study of a big volume well type NaI(Tl) detector by point and voluminous sources and Monte-Carlo simulation, Appl. Radiat. Isot. 99 (2015) 150-154. https://doi.org/10.1016/j.apradiso.2015.03.003
  8. M.I. Abbas, Analytical formulae for well-type NaI (Tl) and HPGe detectors efficiency computation, Appl. Radiat. Isot. 55 (2001) 245-252. https://doi.org/10.1016/S0969-8043(01)00047-1
  9. S. Yalcin, O. Gurler, G. Kaynak, O. Gundogdu, Calculation of total counting efficiency of a NaI(Tl) detector by hybrid Monte-Carlo method for point and disk sources, Appl. Radiat. Isot. 65 (2007) 1179-1186. https://doi.org/10.1016/j.apradiso.2007.05.003
  10. G.A. Brinkman, A.H.W. Aten, J.T. Veenboer, Absolute Standardization with a NaI(Tl) Crystal - I Calibration by Means of a Single Nuclide, Int. J. Appl. Radiat. Isot. 14 (1963) 153-157. https://doi.org/10.1016/0020-708X(63)90110-8
  11. F. Cejnar, I. Kovar, Experimental and calculated photopeak efficiency values for well-type NaI(Tl) detectors, Int. J. Appl. Radiat. Isot. 31 (1980) 79-84. https://doi.org/10.1016/0020-708X(80)90047-2
  12. Silena international S.P.A web page, 2016. http://www.geocities.ws/silenaspa/silena/profile.htm.
  13. K.M.S. Abdullah, R.A. Ramadhan, Determination of Gamma-Emitting Radionuclides in Duhok City, Iraq, Nuclear 48 (2011) 295-300.
  14. D. Demir, A. Un, Y. Sahin, Efficiency determination for NaI (Tl) detectors in the 23 keV to 1333 keV energy range, Instrum. Sci. Technol. 36 (2008) 291-301. https://doi.org/10.1080/10739140801944092
  15. G. Gilmore, Practical Gamma-ray Spectrometry, second ed., John Wiley & Sons Ltd, 2008.
  16. K.M.S. Abdullah, A Method to Determine Doublet Photopeak Area $^{137}Cs\;and\;^{214}Bi$, in NORM Soil by NaI-detector, Int, J. Fundam. Phys. Sci 5 (2015) 32-37.
  17. G.F. Knoll, Radiation Detection and Measurement, third ed., John Wiley & Sons, Inc., 2000.

Cited by

  1. Short-term exposures to particulate matter gamma radiation activities and biomarkers of systemic inflammation and endothelial activation in COPD patients vol.180, pp.None, 2020, https://doi.org/10.1016/j.envres.2019.108841
  2. Shielding design for high-intensity Co-60 and Ir-192 gamma sources used in industrial radiography based on PHITS Monte Carlo simulations vol.135, pp.10, 2018, https://doi.org/10.1140/epjp/s13360-020-00797-8
  3. Determination of Radium and Radon Exhalation Rate as a Function of Soil Depth of Duhok Province - Iraq vol.14, pp.1, 2018, https://doi.org/10.1080/16878507.2021.1999719
  4. Effects of particulate matter gamma radiation on oxidative stress biomarkers in COPD patients vol.31, pp.4, 2021, https://doi.org/10.1038/s41370-020-0204-8