A Copper Shield for the Reduction of X-γ True Coincidence Summing in Gamma-ray Spectrometry

  • Received : 2018.09.03
  • Accepted : 2018.11.30
  • Published : 2018.12.31

Abstract

Background: Gamma-ray detectors having a thin window of a material with low atomic number can increase the true coincidence summing effects for radionuclides emitting X-rays or gamma-rays. This effect can make efficiency calibration or spectrum analysis more complicated. In this study, a Cu shield was tested as an X-ray filter to neglect the true coincidence summing effect by X-rays and gamma-rays in gamma-ray spectrometry, in order to simplify gamma-ray energy spectrum analysis. Materials and Methods: A Cu shield was designed and applied to an n-type high-purity germanium detector having an $X-{\gamma}$ summing effect during efficiency calibration. This was tested using a commercial, certified mixed gamma-ray source. The feasibility of a Cu shield was evaluated by comparing efficiency calibration results with and without the shield. Results and Discussion: In this study, the thickness of a Cu shield needed to avoid true coincidence summing effects due to $X-{\gamma}$ was tested and determined to be 1 mm, considering the detection efficiency desired for higher energy. As a result, the accuracy of the detection efficiency calibration was improved by more than 13% by reducing $X-{\gamma}$ summing. Conclusion: The $X-{\gamma}$ summing effect should be considered, along with ${\gamma}-{\gamma}$ summing, when a detection efficiency calibration is implemented and appropriate shielding material can be useful for simplifying analysis of the gamma-ray energy spectra.

Keywords

References

  1. Debertin K, Helmer RG. Gamma- and X-ray spectrometry with semiconductor detectors. 3rd Ed. Amsterdam, Netherlands. North-Holland. 1988;258-270.
  2. Montgomery DM, Montgomery GA. A method for assessing and correcting coincidence summing effects for germanium detector efficiency calibrations. J. Radioanal. Nucl. Chem. 1995;193(5):71-79. https://doi.org/10.1007/BF02041918
  3. Sima O, Arnold D. Self-attenuation and coincidence-summing corrections calculated by Monte Carlo simulations for gammaspectrometric measurement with well-type germanium detectors. Appl. Radiat. Isot. 1996;47(9-10):889-893. https://doi.org/10.1016/S0969-8043(96)00082-6
  4. Garcia-Torano E, Pozuelo M, Salvat F. Monte Carlo calculations of coincidence-sum corrections for volume sources in gammaray spectrometry with Ge detectors. Nucl. Instrum. Methods Phys. Res. Sect. A. 2005;554(3):577-583.
  5. Venkataraman R, Croft S, Russ WR. Calculation of peak-to-total ratios for high purity germanium detectors using Monte-Carlo modeling. J. Radioanal. Nucl. Chem. 2005;264(4):183-191. https://doi.org/10.1007/s10967-005-0692-y

Cited by

  1. A shipboard real-time gamma-ray measurement system for detecting radionuclides in seawater vol.1005, pp.None, 2018, https://doi.org/10.1016/j.nima.2021.165374