A Copper Shield for the Reduction of X-γ True Coincidence Summing in Gamma-ray Spectrometry  

Byun, Jong-In (Korea Institute of Nuclear Safety)
Publication Information
Journal of Radiation Protection and Research / v.43, no.4, 2018 , pp. 137-142 More about this Journal
Abstract
Background: Gamma-ray detectors having a thin window of a material with low atomic number can increase the true coincidence summing effects for radionuclides emitting X-rays or gamma-rays. This effect can make efficiency calibration or spectrum analysis more complicated. In this study, a Cu shield was tested as an X-ray filter to neglect the true coincidence summing effect by X-rays and gamma-rays in gamma-ray spectrometry, in order to simplify gamma-ray energy spectrum analysis. Materials and Methods: A Cu shield was designed and applied to an n-type high-purity germanium detector having an $X-{\gamma}$ summing effect during efficiency calibration. This was tested using a commercial, certified mixed gamma-ray source. The feasibility of a Cu shield was evaluated by comparing efficiency calibration results with and without the shield. Results and Discussion: In this study, the thickness of a Cu shield needed to avoid true coincidence summing effects due to $X-{\gamma}$ was tested and determined to be 1 mm, considering the detection efficiency desired for higher energy. As a result, the accuracy of the detection efficiency calibration was improved by more than 13% by reducing $X-{\gamma}$ summing. Conclusion: The $X-{\gamma}$ summing effect should be considered, along with ${\gamma}-{\gamma}$ summing, when a detection efficiency calibration is implemented and appropriate shielding material can be useful for simplifying analysis of the gamma-ray energy spectra.
Keywords
True coincidence summing; X-ray; Gamma-ray spectrometry;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Debertin K, Helmer RG. Gamma- and X-ray spectrometry with semiconductor detectors. 3rd Ed. Amsterdam, Netherlands. North-Holland. 1988;258-270.
2 Montgomery DM, Montgomery GA. A method for assessing and correcting coincidence summing effects for germanium detector efficiency calibrations. J. Radioanal. Nucl. Chem. 1995;193(5):71-79.   DOI
3 Sima O, Arnold D. Self-attenuation and coincidence-summing corrections calculated by Monte Carlo simulations for gammaspectrometric measurement with well-type germanium detectors. Appl. Radiat. Isot. 1996;47(9-10):889-893.   DOI
4 Garcia-Torano E, Pozuelo M, Salvat F. Monte Carlo calculations of coincidence-sum corrections for volume sources in gammaray spectrometry with Ge detectors. Nucl. Instrum. Methods Phys. Res. Sect. A. 2005;554(3):577-583.
5 Venkataraman R, Croft S, Russ WR. Calculation of peak-to-total ratios for high purity germanium detectors using Monte-Carlo modeling. J. Radioanal. Nucl. Chem. 2005;264(4):183-191.   DOI