초록
We reports improved monitoring performance of Self plasma-optical emission spectroscopy (SP-OES) by augmenting a by-pass tube to a conventional straight (or single) tube type self plasma reactor. SP-OES has been used as a tool for the monitoring of plasma chemistry indirectly in plasma process system. The benefits of SP-OES are low cost and easy installation, but some semiconductor industries who adopted commercialized SP-OES product experiencing less sensitivity and slow sensor response. OH out-gas chemistry monitoring was performed to have a direct comparison of a conventional single type tube and a by-pass type tube, and fluid dynamic simulation on the improved hardware design was also followed. It is observed faster pumping out of OH from the chamber in the by-pass type SP-OES.