Measurement of Spatial Scattered Dose Distribution According to Presence or Absence of Radiation Shielding in the Operating Room

수술실에서 방사선 차폐기 사용 유무에 따른 공간산란선량분포의 측정

  • Do, Sang-Lock (Department of Radiological Science, Daegu Catholic University) ;
  • Cho, Pyong-Kon (Department of Radiological Science, Daegu Catholic University) ;
  • Kim, Seong-Jin (Korea Institute of Medical Device Assessment) ;
  • Jung, Dong Kyung (Department of Radioation Oncology, Daegu Fatima Hospital)
  • 도상록 (대구가톨릭대학교 방사선학과) ;
  • 조평곤 (대구가톨릭대학교 방사선학과) ;
  • 김성진 (한국의료기기평가연구원) ;
  • 정동경 (대구파티마병원 방사선 종양학과)
  • Received : 2017.11.03
  • Accepted : 2017.12.24
  • Published : 2017.12.31

Abstract

This study compared the spatial scattered dose distribution according to whether the recently developed radiation shielding is used or not in order to understand the spatial scattered dose distribution of C-arm. The horizontal side distribution increased by $30^{\circ}$ in the interval of the radius 50 cm on the height of 95 cm based on the head of the patient, and it was measured by increasing $30^{\circ}$ with the interval of 50 cm in the vertical side of each horizontal side. In the same method, the radiation shielding was installed and measured. The result of measurement shows that the horizontal side of 50 cm distance was $0^{\circ}$, $90^{\circ}$ and $180^{\circ}$, was $1.77{\pm}0.12$, $1.90{\pm}0.13$, $2.12{\pm}0.14$, and $2.69{\pm}0.15mSv/h$ in the $270^{\circ}$ direction, and was $1.59{\pm}0.12$, $0.99{\pm}0.09$, $1.47{\pm}0.11$, and $1.37{\pm}0.11mSv/h$ after the use of the radiation shielding. In addition, the vertical distribution in horizontal direction $90^{\circ}$ with 50 cm distance was $30^{\circ}$, $60^{\circ}$, $120^{\circ}$, was $3.85{\pm}0.18$, $9.15{\pm}0.28$, $10.82{\pm}0.31$, and $5.40{\pm}0.22mSv/h$ in $150^{\circ}$, and was $2.03{\pm}0.13$, $4.32{\pm}0.19$, $2.76{\pm}0.16$, and $1.92{\pm}0.13mSv/h\;mR/h$ after the use of the radiation shielding. Both direction showed decrease according to the use of the radiation shielding. Therefore, radiation related workers who work in operating rooms should recognize the spatial scattered dose distribution exactly and need to try to prevent the risk of radiation exposure with proper protective measures.

본 연구는 C-arm의 공간산란선량분포를 파악해 보고자, 최근에 개발된 방사선 차폐기의 사용 유무에 따른 공간산란선량분포를 비교하였다. 수평면 분포는 95 cm 높이에서 환자 머리를 기준으로 반경 50 cm 간격으로 $30^{\circ}$씩 증가하였고, 각 수평면의 수직면 내에서 반경 50 cm 간격으로 $30^{\circ}$씩 증가하며 측정하였다. 동일한 방법으로 방사선 차폐기를 설치 후 측정하였다. 측정결과 50 cm 거리 수평면은 $0^{\circ}$, $90^{\circ}$, $180^{\circ}$ 그리고 $270^{\circ}$ 방향에서 $1.77{\pm}0.12$, $1.90{\pm}0.13$, $2.12{\pm}0.14$, 그리고 $2.69{\pm}0.15mSv/h$, 방사선 차폐기 사용 후 $1.59{\pm}0.12$, $0.99{\pm}0.09$, $1.47{\pm}0.11$, 그리고 $1.37{\pm}0.11mSv/h$로 나타났다. 또한, 50 cm 거리 수평 방향 $90^{\circ}$의 수직 분포는 $30^{\circ}$, $60^{\circ}$, $120^{\circ}$ 그리고 $150^{\circ}$에서 $3.85{\pm}0.18$, $9.15{\pm}0.28$, $10.82{\pm}0.31$, 그리고 $5.40{\pm}0.22mSv/h$, 방사선 차폐기 사용 후 각 각 $2.03{\pm}0.13$, $4.32{\pm}0.19$, $2.76{\pm}0.16$, 그리고 $1.92{\pm}0.13mSv/h$이었다. 방사선 차폐기 사용에 따라 각 방향 모두에서 감소하였다. 따라서 수술실에 근무하는 방사선관계종사자는 공간산란선량 분포를 정확히 인지하고 적절한 방호 대책으로 피폭의 위해를 방지하는 노력이 필요하다고 사료된다.

Keywords

References

  1. Lee JG. Principles of Radiation Protection. Korea Association for Radiation Application. Geoseong Publishing Printing; 2016, 541-93.
  2. Korea Centers for Disease Control and Prevention. Publication of Medical Radiation Occupational Radiation Statistics Statistical Yearbook. 2017 [cited 2017 October 10]; https://www.cdc.go.kr/CDC/contents/CdcKrContentLink.jsp?fid=21&cid=72953&ctype=1.
  3. Shim DM, Kim YM, Oh SK, Lim CM, Kown BT. Radiation Induced Hand Necrosis of an Orthopaedic Surgeon Who Had Treated a Patient with Fluoroscopy-Guided Spine Injection. J Korean Orthop Assoc. 2014;49(3):250-4. https://doi.org/10.4055/jkoa.2014.49.3.250
  4. Kim JW, Kim JJ. Radiation Exposure to the Orthopaedic Surgeon during Fracture Surgery. J Korea Orthop Assoc. 2010;45:107-13. https://doi.org/10.4055/jkoa.2010.45.2.107
  5. Park SH, Park JM, Kim HS. The Study on Scattered Ray by C-arm in Operation Room. Korean journal of digital imaging in medicine. 2011;13: 21-6.
  6. Cho PK. Distribution of the Scatter ray in Fluoroscopy X-ray Room. The Journal of the Korea Contents Association. 2011;11(10):349-54. https://doi.org/10.5392/JKCA.2011.11.10.349
  7. Kown DM, Park MH, Nam HD. Measurement of the Scattered Spatial Dose Distribution for the Mobile X-ray Radiography. Journal of Korean Society of radiological technology. 2001;24(1): 23-6.
  8. Sung KH, Jung YJ, Kwon SS, Lee GW, Chung CY, Lee KM, Cha HM, et al. Performances of a protector against scattered radiation during intraoperative use of a C-arm fluoroscope. Journal of Radiological Protection. 2016;36:629-40. https://doi.org/10.1088/0952-4746/36/3/629
  9. Ministry of Health and Welfare. Rules om Safty Control for Diagnostic Radiographic Unit. 2017 [cited 2017 October 10]; http://www.law.go.kr/lsSc.do?menuId=0&subMenu=1&query=%EC%A7%84%EB%8B%A8%EC%9A%A9%EB%B0%A9%EC%82%AC%EC%84%A0#undefined.
  10. Kim BH, Kim HJ. A Study on knowledge, perception, self-efficacy, and performance on radiation protection among perioperative workers in terms of radiation protection. Journal of the Korea Academia-Industrial cooperation Society. 2017;18(5):343-54. https://doi.org/10.5762/KAIS.2017.18.5.343
  11. Lee JG. Principles of Radiation Protection. Korea Association for Radiation Application. Geoseong Publishing Printing; 2016, 381-431.
  12. Choi SK. A Study of Changes in the Primary Dose Penetrating the Protective Apron on SID in X-ray Radiography. Journal of Radiological Science and Technology. 2016;39(4):501-7. https://doi.org/10.17946/JRST.2016.39.4.03
  13. Cho PK. Distribution of the Scatter Ray on Chest X-ray Examinations. The Joumal of the Korea Contents Association. 2012;12(7):255-60.
  14. Kim KS, Song JN, Kim SO. Study on the Method for Reducing the Operator’s Exposure Dose From a C-Arm System. Journal of Radiological Science and Technology. 2016;39(4):493-9. https://doi.org/10.17946/JRST.2016.39.4.02
  15. Choi SG. A Study of Changes in the Primary Dose Penetrating the Protective Apron on SID in X-ray Radiography. Journal of Radiological Science and Technology. 2016;39(4):501-7. https://doi.org/10.17946/JRST.2016.39.4.03
  16. Choi SK. Relationship between the Distribution of Space doses in X-ray Rooms and the "Inverse Square Law of Distance". The Journal of the Korea Contents Association. 2013;13(8):301-7. https://doi.org/10.5392/JKCA.2013.13.08.301