Accuracy evaluation of metal copings fabricated by computer-aided milling and direct metal laser sintering systems

  • Park, Jong-Kyoung (Department of Dental Laboratory Science and Engineering, College of Health Science, Korea University) ;
  • Lee, Wan-Sun (Department of Dental Laboratory Science and Engineering, College of Health Science, Korea University) ;
  • Kim, Hae-Young (Department of Public Health Science, Graduate School & BK21+ Program in Public Health Science, Korea University) ;
  • Kim, Woong-Chul (Department of Dental Laboratory Science and Engineering, College of Health Science, Korea University) ;
  • Kim, Ji-Hwan (Department of Dental Laboratory Science and Engineering, College of Health Science, Korea University)
  • Received : 2014.06.30
  • Accepted : 2014.10.01
  • Published : 2015.04.30

Abstract

PURPOSE. To assess the marginal and internal gaps of the copings fabricated by computer-aided milling and direct metal laser sintering (DMLS) systems in comparison to casting method. MATERIALS AND METHODS. Ten metal copings were fabricated by casting, computer-aided milling, and DMLS. Seven mesiodistal and labiolingual positions were then measured, and each of these were divided into the categories; marginal gap (MG), cervical gap (CG), axial wall at internal gap (AG), and incisal edge at internal gap (IG). Evaluation was performed by a silicone replica technique. A digital microscope was used for measurement of silicone layer. Statistical analyses included one-way and repeated measure ANOVA to test the difference between the fabrication methods and categories of measured points (${\alpha}$=.05), respectively. RESULTS. The mean gap differed significantly with fabrication methods (P<.001). Casting produced the narrowest gap in each of the four measured positions, whereas CG, AG, and IG proved narrower in computer-aided milling than in DMLS. Thus, with the exception of MG, all positions exhibited a significant difference between computer-aided milling and DMLS (P<.05). CONCLUSION. Although the gap was found to vary with fabrication methods, the marginal and internal gaps of the copings fabricated by computer-aided milling and DMLS fell within the range of clinical acceptance (< $120{\mu}m$). However, the statistically significant difference to conventional casting indicates that the gaps in computer-aided milling and DMLS fabricated restorations still need to be further reduced.

Keywords

References

  1. Brecker SC. Porcelain baked to gold-A new medium in prosthodontics. J Prosthet Dent 1956;6:801-10. https://doi.org/10.1016/0022-3913(56)90077-4
  2. Johnston JF, Dykema RW, Cunningham DM. The use and construction of gold crowns with a fused porcelain veneer-A progress report. J Prosthet Dent 1956;6:811-21. https://doi.org/10.1016/0022-3913(56)90078-6
  3. Christensen GJ. Longevity of posterior tooth dental restorations. J Am Dent Assoc 2005;136:201-3. https://doi.org/10.14219/jada.archive.2005.0142
  4. Miyazaki T, Hotta Y, Kunii J, Kuriyama S, Tamaki Y. A review of dental CAD/CAM: current status and future perspectives from 20 years of experience. Dent Mater J 2009;28:44-56. https://doi.org/10.4012/dmj.28.44
  5. Weiss LE. Processes overview. In: Holdridge GM, Prinz FB, eds. Rapid Prototyping in Europe and Japan. Baltimore; Loyola College of Maryland; 1997. p. 5-20.
  6. Petzold R, Zeilhofer HF, Kalender WA. Rapid protyping technology in medicine-basics and applications. Comput Med Imaging Graph 1999;23:277-84. https://doi.org/10.1016/S0895-6111(99)00025-7
  7. Yadroitsev I, Bertrand Ph, Smurov I. Parametric analysis of the selective laser melting process. Appl Surf Sci 2007;253:8064-9. https://doi.org/10.1016/j.apsusc.2007.02.088
  8. Gardner FM. Margins of complete crowns-Literature review. J Prosthet Dent 1982;48:396-400. https://doi.org/10.1016/0022-3913(82)90072-5
  9. Wataha JC. Alloys for prosthodontic restorations. J Prosthet Dent 2002;87:351-63. https://doi.org/10.1067/mpr.2002.123817
  10. Wataha JC, Messer RL. Casting alloys. Dent Clin North Am 2004;48:vii-viii, 499-512.
  11. Duncan JD. The casting accuracy of nickel-chromium alloys for fixed prostheses. J Prosthet Dent 1982;47:63-8. https://doi.org/10.1016/0022-3913(82)90244-X
  12. Ucar Y, Akova T, Akyil MS, Brantley WA. Internal fit evaluation of crowns prepared using a new dental crown fabrication technique: laser-sintered Co-Cr crowns. J Prosthet Dent 2009;102:253-9. https://doi.org/10.1016/S0022-3913(09)60165-7
  13. Regish KM, Sharma D, Prithviraj DR, Nair A, Raghavan R. Evaluation and comparison of the internal fit and marginal accuracy of base metal (nickelchromium) and zirconia copings before and after ceramic veneering: a sem study. Eur J Prosthodont Restor Dent 2013;21:44-8.
  14. Eames WB, O'Neal SJ, Monteiro J, Miller C, Roan JD Jr, Cohen KS. Techniques to improve the seating of castings. J Am Dent Assoc 1978;96:432-7. https://doi.org/10.14219/jada.archive.1978.0090
  15. Holmes JR, Bayne SC, Holland GA, Sulik WD. Considerations in measurement of marginal fit. J Prosthet Dent 1989;62:405-8. https://doi.org/10.1016/0022-3913(89)90170-4
  16. van Noort R. The future of dental devices is digital. Dent Mater 2012;28:3-12. https://doi.org/10.1016/j.dental.2011.10.014
  17. Beuer F, Schweiger J, Edelhoff D. Digital dentistry: an overview of recent developments for CAD/CAM generated restorations. Br Dent J 2008;204:505-11. https://doi.org/10.1038/sj.bdj.2008.350
  18. Ortorp A, Jönsson D, Mouhsen A, Vult von Steyern P. The fit of cobalt-chromium three-unit fixed dental prostheses fabricated with four different techniques: a comparative in vitro study. Dent Mater 2011;27:356-63. https://doi.org/10.1016/j.dental.2010.11.015
  19. Quante K, Ludwig K, Kern M. Marginal and internal fit of metal-ceramic crowns fabricated with a new laser melting technology. Dent Mater 2008;24:1311-5. https://doi.org/10.1016/j.dental.2008.02.011
  20. Beuer F, Aggstaller H, Edelhoff D, Gernet W, Sorensen J. Marginal and internal fits of fixed dental prostheses zirconia retainers. Dent Mater 2009;25:94-102. https://doi.org/10.1016/j.dental.2008.04.018
  21. Laurent M, Scheer P, Dejou J, Laborde G. Clinical evaluation of the marginal fit of cast crowns-validation of the silicone replica method. J Oral Rehabil 2008;35:116-22. https://doi.org/10.1111/j.1365-2842.2003.01203.x
  22. Sorensen JA. A standardized method for determination of crown margin fidelity. J Prosthet Dent 1990;64:18-24. https://doi.org/10.1016/0022-3913(90)90147-5
  23. Pelekanos S, Koumanou M, Koutayas SO, Zinelis S, Eliades G. Micro-CT evaluation of the marginal fit of different InCeram alumina copings. Eur J Esthet Dent 2009;4:278-92.
  24. Rahme HY, Tehini GE, Adib SM, Ardo AS, Rifai KT. In vitro evaluation of the "replica technique" in the measurement of the fit of Procera crowns. J Contemp Dent Pract 2008;9:25-32.
  25. Council on Dental Materials and Devices. Revised American National Standards Instititute/American Dental Association Specification No. 8 for Zinc Phosphate Cement. J Am Dent Assoc 1978;96:121-3. https://doi.org/10.14219/jada.archive.1978.0032
  26. Assif D, Rimer Y, Aviv I. The flow of zinc phosphate cement under a full-coverage restoration and its effect on marginal adaptation according to the location of cement application. Quintessence Int 1987;18:765-74.
  27. Hung SH, Hung KS, Eick JD, Chappell RP. Marginal fit of porcelain-fused-to-metal and two types of ceramic crown. J Prosthet Dent 1990;63:26-31. https://doi.org/10.1016/0022-3913(90)90260-J
  28. Gulker I. Margins. N Y State Dent J 1985;51:213-5, 217.
  29. McLean JW, von Fraunhofer JA. The estimation of cement film thickness by an in vivo technique. Br Dent J 1971;131:107-11. https://doi.org/10.1038/sj.bdj.4802708

Cited by

  1. Marginal Accuracy and Internal Fit of 3-D Printing Laser-Sintered Co-Cr Alloy Copings vol.10, pp.1, 2017, https://doi.org/10.3390/ma10010093
  2. Misfit and fracture load of implant-supported monolithic crowns in zirconia-reinforced lithium silicate vol.25, pp.3, 2017, https://doi.org/10.1590/1678-7757-2016-0233
  3. Direct metal laser sintering 방식을 이용하여 제작한 다양한 고정성 보철물 수복 증례 vol.32, pp.3, 2015, https://doi.org/10.14368/jdras.2016.32.3.246
  4. CAD/CAM 밀링 시스템을 활용한 단일 치관과 3본 교의치의 3D 적합도 평가 vol.39, pp.1, 2015, https://doi.org/10.14347/kadt.2017.39.1.35
  5. Two different techniques of manufacturing TMJ replacements - A technical report vol.45, pp.9, 2015, https://doi.org/10.1016/j.jcms.2017.06.003
  6. Comparison of the Accuracy of Fit of Metal, Zirconia, and Lithium Disilicate Crowns Made from Different Manufacturing Techniques vol.28, pp.5, 2015, https://doi.org/10.1111/jopr.13029
  7. Evaluation of strain distribution on an edentulous mandible generated by cobalt-chromium metal alloy fixed complete dentures fabricated with different techniques: An in vitro study vol.122, pp.1, 2015, https://doi.org/10.1016/j.prosdent.2018.10.034
  8. Comparative evaluation of the effect of artificial aging on the marginal leakage of cast crowns luted with three cements: An in vitro study vol.12, pp.7, 2015, https://doi.org/10.4103/jioh.jioh_96_19
  9. Adaptability Evaluation of Metal-Ceramic Crowns Obtained by Additive and Subtractive Technologies vol.10, pp.16, 2015, https://doi.org/10.3390/app10165563
  10. Reliability of Metal 3D Printing with Respect to the Marginal Fit of Fixed Dental Prostheses: A Systematic Review and Meta-Analysis vol.13, pp.21, 2020, https://doi.org/10.3390/ma13214781
  11. Applications of three-dimensional printers in prosthetic dentistry vol.63, pp.3, 2015, https://doi.org/10.2334/josnusd.21-0072
  12. Fracture Load of Metal, Zirconia and Polyetheretherketone Posterior CAD-CAM Milled Fixed Partial Denture Frameworks vol.14, pp.4, 2015, https://doi.org/10.3390/ma14040959
  13. Comparison of Fracture Load of the Four Translucent Zirconia Crowns vol.26, pp.17, 2015, https://doi.org/10.3390/molecules26175308
  14. Assessment of internal fit and micro leakage of conventionally fabricated ceramometallic restoration versus CAD wax and press veneering (in-vitro study) vol.7, pp.1, 2021, https://doi.org/10.1038/s41405-021-00072-7
  15. A review on current additive manufacturing technologies and materials used for fabrication of metal-ceramic fixed dental prosthesis vol.35, pp.23, 2021, https://doi.org/10.1080/01694243.2021.1899699