Abstract
It is possible for the nation's largest flowmeter calibration system in K-water to calibrate flow rate up to $2,700m^3/h$ and diameter 800mm. However, the calibration and measurement capability of K-water's system is not satisfied in comparison with other developed countries. In this study, we find the dominant factors related to the uncertainty of weight and time measurement for gravimetric flowmeter calibration system. As a results of improving the system, the combined standard uncertainty has been improved $1.099{\times}10^{-3}$ to $2.332{\times}10^{-4}$. So calibration and measurement capability got 0.08 percent of the relative expanded uncertainty for maximum flow rate using the coverage factor(k=2).