A Study on Accuracy and Usefulness of In-vivo Dosimetry in Proton Therapy

양성자 치료에서 생체 내 선량측정 검출기(In-vivo dosimety)의 정확성과 유용성에 관한 연구

  • Kim, Sunyoung (Department of International Radiological Science, Hallym University of Graduate Studies) ;
  • Choi, Jaehyock (Department of International Radiological Science, Hallym University of Graduate Studies) ;
  • Won, Huisu (Department of International Radiological Science, Hallym University of Graduate Studies) ;
  • Hong, Joowan (Department of International Radiological Science, Hallym University of Graduate Studies) ;
  • Cho, Jaehwan (Department of International Radiological Science, Hallym University of Graduate Studies) ;
  • Lee, Sunyeob (Department of International Radiological Science, Hallym University of Graduate Studies) ;
  • Park, Cheolsoo (Department of International Radiological Science, Hallym University of Graduate Studies)
  • 김선영 (한림국제대학원대학교 국제방사선학과) ;
  • 최재혁 (한림국제대학원대학교 국제방사선학과) ;
  • 원희수 (한림국제대학원대학교 국제방사선학과) ;
  • 홍주완 (한림국제대학원대학교 국제방사선학과) ;
  • 조재환 (한림국제대학원대학교 국제방사선학과) ;
  • 이선엽 (한림국제대학원대학교 국제방사선학과) ;
  • 박철수 (한림국제대학원대학교 국제방사선학과)
  • Received : 2014.04.22
  • Accepted : 2014.06.25
  • Published : 2014.06.30

Abstract

In this study, the authors attempted to measure the skin dose by irradiating the actual dose on to the TLD(Thermo-Luminescence Dosimeter) and EBT3 Film used as the In-vivo dosimetry after planning the same treatment as the actual patient on a Phantom, because the erythema or dermatitis is frequently occurred on the patients' skin at the time of the proton therapy of medulloblastoma patient receiving the proton therapy. They intended to know whether there is the usefulness for the dosimetry of skin by the comparative analysis of the measured dose values with the treatment planned skin dose. The CT scan from the Brain to the Pelvis was done by placing a phantom on the CSI(Cranio-spinal irradiation) Set-up position of Medulloblastoma, and the treatment Isocenter point was aligned by using DIPS(Digital Image Positioning System) in the treatment room after planning a proton therapy. The treatment Isocenter point of 5 areas that the proton beam was entered into them, and Markers of 2 areas shown in the Phantom during CT scans, that is, in all 7 points, TLD and EBT3 Film pre-calibrated are alternatively attached, and the proton beam that the treatment was planned, was irradiated by 10 times, respectively. As a result of the comparative analysis of the average value calculated from the result values obtained by the repeated measurement of 10 times with the Skin Dose measured in the treatment planning system, the measured dose values of 6 points, except for one point that the accurate measurement was lacked due to the measurement position with a difficulty showed the distribution of the absolute dose value ${\pm}2%$ in both TLD and EBT Film. In conclusion, in this study, the clinical usefulness of the TLD and EBT3 Film for the Enterance skin dose measurement in the first proton therapy in Korea was confirmed.

본 연구는 양성자 치료를 시행하고 있는 수모세포종 환자의 양성자 치료 시 환자들의 Skin에서 홍반이나 피부염이 발생하는 경우가 빈번하여 Phantom에서 실제 환자와 동일한 치료계획 후 생체 내 선량측정 검출기(In-vivo dosimetry)로 사용되는 열 형광선량계(Thermo-luminescence Dosimeter, TLD)와 EBT3 Film으로 실제선량을 조사 측정한다. 측정된 선량값을 치료계획 된 피부선량과 비교 분석하여 양성자 치료 시 임상적으로 피부선량 측정 사용에 유용성이 있는지 알아보고자 한다. Phantom을 수모세포종의 전뇌척수치료(Cranio-spinal irradiation, CSI) Set-up position으로 위치하여 Brain에서 부터 Pelvis까지 CT촬영하고 양성자 치료계획 후 치료실에서 디지털 영상 포지셔닝 시스템(Digital Imaging Positioning System, DIPS)을 이용하여 정확하게 치료 Isocenter지점을 맞춘다. 양성자 빔이 들어가는 5부위의 치료 Isocenter 지점과 CT촬영 시 Phantom에 표시한 Marker 2부위, 모두 7지점에 미리 교정을 마친 열 형광선량계와 EBT3 Film을 교차해가면서 위치하여 붙이고 치료 계획된 양성자 빔을 각각 10회씩 조사한다. 10회 반복 측정하여 얻은 값을 평균산출하여 치료계획 시스템 에서 측정된 Skin Dose와 비교 분석한 결과, 측정위치의 어려움이 있는 정확한 측정이 결여된 한 지점을 제외한 나머지 6지점의 측정 선량값은 열 형광선량계와 EBT3 Film 모두 절대 선량값과 ${\pm}2%$이내의 분포를 보였다. 결론적으로, 본 연구에서는 국내 최초의 양성자 치료에서 Enterance skin dose 측정을 위한 In-vivo dosimetry로 열 형광선량계와 EBT3 Film의 임상적인 유용성을 확인하였다.

Keywords

References

  1. JooYoung Kim "Proton therapy of brain tumors in children", J., Korean Med. Assoc. 55.5, pp. 454-462, 2012. https://doi.org/10.5124/jkma.2012.55.5.454
  2. Vatnitsky, Stanislav M.,. "Radiochromic film dosimetry for clinical proton beams", Applied radiation and isotopes 48.5, pp. 643-651, 1997. https://doi.org/10.1016/S0969-8043(97)00342-4
  3. Seongsu Shin, et al., "A Study on the Skin Dose Measurement of Glass Dosimeter Use During Rectal Cancer 3field Technic and Cervical Cancer 4field Technic Radiation Therapy" Radiation Science and Technology, pp. 157-163, 2011.
  4. Incheol Im, et al. "Optical stimulation luminescence dosimeters (OLSDs) radiation therapy to the patient's skin using a rectal dose measurements", Journal of Radiation Protection 36.2, pp. 86-93, 2011.
  5. Sorriaux, Jefferson, et al. "Evaluation of Gafchromic< sup>(R) EBT3 films characteristics in therapy photon, electron and proton beams", Physica Medica 29.6 pp. 599-606, 2013. https://doi.org/10.1016/j.ejmp.2012.10.001
  6. Eujung Hwang et al. "Total body skin electron beam therapy (TSET) was used in the film dosimetry EBT2." Medical Physics, 21.1 pp. 60-69, 2010.
  7. JeongEun Rah, Dongho Shin b, Jeonghoon Park "Dosimetric verification of treatment planning system superficial dose calculations for proton beam", Radiation Measurements 55, pp. 60-63, 2013. https://doi.org/10.1016/j.radmeas.2013.02.015
  8. Weaver R.D., Gerbi B.J., Dusenbery K.E., "Evaluation of dose variaton during total skin electron irradiation using variation thermoluminescent dosimeters", Int. J. Radiat. Oncol. Biol. Phys 33, pp. 475-478
  9. Park. S. A., et al., "Dose verification of proton beam therapy using the Gafchromic EBT film.", Radiation Measurements 46.8 pp. 717-721, 2011. https://doi.org/10.1016/j.radmeas.2011.06.027
  10. Butson, Martin J., et al. "Energy response of the new EBT2 radiochromic film to x-ray radiation." Radiation Measurements 45.7, pp. 836-839, 2010. https://doi.org/10.1016/j.radmeas.2010.02.016
  11. Dreindl, Ralf, Dietmar Georg, Markus Stock. "Radiochromic film dosimetry: Considerations on precision and accuracy for EBT2 and EBT3 type films", Zeitschrift fur Medizinische Physik, 2013.
  12. Butson, Martin J., et al. "Radiochromic film for medical radiation dosimetry." Materials Science and Engineering: R: Reports 41.3, pp. 61-120, 2003. https://doi.org/10.1016/S0927-796X(03)00034-2
  13. Fiorini, F., et al. "Under-response correction for EBT3 films in the presence of proton spread out Bragg peaks." Physica Medica, 2014.
  14. Bilski, P., et al. "Thermoluminescence efficiency of LiF: Mg, Cu, P (MCP-N) detectors to photons, beta-electrons, alpha particles and thermal neutrons." Radiation protection dosimetry 55.1 pp. 31-38, 1994. https://doi.org/10.1093/oxfordjournals.rpd.a082372
  15. Martisikova, M., et al. "Use of Gafchromic(R) EBT films in heavy ion therapy." Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 591.1, pp. 171-173, 2008.
  16. Devic, S., et al. "Accurate skin dose measurements using radiochromic film in clinical applications." Medical physics 33.4, pp. 1116-1124, 2006. https://doi.org/10.1118/1.2179169
  17. Zullo, John R., et al. "LiF TLD-100 as a dosimeter in high energy proton beam therapy-can it yield accurate results?." Medical Dosimetry 35.1, pp. 63-66, 2010. https://doi.org/10.1016/j.meddos.2009.03.001
  18. Su, Fan-Chi, Chengyu Shi, and Niko Papanikolaou. "Clinical application of GAFCHROMIC< sup>(R) EBT film for< i> in vivo dose measurements of total body irradiation radiotherapy." Applied Radiation and Isotopes 66.3, pp. 389-394, 2008. https://doi.org/10.1016/j.apradiso.2007.09.015
  19. Hsi, Wen C., et al. "Image-guided method for TLD-based in vivo rectal dose verification with endorectal balloon in proton therapy for prostate cancer." Medical physics 40.5, 051715, 2013. https://doi.org/10.1118/1.4801901
  20. Fuss M, Sturtewagen E, De Wagter C, Georg D: Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance. Phys Med Biol 52, pp. 4211-4225, 2007. https://doi.org/10.1088/0031-9155/52/14/013
  21. Bufacchi, A., et al. "< i> In vivo EBT radiochromic film dosimetry of electron beam for Total Skin Electron Therapy (TSET)." Physica Medica 23.2, pp. 67-72, 2007. https://doi.org/10.1016/j.ejmp.2007.03.003
  22. Choi, Joon Yong, et al. "Development of a Thermoplastic Oral Compensator for Improving Dose Uniformity in Radiation Therapy for Head and Neck Cancer." Progress in Medical Physics 23.4, pp. 269-278, 2012.
  23. Sądel, M., et al. "Relative thermoluminescent efficiency of LiF detectors for proton radiation: Batch variability and energy dependence." Radiation Measurements 56, pp. 205-208, 2013. https://doi.org/10.1016/j.radmeas.2013.01.052
  24. Soares, Christopher G. "Radiochromic film dosimetry." Radiation measurements 41, S100-S116, 2006. https://doi.org/10.1016/j.radmeas.2007.01.007
  25. Dempsey, James F., et al. "Validation of a precision radiochromic film dosimetry system for quantitative two-dimensional imaging of acute exposure dose distributions." Medical physics 27.10, pp. 2462-2475, 2000. https://doi.org/10.1118/1.1290488

Cited by

  1. Characterization of light transmissions in various optical fibers with proton beam vol.71, pp.11, 2017, https://doi.org/10.3938/jkps.71.842