Abstract
We have been unwittingly exposed to magnetic field. Biological effects due to electromagnetic field exposure has been studied over the past several decades. There has been epidemiological studies and laboratory studies. In the case of laboratory studies, the effect has been observed through the cell stimulation of an artificial magnetic field. The used frequency and waveform are various in the cell experiment, but the intensity of the magnetic field is usually around 10G. Intensity of Earth's magnetic field is about 300-400mG. Although this intensity is not much higher than artificial magnetic field as 3-4% amplitude, we can't ignore the effect because every living thing on earth has already adapted for long time ago. In this study, we have designed and implemented the 3-axis magnetic field generation system in oder to remove the static magnetic field in the real-time and kept its intensity less than ${\pm}5mG({\pm}0.5uT)$ using 3-axis helmholtz coil.
자기장에 대하여 우리는 부지불식간에 많이 노출되고 있고 이와 관련된 영향에 대한 연구도 많이 진행되어 왔다. 역학 조사를 통한 연구도 있고, 실험실 수준의 연구도 있다. 실험실 수준의 연구의 경우, 인위적인 자기장 생성을 통해서 세포에 자극을 가하고 이에 대한 영향을 관찰한다. 세포 실험에서 많이 사용되는 주파수와 파형은 다양하지만, 그 크기는 보통 10G 내외이다. 지구 자계의 크기인 300-400mG 정도와 비교하면 크다고 할 수 없으나, 오래 전부터 이 환경에 적응되어 왔기 때문에 인위적으로 가하는 자기장 세기의 3-4%의 크기지만 무시할 수 없다. 본 연구에서는 주변 자기장에 대한 영향을 고려하여 실험할 수 있도록 양방향 정자계 생성이 가능한 자기장 발생 장치를 설계하고 구현하였다. 제안된 시스템으로 조절 가능한 정자계의 크기는 ${\pm}500mG({\pm}50uT)$까지이며 최소 ${\pm}5mG({\pm}0.5uT)$까지 주변 자기장을 실시간으로 상쇄시킬 수 있었다.