자동혈압계의 오실레이션 신호를 이용한 혈압 측정

Measuring Blood Pressure Using Oscillation Signal from an Automatic Sphygmomanometer

  • 투고 : 2012.04.04
  • 심사 : 2012.10.11
  • 발행 : 2012.11.01

초록

This study describes an oscillometric-based blood pressure measuring algorithm by detecting turning points of oscillation signal from digitally filtered cuff signals of an automatic sphygmomanometer. The blood pressure measuring algorithm uses a characteristic ratios method from the turning points. The accurate values of the systolic/diastolic blood presures(SBP/DBP) are calculated using the peaks in the ranges of characteristic ratios. Performances of the proposed algorithm and four automatic sphygmomanometers are compared with the mercury manometer(manual type sphygmomanometer), regarding the SBP and DBP values of manual sphygmomanometer as the reference values. The performance test showed the proposed algorithm revealed the best results in errors and a statistical analysis. Therefore this algorithm can be usable in any automatic sphygmomanometers.ssure states. This may be compromising results for subject-independent sensibility evaluation using EEG signal.

키워드

참고문헌

  1. H. Asada, A. Reisner, P. Shaltis, and D. McCombie, "Towards the development of wearable blood pressure sensors : A photo-plethysmograph approach using conducting polymer actuators", Int'l Conf. IEEE-EMBS, pp. 4156-4159, 2005.
  2. J. V. Edmond, M. Senbos, J. F. Crul, "Invasive vs non-invasive measurement of arterial pressure", Br J Anesth, vol. 57, pp. 434-444, 1985. https://doi.org/10.1093/bja/57.4.434
  3. J. I. Carr, J. M. Brown, Introduction to biomedical equipment technology, Prentice Hall, pp. 197-210, 1993.
  4. JCTB Moreas, M. Cerulli, PS. Ng, "Development of a new oscillometric blood pressure system", IEEE Computers in Cardiology Conf., pp. 467-470. 1999.
  5. JCTB Moreas, M. Cerulli, PS. Ng, "A strategy for determination of systolic, mean and diastolic blood pressures from oscillometric pulse profiles", IEEE Computers in Cardiology Conf., vol. 27, pp. 211-214, 2000.
  6. G. M. Drzewiecki, R. Hood, and H. Apple. "Theory of the oscilometric maximum and the systolic and diastolic detection ratios", Ann Biomed. Eng, vol. 22, pp. 88-96, 1994. https://doi.org/10.1007/BF02368225
  7. S. S. Lim, K. J. Lee, "Analysis of Oscillometric Model based on Shape of Arterial Pressure", J. Biomed. Eng. Res., vol. 21, no.4, pp. 411-417, 2000.
  8. L. A. Geddes, M. Voelz, C. Combs, D. Reiner, and C. F. Babbs, "Characterization of the oscillometric method for measuring indirect blood pressure", Ann. Biomed. Eng., vol. 10. pp. 271-280. 1983.