Abstract
This study was performed to find a method to decrease the radiation exposure of patients when obtaining anteroposterior pelvic images maintaining the quality of the image by using the copper filter built in the DR equipment now being used in our hospital. We measured the level of radiation by changing the filters from none filter to 0.1mmCu, 0.2mmCu, and 0.3mmCu and detected the organ level of radiation by PCXMC program. As a result, high levels of radiation were detected in the bladders and testicles and the levels were decreased as the thickness of the filter was increased and the amount of decrease was the highest in 0.1mmCu. And we asked for the expert opinions to 3 radiologists and as a result, only images obtained by 0.1mmCu filter out of all the images on which copper filters were used were accepted as the ones with diagnostic value same as none filter. At this time, the incident dose on the pelvic region was 0.895mGy which was smaller than the one in none filter by 47%. Therefore, using 0.1mmCu when obtaining anteroposterior pelvic images can effectively decrease the radiation exposure of patients.
본 연구는 본원에서 사용 중인 DR 장비에 내장되어 있는 구리필터를 이용하여 골반전후방향 촬영 시 영상의 화질을 저하시키지 않으면서 환자의 피폭선량을 줄이고자 실시하였다. 동일 조사선량으로 none filter, 0.1mmCu, 0.2mmCu, 0.3mmCu로 변화시켜 선량을 측정하고 PCXMC 프로그램으로 장기선량을 산출하였다. 결과는 고환과 방광에서 높은 선량이 검출 되었으며 필터 두께가 증가할수록 선량이 감소하였고 감소폭은 0.1mmCu 일 때 가장 크게 나타났다. 또한 촬영된 영상을 3명의 영상의학과 전문의에게 의뢰하여 평가해 본 결과 0.1mmCu를 사용한 영상이 none filter시와 가장 동일한 진단적 가치가 있는 영상으로 평가 되었고 이때 골반부 입사표면선량은 0.895mGy로 none filter 대비 47%의 선량을 줄일 수 있었다. 따라서 골반전후방향 촬영 시 0.1mmCu를 사용하여 촬영하면 환자의 피폭선량을 효과적으로 감소시킬 수 있을 것이다.