Abstract
We have investigated temperature dependence and long-term change of humidity measurement from 32 relative humidity sensors. The readings of the humidity sensors depended not only the reference humidity, but also temperature of the chamber. Approximately, the temperature dependence of the humidity sensor in average was 0.05 %R.H./$^{\circ}C$ in the temperature range from $5^{\circ}C$ to $55^{\circ}C$. For humidity sensors that have an internal temperature compensation circuit, the resulting temperature dependence was weaker by 20%. It should be also noted that for the humidity sensors used in this work underwent ${\pm}3$ %R.H. change per year for level of confidence of 95%. The users of relative humidity sensors may refer this value as a minimum change when they set the calibration interval of the humidity sensors.