Browse > Article
http://dx.doi.org/10.5369/JSST.2012.21.6.420

Statistical Analysis on the Temperature Dependence and Long-Term Change of Relative Humidity Sensors  

Kim, Jong Chul (Center for Thermometry, Div. of Physical Metrology, Korea Research Institute of Standards and Science)
Choi, Byung Il (Center for Thermometry, Div. of Physical Metrology, Korea Research Institute of Standards and Science)
Woo, Sangbong (Center for Thermometry, Div. of Physical Metrology, Korea Research Institute of Standards and Science)
Yang, Inseok (Center for Thermometry, Div. of Physical Metrology, Korea Research Institute of Standards and Science)
Publication Information
Journal of Sensor Science and Technology / v.21, no.6, 2012 , pp. 420-424 More about this Journal
Abstract
We have investigated temperature dependence and long-term change of humidity measurement from 32 relative humidity sensors. The readings of the humidity sensors depended not only the reference humidity, but also temperature of the chamber. Approximately, the temperature dependence of the humidity sensor in average was 0.05 %R.H./$^{\circ}C$ in the temperature range from $5^{\circ}C$ to $55^{\circ}C$. For humidity sensors that have an internal temperature compensation circuit, the resulting temperature dependence was weaker by 20%. It should be also noted that for the humidity sensors used in this work underwent ${\pm}3$ %R.H. change per year for level of confidence of 95%. The users of relative humidity sensors may refer this value as a minimum change when they set the calibration interval of the humidity sensors.
Keywords
Humidity sensor; Temperature dependence; Long-term change;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 KS A 0078 "Humidity-measurement methods", 1999.
2 C. B. Park and D. C. Lee,"Technological trends in humidity sensors", Proceeding of Korean Institute of Electrical Engineers, Vol. 41, No. 2, pp. 12-21, 1992.
3 S. P. Lee, J. Y. Rim, and Y. K. Yoon, "A resistivetype humidity sensor using PMMA thin film", J. Sensor Sci. & Tech., Vol. 1, No. 2, pp. 125-130, 1992.
4 W.-S. Chung, "A temperature-compensated hygrometer using resistive humidity sensors", Proceedings of the Institute of Electronics Engineers of Korea, Vol. 43-SC, No. 6, pp. 27-32, 2006.   과학기술학회마을
5 T. Isao, "Design of driving circuits of humidity sensors based on the change in electrical resistance", EE Report Monthly, No. 2, pp. 73-84. 2009.
6 KS B 5344 "Hygrometers-test method", 1997.