Effect of cutting flute length and shape on insertion and removal torque of orthodontic mini-implants

교점용 미니 임플랜트의 cutting flute의 길이 및 형태에 따른 식립 및 제거 토크의 비교

  • Yun, Soon-Dong (Department of Orthodontics, School of Dentistry, Chosun University) ;
  • Lim, Sung-Hoon (Department of Orthodontics, School of Dentistry, Chosun University)
  • 윤순동 (조선대학교 치과대학 교정학교실) ;
  • 임성훈 (조선대학교 치과대학 교정학교실)
  • Published : 2009.04.30

Abstract

Objective: The purpose of this study was to evaluate the effect of length and shape of cutting flute on mechanical properties of orthodontic mini-implants. Methods: Three types of mini-implants with different flute patterns (Type A with 2.6 mm long flute, Type B with 3.9 mm long and straight flute, Type C with 3.9 mm long and helical flute) were inserted into the biomechanical test blocks (Sawbones Inc., USA) with 2 mm and 4 mm cortical bone thicknesses to test insertion and removal torque. Results: In 4 mm cortical bone thickness, Type C mini-implants showed highest maximum insertion torque, then Type A and Type B in order. Type C also showed shortest total insertion time and highest maximum removal torque, but Type A and B didn't showed statistically significant difference in insertion time and removal torque. In 2 mm cortical bone thickness, there were no significant difference in total insertion time and maximum removal torque in three types of mini-implants, but maximum insertion torque of Type A was higher than two other Types of mini-implants. Conclusions: Consideration about length and shape of cutting flute of mini-implant is also required when the placement site has thick cortical bone.

본 연구는 self-tapping screw의 특징적 구조인 cutting flute의 길이 및 형태 변화에 따른 미니 임플랜트의 식립 및 제거 토크를 비교해보기 위해 시행되었다. 외경, 내경, 길이 및 pitch 등 다른 조건이 모두 동일하고 cutting flute의 형태만 다른 세 종류의 미니 임플랜트를 사용하였다(A군; 2.6 mm 길이의 flute가 형성된 그룹, B군; 3.9 mm의 길이이면서 직선형의 flute를 갖는 그룹, C군; 3.9 mm의 길이이면서 나선형의 flute를 갖는 그룹). 골밀도를 균일하게 하기 위해 실험용 인공골(Sawbones Inc., USA)을 사용하였으며, 수직력 및 회전속도를 일정하게 하기 위해 구동식 토크시험기를 이용하였다. 그 결과 다음과 같은 결론을 얻었다. 총 삽입 시간은 2 mm 두께의 피질골 시편에서는 세 군간에 차이를 보이지 않으나, 4 mm 두께의 피질골 시편에서는 C군 > B군 > A군 순으로 짧은 삽입 시간을 보였다. 최대 삽입 토크는 2 mm 두께의 시편에서는 A군이 다른 두 군에 비해 높은 값을 보이나, 4 mm 두께의 피질골 시편에서는 C군 > A군 > B군 순으로 높은 값을 보였다. 최대 제거 토크는 2 mm 두께의 피질골 시편에서는 세 군 사이에 차이가 없으나, 4 mm 두께의 피질골 시편에서는 C군이 다른 두 군에 비해 높은 값을 보여주었다. 이상의 결과로 보아 피질골이 두꺼운 부위에 미니 임플랜트를 식립할 경우 cutting flute의 길이 및 형태 또한 고려하여 미니 임플랜트를 선택하는 것이 바람직하다.

Keywords

References

  1. Kanomi R. Mini-implant for orthodontic anchorage. J Clin Orthod 1997;31:763-7
  2. Costa A, Raffaini M, Melsen B. Miniscrews as orthodontic anchorage a preliminary report. Int J Adult Orthodon Orthognath Surg 1998;13:201-9
  3. Park HS. A new proctocol of the sliding mechanics with micro-implant anchorage (MIA). Korean J Orthod 2000;30:677-85
  4. Park HS. The skeletal cortical anchorage using titanium microscrew implants. Korean J Orthod 1999;29:699-706
  5. Oh MY, Chung KR, Kwon YD, Ryu DM, Lee BS. The clinical use of miniscrew for intraoral anchorage (1). J Korean Dent Assoc 2000;38:18-21
  6. Park HS, Jeong SH, Kwon OW. Factors affecting the clinical success of screw implants used as orthodontic anchorage. Am J Orthod Dentofacial Orthop 2006;130:18-25 https://doi.org/10.1016/j.ajodo.2004.11.032
  7. Moon CH, Lee DG, Lee HS, Im JS, Baek SH. Factors associated with the success rate of orthodontic miniscrews placed in the upper and lower posterior buccal region. Angle Orthod 2008;78:101-6 https://doi.org/10.2319/121706-515.1
  8. Kuroda S, Sugawara Y, Deguchi T, Kyung HM, Takano-Yamamoto T. Clinical use of miniscrew implants as orthodontic anchorage: success rate and postoperative discomfort. Am J Orthod Dentofacial Orthop 2007;131:9-15 https://doi.org/10.1016/j.ajodo.2005.02.032
  9. Cheng SJ, Tseng IY, Lee JJ, Kok SH. A prospective study of the risk factors associated with failure of mini-implants used for orthodontic anchorage. Int J Oral Maxillofac Implants 2004;19:100-6
  10. Kuroda S, Yamada K, Deguchi T, Hashimoto T, Kyung HM, Takano-Yamamoto T. Root proximity is a major factor for screw failure in orthodontic anchorage. Am J Orthod Dentofacial Orthop 2007;131(Suppl 4):68S-73S https://doi.org/10.1016/j.ajodo.2006.06.017
  11. You ZH, Bell WH, Schneiderman ED, Ashman RB. Biomechanical properties of small bone screws. J Oral Maxillofac Surg 1994;52:1293-302 https://doi.org/10.1016/0278-2391(94)90052-3
  12. Kim JW, Cho IS, Lee SJ, Kim TW, Chang YI. Mechanical analysis of the taper shape and length of orthodontic mini-implant for initial stability. Korean J Orthod 2006;36:55-62
  13. Kim JW, Cho IS, Lee SJ, Kim TW, Chang YI. Effect of dual pitch mini-implant design and diameter of an orthodontic mini-implant on the insertion and removal torque. Korean J Orthod 2006;36:275-83
  14. Yano S, Motoyoshi M, Uemura M, Ono A, Shimizu N. Tapered orthodontic miniscrews induce bone-screw cohesion following immediate loading. Euro J Orthod 2006;28:541-6 https://doi.org/10.1093/ejo/cjl044
  15. Song YY, Cha JY, Hwang CJ. Evaluation of insertion torque and pull-out strength of mini-screws according to different thickness of artificial cortical bone. Korean J Orthod 2007;37:5-15
  16. Lim SA, Cha JY, Hwang CJ. Insertion torque of orthodontic miniscrews according to changes in shape, diameter and length. Angle Orthod 2008;78:234-40 https://doi.org/10.2319/121206-507.1
  17. Heidemann W, Gerlach KL, Gröbel KH, Köllner HG. Drill free screws: a new form of osteosynthesis screw. J Craniomaxillofac Surg 1998;26:163-8 https://doi.org/10.1016/S1010-5182(98)80007-3
  18. Heidemann W, Gerlach KL, Grobel KH, Kollner HG. Influence of different pilot hole sizes on torque measurements and pullout analysis of osteosynthesis screws. J Craniomaxillofac Surg 1998;26:50-5 https://doi.org/10.1016/S1010-5182(98)80035-8
  19. Kim JW, Ahn SJ, Chang YI. Histomorphometric and mechanical analysis of the drill-free screw as orthodontic anchorage. Am J Orthod Dentofacial Orthop 2005;128:190-4 https://doi.org/10.1016/j.ajodo.2004.01.030
  20. Kim JW, Chang YI. Effect of drilling process in stability of micro-implants used for orthodontic anchorage. Korean J Orthod 2002;32:107-15
  21. Kim YH, Choi JH. The study about retention of miniscrews used for intraoral anchorage. J Korean Dent Assoc 2001;39:684-7
  22. Ansell RH, Scales JT. A study of some factors which affect the strength of screws and their insertion and holding power in bone. J Biomech 1968;1:279-302 https://doi.org/10.1016/0021-9290(68)90023-7
  23. Foley WL, Frost DE, Paulin WB Jr, Tucker MR. Uniaxial pullout evaluation of internal screw fixation. J Oral Maxillofac Surg 1989;47:277-80 https://doi.org/10.1016/0278-2391(89)90231-0
  24. Evans M, Spencer M, Wang Q, White SH, Cunningham JL. Design and testing of external fixator bone screws. J Biomed Eng 1990;12:457-62 https://doi.org/10.1016/0141-5425(90)90054-Q
  25. Yerby S, Scott CC, Evans NJ, Messing KL, Carter DR. Effect of cutting flute design on cortical bone screw insertion torque and pullout strength. J Orthop Trauma 2001;15:216-21 https://doi.org/10.1097/00005131-200103000-00012
  26. Wikenheiser MA, Markel MD, Lewallen DG, Chao EYS. Thermal response and torque resistance of five cortical half-pins under simulated insertion technique. J Orthop Res 1665;13:615-9 https://doi.org/10.1002/jor.1100130418
  27. Koistinen A, Santavirta SS, Kröger H, Lappalainen R. Effect of bone mineral density and amorphous diamond coatings on insertion torque of bone screws. Biomaterials 2005;26:5687-94 https://doi.org/10.1016/j.biomaterials.2005.02.003
  28. Seebeck J, Goldhahn J, Städele H, Messer P, Morlock MM, Schneider E. Effect of cortical thickness and cancellous bone density on the holding strength of internal fixator screws. J Orthop Res 2004;22:1237-42 https://doi.org/10.1016/j.orthres.2004.04.001
  29. Deguchi T, Nasu M, Murakami K, Yabuuchi T, Kamioka H, Takano-Yamamoto T. Quantitative evaluation of cortical bone thickness with computed tomographic scanning for orthodontic implants. Am J Orthod Dentofacial Orthop 2006;129:721.e7-721.e12 https://doi.org/10.1016/j.ajodo.2006.02.026
  30. Liou EJW, Chen PH, Wang YC, Lin JC. A computed tomographic image study on the thickness of the infrazygomatic crest of the maxilla and its clinical implications for miniscrew insertion. Am J Orthod Dentofacial Orthop 2007;131:352-6 https://doi.org/10.1016/j.ajodo.2005.04.044
  31. Motoyoshi M, Hirabayashi M, Uemura M, Shimizu N. Recommended placement torque when tightening an orthodontic mini-implant. Clin Oral Implants Res 2006;17:109-14 https://doi.org/10.1111/j.1600-0501.2005.01211.x

Cited by

  1. The effects of different pilot-drilling methods on the mechanical stability of a mini-implant system at placement and removal: a preliminary study vol.41, pp.5, 2009, https://doi.org/10.4041/kjod.2011.41.5.354
  2. Insertion torque and orthodontic mini-implants: A systematic review of the artificial bone literature vol.227, pp.11, 2009, https://doi.org/10.1177/0954411913495986