유한요소법을 이용한 나이티놀 와이어 스텐트의 전산모사 실험 데이터 검증

The Verification of Computer Simulation of Nitinol Wire Stent Using Finite Element Analysis

  • 김진영 (가톨릭대학교 의과대학 의공학교실) ;
  • 정원균 (가톨릭대학교 의과대학 의공학교실) ;
  • 전동민 (가톨릭대학교 의과대학 의공학교실) ;
  • 신일균 ((주)엠아이텍 중재의학연구소) ;
  • 김한기 ((주)엠아이텍 중재의학연구소) ;
  • 신동오 (경희대학교 의학전문대학원 방사선종양학교실) ;
  • 김상호 (가톨릭대학교 의과대학 의공학교실) ;
  • 서태석 (가톨릭대학교 의과대학 의공학교실)
  • Kim, Jin-Young (Department of Biomedical Engineering, College of Medicine, The Catholic University of Korea) ;
  • Jung, Won-Gyun (Department of Biomedical Engineering, College of Medicine, The Catholic University of Korea) ;
  • Jeon, Dong-Min (Department of Biomedical Engineering, College of Medicine, The Catholic University of Korea) ;
  • Shin, Il-Gyun (Institute of Interventional Medicine, M. I. Tech Co., Lnc) ;
  • Kim, Han-Ki (Institute of Interventional Medicine, M. I. Tech Co., Lnc) ;
  • Shin, Dong-Oh (Department of Radiation Oncology, School of Medicine, Kyung Hee University) ;
  • Kim, Sang-Ho (Department of Biomedical Engineering, College of Medicine, The Catholic University of Korea) ;
  • Suh, Tae-Suk (Department of Biomedical Engineering, College of Medicine, The Catholic University of Korea)
  • 발행 : 2009.09.30

초록

최근 스텐트 시술의 수학적인 분석은 스텐트의 위치와 기계적인 반응을 측정하는 여러 가지 툴이 개발됨에 따라 발전하고 있다. 그러나 기존의 연구의 경우, 전산모사 실험을 통해 여러 요소에 따른 변형 정도 및 응력 분포, 팽창력 등과 같은 성능평가에 대한 연구가 주류를 이루며 전산모사 실험을 통한 응력 분포 및 변형 정도에 대한 소프트웨어의 검증은 거의 없는 실정이다. 따라서 본 연구에서는 만능재료 시험기를 이용한 실험을 통해 얻은 결과와 전산모사 실험 결과의 유의성 검증을 통해 해당 전산 모사용 소프트웨어를 통한 전산모사 실험 결과에 신뢰성을 부여하였다. 또한 이를 토대로 전산모사 실험을 통해 스텐트 피막 유무에 따른 스텐트의 응력분포를 분석하고자 하였다. 본 연구에서는 전산모사 실험 데이터와 실제 인장 시험 데이터를 통한 유의성 검증을 하고자 하였다. 스텐트의 연결 부분에 대해 유한요소 해석법을 적용하여 전산모사 실험을 수행한 결과, 178.93 MPa의 최대응력으로 인장시험 결과인 184.23 MPa와 거의 유사한 값을 가졌으며, 이를 통해 인장시험 결과와 전산모사 결과가 유의성을 가짐을 검증할 수 있었다. 또한 피막 유무에 따른 스텐트의 전산모사 실험을 통해 피막에서는 접촉부, 나이티놀에서는 연결부에서 최대 응력이 발생함을 알 수 있었다. 그리고 피막이 있는 경우에는 피막의 장력으로 인해 연결부분에서 형상을 유지하고 있으나 피막이 없는 경우에는 이격 및 교차부에서 슬립이 발생함을 알 수 있었다. 이러한 결과들을 통해 스텐트 설계 초기에 있어 해당 전산모사용 소프트웨어를 통해 해석을 수행하여 그에 따른 해석 결과를 설계에 반영함으로써 설계 변경을 최소화할 수 있고, 이에 따라 양질의 설계 품질 확보가 가능할 것으로 사료된다.

Recently, the mathematical analysis of stent simulation has been improved, with the help of development of various tool which measure mechanical property and location of stent in artery. The most crucial part of the stent modeling is how to design ideal stent and to evaluate the interaction between stent and artery. While there has been great deal of researches on the evaluation of the expansion, stress distribution, deformation of the stent in terms of the various parameters, few verification through computer simulation has been performed about deformation and stress distribution of the stent. In this study, we have produced the corresponding results between experimental test using Universal Testing Machine and computer simulation for the ideal model of stent. Also, we have analyzed and compared stress distribution of stent in the cases of that with membrane and that without membrane. The results of this study would provide minimum change of plan and good quality for ideal stent replacing damaged artery through the analysis using computer simulation in the early stage of stent design.

키워드

참고문헌

  1. Hedin M: The origin of the word stent. Acta Radiol 6:937-939 (1997)
  2. Capelli Claudio, Gervaso Francesca, Petrini L, et al: Assessment of tissue prolapse after balloon-expandable stenting: Influence of stent cell geometry. Med Eng Phys Article 31:441-447 (2008)
  3. Whitcher FD: Simulation of in vii/o loading conditions of Nitinol vascular stent structures. Computers & Structures 64:100-101(1997)
  4. Harnek J, Zoucas E, Stenram U, Cwikiel W: Insertion of self-expandable nitinol stents without previous balloon angioplasty reduces restenosis compared with PTA prior to stenting. Cardiovasc Intervent Radiol 25:430–436 (2002)
  5. Morton AC, Crossman D, Gunn J: The influence of physical stent parameters upon restenosis. Pathol Biol 52:196–205 (2004)
  6. Ioannis Iakovou, Thomas Schmidt, Erminio Bonizzoni, et al: Incidence, predictors, and outcome of thrombosis after successful implantation of drug-eluting stents. JAMA 293:2126-2130 (2005) https://doi.org/10.1001/jama.293.17.2126
  7. Schwartz RS, Huber KC, Murphy JG, et al: Restenosis and the proportional neointimal response to coronary artery injury: results in a porcine model. J Am Coll Cardiol 19:267–274(1992)
  8. Kornowski R, Hong MK, Tio FO, et al: In-stent restenosis: contributions of inflammatory responses and arterial injury to neointimal hyperplasia. J Am Coll Cardiol 31:224–230 (1998)
  9. Rogers C, Edelman ER: Endovascular stent design dictates experimental restenosis and thrombosis. Circulation 91:2995–3001 (1995)
  10. Liang D, Yang D, Qi M, et al: Finite element analysis of the implantation of a balloon-expandable stent in a stenosed artery. Int J Cardiol 104:314- 318 (2005) https://doi.org/10.1016/j.ijcard.2004.12.033
  11. SN David Chua, BJ Mac Donald, MSJ Hashmi: Finite-element simulation of stent expansion. J Mater Process Technol 120:335-340 (2002) https://doi.org/10.1016/S0924-0136(01)01127-X
  12. Ling DK, Yang DZ, Qi M, et al: Finite element analysis of the implantation of a ballon-expandable stent in a stenosed artery. Inter J Cardio 104:314-318 (2005) https://doi.org/10.1016/j.ijcard.2004.12.033
  13. Auricchio F, Di Loreto M, Sacco E: Finite-element analysis of a stenotic artery revascularization through a stent insertion. Comput Methods Biomech Biomed Eng 4:249–263 (2001)
  14. Joshi HB, Chitale SV, Nagarajan M, et al: A prospective randomized single-blind comparison of ureteral stents composed of firm and soft polymer. J Urol 174:2303 (2005) https://doi.org/10.1097/01.ju.0000181815.63998.5f
  15. Hibbeler RC: Mechanics of Materials. 2nd ed. Macmillan, New York, 1994
  16. Migliavacca F, Petrini L, Colombo M, Auricchio F, Pietrabissa R: Mechanical behavior of coronary stents investigated through the finite element method. J Biomech 35:803–811 (2002)