DOI QR코드

DOI QR Code

A Study on Error Characteristic of Flow Disturbance and Velocity for Electromagnetic and Ultrasonic Flowmeters

유속과 유동교란인자에 의한 전자식 및 초음파식 유량계의 오차특성 연구

  • Published : 2009.10.01

Abstract

In this study, the effect of flow disturbance such as contraction, expansion pipe and velocity deviation from low velocity of $0.1\;^m/s$ to $2.5\;^m/s$ on the error characteristics of the flowmeter was studied. Flow experiments using flowmeter calibration facility of K-water were undertaken for the cases of ultrasonic flowmeter based on transit-time method and electromagnetic flowmeter. Experimental results are presented that measurement error of expansion pipe are larger than contraction pipe. It is shown that the minimum straight length were required to remain of ${\pm}0.5%$ error for electromagnetic flowmeter and ${\pm}2.0%$ error for ultrasonic flowmeter.

Keywords

References

  1. 일본전기계측기공업협회 규격, 1987, '초음파유량계에 의한 유량측정방법,' JEMIS 032
  2. 일본계량기기공엽연합회, 1997, '계장 엔지니어를 위한 유량 계측 A to 2,' 테크하우스
  3. 한국수자원공사 2000, '수돗물공급규정'
  4. 한국표준과학연구원, 1991, '형식별 유량계 설치조건에 따른 특성 연구(II)'
  5. 황상윤, 2006, '초음파 유량계 기술개발 동향' 유체기계저널 제9권, 제3호, pp. 51-57 https://doi.org/10.5293/KFMA.2006.9.3.051