Development of the calibration system for very accurate thermometers at low temperatures

초정밀 온도센서의 저온 영역 교정 장치 개발

  • Yang, In-Seok (Division of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Song, Chang-Ho (Division of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Kang, Kee-Hoon (Division of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Kim, Yong-Gyoo (Division of Physical Metrology, Korea Research Institute of Standards and Science)
  • 양인석 (한국표준과학연구원 기반표준본부) ;
  • 송창호 (한국표준과학연구원 기반표준본부) ;
  • 강기훈 (한국표준과학연구원 기반표준본부) ;
  • 김용규 (한국표준과학연구원 기반표준본부)
  • Published : 2008.05.31

Abstract

For the accurate calibrations of thermometers at low temperatures, triple points of oxygen and argon have been realized using sealed-type triple point cells. The triple points were realized with both adiabatic and continuous heating methods. The resistance of three capsule-type standard platinum resistance thermometers were measured at the triple points of oxygen, argon, mercury and water for the calibrations of these thermometers in the range from $-220^{\circ}C$ to $0.01^{\circ}C$. The extended uncertainty, with the coverage factor of 2, of the calibration in this range is smaller than 1 mK.

Keywords

References

  1. H. Preston-Tomas, 'The International Temperature Scale of 1990 (ITS-90)', Metrologia, vol. 27, no. 1, pp. 3-10, 1990 https://doi.org/10.1088/0026-1394/27/1/002
  2. 강기훈, 김용규, 감기술, '캡슐형 백금저항온도계 국 제비교를 위한 네온, 산소, 아르곤, 수은 및 물의 삼 중점 실현', 센서학회지, 제9권, 제3호, pp. 153-162, 2000
  3. F. Pavese, J. Ancsin, D. N. Astrov, J. Bonhoure, G. Bonnier, G. T. Furukawa, R. C. Kemp, H. Mass, R. L. Rusby, H. Sakurai and Ling Shan-Kang, 'An international intercomparison of fixed points by means of sealed cells in the range 13.81 K - 90,686 K', Metrologia, vol. 20, no. 4, pp. 127-144, 1984 https://doi.org/10.1088/0026-1394/20/4/002
  4. 강기훈, 감기술, 김용규, 송창호, '전기로 및 봉입형 응고점 셀 내의 온도구배가 미치는 표준백금저항온 도계 온도측정의 불확도 요소 평가', 센서학회지, 제 13권 제6호, pp. 411-416, 2004
  5. 강기훈, 김용규, 감기술, '핵심국제비교용 물의 삼중 점 셀 제작과 비교 실험 및 불확도 평가', 센서학회 지, 제13권, 제2호, pp. 114-120, 2004
  6. B.W. Mangum, P. Bloembergen, M. V. Chattle, B. Fellmuth, P. Marcarino and A. I. Pokhodun, 'On the international temperature scale of 1990 (ITS-90)', Metrologia, vol. 34, no. 5, pp. 427-429, 1997 https://doi.org/10.1088/0026-1394/34/5/8

Cited by

  1. Calibration by Comparison and Uncertainty Assessment of Industrial Thermometers at the Boiling Point of Nitrogen vol.22, pp.6, 2013, https://doi.org/10.5369/JSST.2013.22.6.404
  2. Precise Temperature Control by Adjusting Flow of Liquid Nitrogen vol.25, pp.1, 2016, https://doi.org/10.5369/JSST.2016.25.1.65