Dose Measurements using Phantoms for Tube Voltage, Tube Current, Slice Thickness in MDCT

MDCT의 관전압, 관전류, 슬라이스 두께 변화에 따른 팬텀의 선량 분포 측정

  • Lee, Chang-Lae (Department of Radiological Science, College of Health Science and Research Institute of Health Science, Yonsei University) ;
  • Jeon, Seong-Su (Department of Radiology, Wonju Christian Hospital, Yonsei University) ;
  • Nam, So-Ra (Department of Radiological Science, College of Health Science and Research Institute of Health Science, Yonsei University) ;
  • Cho, Hyo-Min (Department of Radiological Science, College of Health Science and Research Institute of Health Science, Yonsei University) ;
  • Jung, Ji-Young (Department of Radiological Science, College of Health Science and Research Institute of Health Science, Yonsei University) ;
  • Kim, Hee-Joung (Department of Radiological Science, College of Health Science and Research Institute of Health Science, Yonsei University)
  • 이창래 (연세대학교 방사선학과, 보건과학연구소) ;
  • 전성수 (연세대학교 원주기독병원 영상의학과) ;
  • 남소라 (연세대학교 방사선학과, 보건과학연구소) ;
  • 조효민 (연세대학교 방사선학과 보건과학연구소) ;
  • 정지영 (연세대학교 방사선학과, 보건과학연구소) ;
  • 김희중 (연세대학교 방사선학과, 보건과학연구소)
  • Published : 2007.09.30

Abstract

The purpose of this study was to measure and evaluate radiation dose for MDCT parameters. Patient dose for various combination of MDCT parameters were experimentally measured, using MDCT (GE light speed plus 4 slice, USA), model 2026C electrometer (RADICAL 2026C, USA), standard Polymethylmethacrylate (PMMA) head and body CT dosimetry phantoms. In clinical situations, for a typical abdominal scan performed with MDCT at 120 kVp, 180 mAs, 20 mm collimation, and a pitch of 0.75 $CTDI_w,\;CTDI_{vol}$ were measured as 20.2 mGy, 26.9 mGy, respectively. When scan length is assumed as 271.3 mm, DLP and measured effective dose of the abdominal would be calculated as $729.1\;mGy{\cdot}cm$, 10.9 mSv, respectively.

본 연구는 MDCT (multi-detector computed tomography) 파라미터 변화에 따른 환자선량을 측정하고 평가 하고자 하였다. MDCT 파라미터의 다양한 변화에 의한 환자선량은 MDCT (GE light speed plus 4 slice, USA)와 model 2026C electrometer (RADICAL 2026C, USA), 그리고 head와 body의 CT선량 표준팬텀(standard polymethylmethacrylate)을 사용하여 측정 하였다. 그 결과 환자선량 $CTDI_w$ 값은 관전압과 관전류가 증가할수록 선형적으로 증가하였고 beam collimation이 증가 할수록 감소하였다. 따라서 관전압, 관전류는 환자의 피폭선량에 직접적인 영향을 주고 슬라이스 두께는 영향이 적은 것을 알 수 있었다. 임상의 전형적인 MDCT 복부 scan에서 120 kVp, 180 mAs, 20 mm collimation과 0.75 pitch의 조건하에 $CTDI_w$$CTDI_{vol}$의 측정값은 각각 20.2 mGy, 26.9 mGy이었고 스캔 길이가 271.3 mm인 환자의 DLP와 유효선량은 각각 $729.1\;mGy{\cdot}cm$, 10.9 mSv였다.