APPLICATION OF FINITE ELEMENT ANALYSIS TO EVALUATE IMPLANT FRACTURES

  • Kim Yang-Soo (Department of Prosthodontics, Graduate School, Seoul National University) ;
  • Kim Chang-Whe (Department of Prosthodontics, Graduate School, Seoul National University) ;
  • Lim Young-Jun (Department of Prosthodontics, Graduate School, Seoul National University) ;
  • Kim Myung-Joo (Department of Prosthodontics, Graduate School, Seoul National University)
  • Published : 2006.06.01

Abstract

Statement of problem. Higher fracture rates were reported for Branemark implants placed in the maxilla and for 3.75 mm diameter implants installed in the posterior region. Purpose. The purpose of this study was to investigate the fracture of a fixture by finite element analysis and to compare different diameter of fixtures according to the level of alveolar bone resorption. Material and Methods. The single implant and prosthesis was modeled in accordance with the geometric designs for the 3i implant systems. Models were processed by the software programs HyperMesh and ANSA. Three-dimensional finite element models were developed for; (1) a regular titanium implant 3.75 mm in diameter and 13 mm in length (2) a regular titanium implant 4.0 mm in diameter and 13 mm in length (3) a wide titanium implant 5.0 mm in diameter and 13 mm in length each with a cementation type abutment and titanium alloy screw. The abutment screws were subjected to a tightening torque of 30 Ncm. The amount of preload was hypothesized as 650 N, and round and flat type prostheses were 12 mm in diameter, 9 mm in height were loaded to 600 N. Four loading offset points (0, 2, 4, and 6 mm from the center of the implants) were evaluated. To evaluate fixture fracture by alveolar bone resorption, we investigated the stress distribution of the fixtures according to different alveola. bone loss levels (0, 1.5, 3.5, and 5.0 mm of alveolar bone loss). Using these 12 models (four degrees of bone loss and three implant diameters), the effects of load-ing offset, the effect of alveolar bone resorption and the size of fixtures were evaluated. The PAM-CRASH 2G simulation software was used for analysis of stress. The PAM-VIEW and HyperView programs were used for post processing. Results. The results from our experiment are as follows: 1. Preload maintains implant-abutment joint stability within a limited offset point against occlusal force. 2. Von Mises stress of the implant, abutment screw, abutment, and bone was decreased with in-creasing of the implant diameter. 3. With severe advancing of alveolar bone resorption, fracture of the 3.75 and the 4.0 mm diameter implant was possible. 4. With increasing of bending stress by loading offset, fracture of the abutment screw was possible.

Keywords

References

  1. Goodacre CJ, Kan JYK, Rungcharassaeng K. Clinical complication of osseointegrated implants. J Prosthet Dent 1999;81:537-552 https://doi.org/10.1016/S0022-3913(99)70208-8
  2. Rangert B, Krogh PHJ, Langer B, Van Roekel NB. Bending overload and implant fractures: A retrospective clinical analysis. Int J Oral Maxillofac Implants 1995;10:326-334
  3. Eckert SE, Meraw SJ, Cal E. Analysis of incidence and associated factors with fractured implants: A retrospective study. Int J Oral Maxillofac Implants 2000;15:662-667
  4. Naert I, Quirynen M, van Steenberghe D, Darius P. A six-year prosthodontic study of 509 consecutively inserted implants for the treatment of partial edentulism. J Prosthet Dent 1992;67:236-245 https://doi.org/10.1016/0022-3913(92)90461-I
  5. Quirynen M, Naert I, van 5teenberghe D, Dekeyser C, Callens A. Periodontal aspects of osseointegrated fixtures supporting a partial bridge. An up to 6 years retrospective study. J Clin Periodontol 1992;19:118-126 https://doi.org/10.1111/j.1600-051X.1992.tb00450.x
  6. Balshi TJ, Hernandez RE, Pryszlak MC, Rangert BO. A comparative study of one implant versus two replacing a single molar. Int J Oral Maxillofac Implants 1996;11:372-378
  7. Lekholm U, van Steenberghe D, Herrmann I, Bolender C, Folmer T,Gunne J. Osseointegrated implants in the treatment of partially edentulous jaws: a prospective 5-year multicenter study. Int J Oral Maxillofac Implants 1994;9:627- 635
  8. Babbush CA, Shimura M. Five-year statistical and clinical observations with the IMZ two-stage osteointegrated implant system. Int J Oral Maxillofac Implants 1993;8:245-253
  9. Adell R, Eriksson B, Lekholm U, Branemark PL Jemt T. Long-term follow up study of osseointegrated implants in the treatment of totally edentulous jaws. Int J Oral Maxillofac Implants 1990;5: 347-359
  10. Naert I, Quirynen M, van Steenberghe D, Darius P. A study of 589 consecutive implants supporting complete fixed prostheses. Part II: prosthetic aspects. J Prosthet Dent 1992;68:949-956 https://doi.org/10.1016/0022-3913(92)90557-Q
  11. Jemt T, Lekholm U. Oral implant treatment in posterior partially edentulous jaws: a 5-year followup report. Int J Oral Maxillofac Implants 1993;8:635-640
  12. Jemt T, Linden B, Lekholm U. Failure and complications in 127 consecutively placed fixed partial prostheses supported by Branemark implants: from prosthetic treatment to first annual checkup. Int J Oral Maxillofac Implants 1992;7:40-44
  13. Gunne J, Jemt T, Linden B. Implant treatment in partially edentulous patients: a report on prostheses after 3 years. Int J Prosthodont 1994;7:143-148
  14. van Steenberghe D, Lekholm U, Bolender C, Folmer T, Henry P, Herrmann I, et al. Applicability of osseointegrated oral implants in the rehabilitation of partially edentulism: a prospective multicenter study on 558 fixtures. Int J Oral Maxillofac Implants 1990;5:272-281
  15. Piattelli A, Piattelli M, Scarano A, Montesani L. Light and scanning electron microscopic report of four fractured implants. Int J Oral Maxillofac Implants 1998;13:561-564
  16. Morgan MJ, James DF, Pilliar RM. Fractures of the fixture component of an osseointegrated implant. Int J Oral Maxillofac Implants 1993;8:409-414
  17. McGlumphy EA, Robinson DM, Mendel DA. Implant superstructures: a comparison of ultimate failure force. Int J Oral Maxillofac Implants 1992;7:35-39
  18. Balfour A, O'Brien GR. Comparative study of antirotational single tooth abutments. J Prosthet Dent 1995;73:36-43 https://doi.org/10.1016/S0022-3913(05)80270-7
  19. Wiskott HW, Belser UC. Lack of integration of smooth titanium surfaces: a working hypothesis based on strains generated in the surrounding bone. Clin Oral Implants Res 1999;10:429-444 https://doi.org/10.1034/j.1600-0501.1999.100601.x
  20. Meijer HJ, Starmans FJ, Steen WH, Bosman F. Loading conditions of endosseous implants in an edentulous human mandible: a three-dimensional finite element study. J Oral Rehabil 1996;23:757-763 https://doi.org/10.1046/j.1365-2842.1996.d01-185.x
  21. Quirynen M, Naert I, van Steenbergh D. Fixture design and overload influence marginal bone loss and fixture success in the Branemark system. Clin Oral Implants Res 1992;3:104-111 https://doi.org/10.1034/j.1600-0501.1992.030302.x
  22. Burguete RL, Johns RB, King T, Patterson EA. Tightening characteristics for screwed joints in osseointegrated dental implant. J Prosthet Dent 1994;71:592-599 https://doi.org/10.1016/0022-3913(94)90443-X
  23. Patterson EA, Johns RB. Theoretical analysis of the fatigue life of fixture screws in osseointegrated dental implants. Int J Oral Maxillofac Implants 1992;7:26-33
  24. Larson F, Bland B, Beaty K, Heylmun T. Relation between titanium alloy properties and dental implant fatigue strength. J Dent Res 1995;74(special issue):112
  25. Tagger GN, Machtei EE, Horowitz J, Peled M. Fracture of dental implants: Literature review and report of a case. Implant Dent 2002;11:137-143 https://doi.org/10.1097/00008505-200204000-00014
  26. Chun-Li Lin, Jen-Chyan Wang. Nonlinear finite element analysis of a splinted implant with various connectors and occlusal forces. Int J Oral Maxillofac Implants 2003;18:331-340
  27. Merz BR, Hunenbart S, Belser UC. Mechanics of the implant-abutment connection: an 8-degree taper compared to a butt joint connection. Int J Oral Maxillofac Implants 2000;15:519-526
  28. Tan KB, Nicholls JI. Implant-abutment screw joint preload of 7 hex-top abutment systems. Int J Oral Maxillofac Implants 2001;16:367-377
  29. Griffith HT. Suggested tightening torques for structural bolts. Fastener Technology / February 1987
  30. Lang LA, Kang B, Wang RF, Lang BR. Finite element analysis to determine implant preload. J Prosthet Dent 2003;90:539-546 https://doi.org/10.1016/j.prosdent.2003.09.012
  31. Martin WC, Woody RD, Miller BH, Miller AW. Implant abutment screw rotations and preloads for four different screw materials and surfaces. J Prosthet Dent 2001;86:24-32 https://doi.org/10.1067/mpr.2001.116230
  32. Mericske-Stem R, Piotti M, Sirtes G. 3-D in vivo force measurements on mandibular implants supporting overdentures: a comparative study. Clin Oral Implants Res 1996;7:387-396 https://doi.org/10.1034/j.1600-0501.1996.070413.x
  33. Kikuchi M, Takada Y, Kiyosue S, Yoda M, Woldu M, Cai Z et al. Mechanical properties and microstructures of cast Ti-Cu alloys. Dent Mater 2003;19:174-181 https://doi.org/10.1016/S0109-5641(02)00027-1
  34. Zavanelli RA, Pessanha Henriques GE, Ferreira I, De Almeida Rollo JM. Corrosion-fatigue life of commercially pure titanium and Ti-6Al-4V alloys in different storage environments. J Prosthet Dent 2000;84:274-279 https://doi.org/10.1067/mpr.2000.108758
  35. Cantwell A, Hobkirk JA. Preload loss in gold prosthesis-retaining screws as a function of time. Int J Oral Maxillofac Implants 2004;19:124-132
  36. Yokoyama K, Ichikawa T, Murakami H, Miyamoto Y, Asaoka K. Fracture mechanisms of retrieved titanium screw thread in dental implant. Biomaterials 2002;23:2459-2465 https://doi.org/10.1016/S0142-9612(01)00380-5
  37. Bickford JH. An introduction to the design and behavior of bolted joints. New York: Marcel Decker, 1981
  38. Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-536
  39. Akpinar I, Demirel F, Parnas L, Sahin S. A comparison of stress and strain distribution chracteristics of two different rigid implant design for distal extension fixed prosthesis. Quintessence Int 1996;27:11-17
  40. Holmgren EP, Seckinger RJ, Kilgren LM, Mante F. Evaluating parameters of osseointegrated dental implants using finite element analysis a two dimensional comparative study examining the effects of implant diameter, implant shape, and load direction. J Oral Implantol 1998;24:80-88 https://doi.org/10.1563/1548-1336(1998)024<0080:EPOODI>2.3.CO;2
  41. Sethi A, Kaus T, Sochor P. The use of angulated abutments in implant dentistry: five year clinical results of an ongoing prospective study. Int J Oral Maxillofac Implants 2000;15:801-810
  42. Himmlova L, Dostalova T, Kacovsky A Konvickova S. Influence of implant length and diameter on stress distribution: A finite element analysis. J Prosthet Dent 2004;91:20-25 https://doi.org/10.1016/j.prosdent.2003.08.008
  43. Davarpanah M, Martinez H, Tecucianu JF. Apicalcoronal implant position: recent surgical proposals. technical note. Int J Oral Maxillofac Implants 2000;15:865-872
  44. Meijer HJ, Starmans FJ, Steen WH, Bosman F. Loading conditions of endosseous implants in an edentulous human mandible: a three dimensional finite element study. J Oral Rehabil 1996;23:757-763 https://doi.org/10.1046/j.1365-2842.1996.d01-185.x
  45. Holmes DC, Loftus JT, Influence of bone quality on stress distribution for endosseous implants. J Oral Implantol 1997;23:104-111
  46. Smith EL, Raab DM. Osteoporosis and physical activity. Acta Med Scand 1996; 711(Suppl):149-156
  47. Papavasiliou G, Kamoisiora P, Bayne SC, Felton DA. Three-dimensional finite element analysis of stressdistribution around single tooth implants as a function of bony support, prosthesis type, and loading during function. J Prosthet Dent 1996;76:633-640 https://doi.org/10.1016/S0022-3913(96)90442-4
  48. Jung YC, Han CH, Lee KW. A 1-year radiographic evaluation of marginal bone around dental implants. Int J Oral Maxi-llofac Implants 1996;11:811-818
  49. Bragger U. Use of radiographs in evaluating success, stability and failure in implant dentistry. Periodontol 2000 1998;17:77-88 https://doi.org/10.1111/j.1600-0757.1998.tb00125.x
  50. Cochran DL, Nummikoski PV, Higginbottom FL, Hermann JS, Makins SR, Buser D. Evaluation of an endosseous titanium implant with a sandblasted and acid-etched surface in the canine mandible: radiographic results. Clin Oral Implants Res 1996;7:240-252 https://doi.org/10.1034/j.1600-0501.1996.070306.x
  51. Kitamura E, Stegaroiu R, Nomura S, Miyakawa O. Biomechanical aspects of marginal bone resorption around osseointegrated implants: considerations based on a three-dimensional finite element analysis. Clin Oral Implants Res 2004;15:401-412 https://doi.org/10.1111/j.1600-0501.2004.01022.x
  52. Tada S, Stegaroiu R, Kitamura E, Miyakawa O, Kusakari H. Influence of implant design and bone quality on stress / strain distribution in bone around impla-nts: a 3-dimensional finite element analysis. Int J Oral Maxillofac Implants 2003;18:357-368
  53. Van Oosterwyck H, Duyck J, Vander Sloten J, Van der Perre G, De Cooman M, Lievens S et al. The influence of bone mechanical properties and implant fixation upon bone loading around oral implants. Clin Oral Implants Res 1998;9:407-418 https://doi.org/10.1034/j.1600-0501.1996.090606.x
  54. Frost HM. Wolff's Law and bone's structural adaptations to mechanical usage: an overview for clinicians. Angle Orthod 1994; 64:175-188
  55. Albrektsson T, Sennerby L. State of the art in oral implants. J Clin Periodontol 1991;18:474-481 https://doi.org/10.1111/j.1600-051X.1991.tb02319.x
  56. Duyck J, Ronold HJ, Van Oosterwyck H, Naert I, Vander SJ, Ellingsen JE. The influence of static and dynamic loading on marginal bone reactions around osseointegrated implants: an animal experimental study. 2001; 12:207-218 https://doi.org/10.1034/j.1600-0501.2001.012003207.x
  57. Quirynen M, Naert I, van Steenberghe D, Nys L. A study of 589 consecutive implants supporting complete fixed prostheses. Part I: periodontal aspects. J Prosthet Dent 1992;68:655-663 https://doi.org/10.1016/0022-3913(92)90383-L
  58. Hansson S. The implant neck: smooth or provided with retention elements. A biomechanical approach. Clin Oral Implants Res 1999;10:394-405 https://doi.org/10.1034/j.1600-0501.1999.100506.x
  59. Lindquist LW, Carlsson GE, Jemt T. A prospective 15 year follow up study of mandibular fixed prostheses supported by osseointegrated implants. Clinical results and marginal bone loss. Clin Oral Implants Res 1996;7:329-336 https://doi.org/10.1034/j.1600-0501.1996.070405.x
  60. Lindquists LW, Rockler B, Carlsson GE. Bone resorption around fixtures in edentulous patients treated with mandibular fixed tissue integrated prostheses. J Prosthet Dent 1988;59:59-63
  61. Geng JP, Tan KB, Liu GR. Application of finite element analysis in implant dentistry: A review of the literature. J Prosthet Dent 2001;85:585-598 https://doi.org/10.1067/mpr.2001.115251