Evaluation of the Positional Accuracy of the Delivered Beams from the Target: A Phantom Study
방사선 치료에서 치료 표적과 조사 빔의 일치 정도 평가: 팬텀 연구
- Kang, Sei-Kwon (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Cho, Byung-Chul (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Cheong, Kwang-Ho (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Ju, Ra-Hyeong (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Kim, Su-Ssan (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Kim, Kyoung-Ju (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Choi, Sang-Gyu (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Bae, Hoon-Sik (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Lee, Re-Na (Department of Radiation Oncology, Ewha Womans University College of Medicine) ;
- Oh, Do-Hoon (Department of Radiation Oncology, College of Medicine, Hallym University)
- 강세권 (한림대학교 의과대학 방사선종양학교실) ;
- 조병철 (한림대학교 의과대학 방사선종양학교실) ;
- 정광호 (한림대학교 의과대학 방사선종양학교실) ;
- 주라헝 (한림대학교 의과대학 방사선종양학교실) ;
- 김수산 (한림대학교 의과대학 방사선종양학교실) ;
- 김경주 (한림대학교 의과대학 방사선종양학교실) ;
- 최상규 (한림대학교 의과대학 방사선종양학교실) ;
- 배훈식 (한림대학교 의과대학 방사선종양학교실) ;
- 이레나 (이화여자대학교 의과대학 방사선종양학교실) ;
- 오도훈 (한림대학교 의과대학 방사선종양학교실)
- Published : 2006.12.31
Abstract
We evaluated the positional accuracy of the delivered beams to the target in a phantom by simulating the whole process of the radiation treatments Including CT scanning, planning and beam exposures with MLCs. For this purpose, a phantom was made to calibrate the alignment between the CT and the attached laser system. A new, convenient method was also devised to align the setup lasers in the treatment room. Film was used for the Identification of the delivered beam and analyzed with a homemade computer program. The positional differences between the target and the beam centers varied with the couch rotations. The accelerator we used showed a maximum discrepancy of 2.0 mm at the table angle of
CT와 치료계획시스템 및 다엽콜리메이터(Multi Leaf Collimator, MLC)를 이용한 방사선 치료의 전 과정을 팬텀에 대해 재현해 봄으로써, 치료계획상의 표적과 조사빔의 일치 정도를 평가해 보았다. 이를 위해, CT 및 그 레이저 정렬을 위한 팬텀을 제작하였으며, 치료실의 레이저 정렬을 손쉽게 할 수 있는 방법을 고안하였다. 조사빔의 확인에는 필름을 이용하였고, 조사빔과 실제 표적위치 사이의 정량적인 분석은 자체 제작한 프로그램을 이용하였다. 표적과 빔 중심의 불일치 정도는 테이블 각도에 따라 차이가 있었으며 시험에 이용된 치료기의 경우, 테이블 각도 에서 최대 2.0 mm의 차이를 보였다. MLC field에 대해 Winston-Lutz 테스트를 시행하여