Evaluation of the Positional Accuracy of the Delivered Beams from the Target: A Phantom Study

방사선 치료에서 치료 표적과 조사 빔의 일치 정도 평가: 팬텀 연구

  • Kang, Sei-Kwon (Department of Radiation Oncology, College of Medicine, Hallym University) ;
  • Cho, Byung-Chul (Department of Radiation Oncology, College of Medicine, Hallym University) ;
  • Cheong, Kwang-Ho (Department of Radiation Oncology, College of Medicine, Hallym University) ;
  • Ju, Ra-Hyeong (Department of Radiation Oncology, College of Medicine, Hallym University) ;
  • Kim, Su-Ssan (Department of Radiation Oncology, College of Medicine, Hallym University) ;
  • Kim, Kyoung-Ju (Department of Radiation Oncology, College of Medicine, Hallym University) ;
  • Choi, Sang-Gyu (Department of Radiation Oncology, College of Medicine, Hallym University) ;
  • Bae, Hoon-Sik (Department of Radiation Oncology, College of Medicine, Hallym University) ;
  • Lee, Re-Na (Department of Radiation Oncology, Ewha Womans University College of Medicine) ;
  • Oh, Do-Hoon (Department of Radiation Oncology, College of Medicine, Hallym University)
  • 강세권 (한림대학교 의과대학 방사선종양학교실) ;
  • 조병철 (한림대학교 의과대학 방사선종양학교실) ;
  • 정광호 (한림대학교 의과대학 방사선종양학교실) ;
  • 주라헝 (한림대학교 의과대학 방사선종양학교실) ;
  • 김수산 (한림대학교 의과대학 방사선종양학교실) ;
  • 김경주 (한림대학교 의과대학 방사선종양학교실) ;
  • 최상규 (한림대학교 의과대학 방사선종양학교실) ;
  • 배훈식 (한림대학교 의과대학 방사선종양학교실) ;
  • 이레나 (이화여자대학교 의과대학 방사선종양학교실) ;
  • 오도훈 (한림대학교 의과대학 방사선종양학교실)
  • Published : 2006.12.31

Abstract

We evaluated the positional accuracy of the delivered beams to the target in a phantom by simulating the whole process of the radiation treatments Including CT scanning, planning and beam exposures with MLCs. For this purpose, a phantom was made to calibrate the alignment between the CT and the attached laser system. A new, convenient method was also devised to align the setup lasers in the treatment room. Film was used for the Identification of the delivered beam and analyzed with a homemade computer program. The positional differences between the target and the beam centers varied with the couch rotations. The accelerator we used showed a maximum discrepancy of 2.0 mm at the table angle of $295^{\circ}$. The same measurements based on the new isocenter from the Winston-Lutz test resulted in the maximum of 1.35 mm for all rotation angles. The evaluation of the differences between the target and the beam centers is useful for the treatment planning.

CT와 치료계획시스템 및 다엽콜리메이터(Multi Leaf Collimator, MLC)를 이용한 방사선 치료의 전 과정을 팬텀에 대해 재현해 봄으로써, 치료계획상의 표적과 조사빔의 일치 정도를 평가해 보았다. 이를 위해, CT 및 그 레이저 정렬을 위한 팬텀을 제작하였으며, 치료실의 레이저 정렬을 손쉽게 할 수 있는 방법을 고안하였다. 조사빔의 확인에는 필름을 이용하였고, 조사빔과 실제 표적위치 사이의 정량적인 분석은 자체 제작한 프로그램을 이용하였다. 표적과 빔 중심의 불일치 정도는 테이블 각도에 따라 차이가 있었으며 시험에 이용된 치료기의 경우, 테이블 각도 에서 최대 2.0 mm의 차이를 보였다. MLC field에 대해 Winston-Lutz 테스트를 시행하여 $295^{\circ}$얻은 isocenter를 기준으로 다시 시험한 결과는 측정한 모든 테이블 각도에서 1.35 mm 이하였다. 표적과 빔 중심의 차이에 관한 이러한 평가는 실제 환자의 치료 계획 설정에 유용하게 이용될 수 있다.