Abstract
The precise measurement of ultrasonic power is important to the qualify assurance and the safety of medical ultrasonic equipments In the Present work, a brier theory was introduced to determine the ultrasonic Power from the time valving balance-indication due to the radiation force acting on an absorbing target and/or other causes such as buoyancy during the repetition of on/off behavior of ultrasonic irradiation. The developed automated system measuring the ultrasonic power was described in detail with the precise mechanical alignment tool , the electric signal generation network, the control and measurement network and the appropriate procedure. The ultrasonic power measured by the developed system was compared to the reference data calibrated by the other national metrology institute at 1 MHz, 5 MHz, 10 MHz, and 15 MHz over the range 10 mW to 10 W. Their relative differences are within $5\%$.
초음파 파워의 정확한 측정은 초음파 의료 장비의 품질 보증과 안전성에 중요하다. 본 논문에서는 초음파 조사가 on/off를 반복하는 동안의 표적에 작용하는 방사 힘과 부력과 같은 다른 원인에서 기인하는 저울 지시 값의 시간에 따른 변화로부터 초음파 파워를 결정하는 간단한 이론을 제시하였다. 아울러 개발된 초음파 파워 측정 시스템에 대해 정밀 기계적 정렬장치, 전기신호 발생부 구성, 제어 및 측정부 구성 및 적절한 절차와 함께 상세히 기술하였다. 본 시스템으로 1 MHz, 5 MHz, 10 MHz, 15 MHz에서 10 mW에서 10 W범위에 걸쳐 측정된 결과를 다른 나라에서 교정된 기준 데이터와 상호 비교한 결과 상대 편차가 $5\%$ 이내로 확인되었다.