• Title/Summary/Keyword: 흡음형 표적

Search Result 1, Processing Time 0.015 seconds

Construction of Ultrasonic Power Measurement System and Its Performance Evaluation (초음파 파워 측정 시스템 구성 및 성능평가)

  • Jho Moon Jae;Kim Yong Tae;Yun Yong Hyeon;Jung Suug Soo
    • The Journal of the Acoustical Society of Korea
    • /
    • v.24 no.8
    • /
    • pp.431-440
    • /
    • 2005
  • The precise measurement of ultrasonic power is important to the qualify assurance and the safety of medical ultrasonic equipments In the Present work, a brier theory was introduced to determine the ultrasonic Power from the time valving balance-indication due to the radiation force acting on an absorbing target and/or other causes such as buoyancy during the repetition of on/off behavior of ultrasonic irradiation. The developed automated system measuring the ultrasonic power was described in detail with the precise mechanical alignment tool , the electric signal generation network, the control and measurement network and the appropriate procedure. The ultrasonic power measured by the developed system was compared to the reference data calibrated by the other national metrology institute at 1 MHz, 5 MHz, 10 MHz, and 15 MHz over the range 10 mW to 10 W. Their relative differences are within $5\%$.