Comparison of Calibration Methods of $^{192}\textrm{Ir}$ Sources for High Dose Rate Brachytherapy

고선량률 근접조사치료용 이리듐-192 방사성동위원소의 교정방법 비교연구

  • 허현도 (인하대학교 의과대학 방사선종양학교실) ;
  • 박성용 (국립암센터 양성자치료센터) ;
  • 이레나 (이화여자대학교 의과대학 방사선종양학교실) ;
  • 신동오 (경희대학교 의과대학 방사선종양학교실) ;
  • 권수일 (경기대학교 의학물리학과) ;
  • 노준규 (인하대학교 의과대학 방사선종양학교실) ;
  • 최진호 (가천의과대학 방사선종양학교실)
  • Published : 2004.12.01

Abstract

The activity of Ir-192 sources for high dose rate (HDR) Brachytherapy in Korea were measured by using the well-type chamber and using the calibration Jig with the Farmer-type ionization chamber to compare the manufacturer certificated source strength which is supplied with each new Ir-192 source. The activity of two different source models used in six hospitals were measured. The range of measured activities to the manufacturer's suggested ones was -2.40% to +3.31% for the calibration Jig and -3.12% to 0.00% for the well-type chamber system. The source strength values given by the manufacturer for the 6 sources were within ${\pm}5%$ for the two different measuring equipment. Our results demonstrate that well-type chamber as wall as Farmer-type chamber system are appropriate system for the routine source calibration procedures in HDR brachytherapy. Whenever a new source is installed to use in clinics, a source calibration should be carried out.

본 연구는 국내에서 이용되고 있는 고선량률 근접치료기에 사용되는 Ir-192 선원에 대하여 Farmer type전리함과 우물형 전리함을 이용하여 선원의 세기를 측정하여 제조사에서 제공된 값과의 오차를 확인하고 두 증기간의 오차를 비교하였다. 동일한 Ir-192가 장착되어 가동 중인 6개의 병원의 고선량률 근접치료기(microSelctron, Nucletron, Netherlands)를 대상으로 선원의 모델별로 구분하여 상품화된 교정 Jig와 우물형 전리함 시스템을 이용, 선원의 세기를 측정하고 제조사에서 제공된 값과 비교하여 오차를 구하였다. 교정용 Jig를 이용한 선원 세기의 오차범위는 -2.40~+3.12%이고 우물형 전리함의 경우 오차범위는 -3.31~0.00%를 나타내었다. 측정 결과 두 장비 모두 제조사에서 제공된 값과 비교하였을 때 ${\pm}5%$의 오차를 넘지 않았으므로 임상 적용에는 문제가 없다. 외국의 사고 사례 보고에 따르면 계산 값에 의존하지 말고 새로운 선원이 교체될때 반드시 치료에 사용되기 전에 측정 도구를 이용하여 방사능의 세기를 측정하고 계산된 값과 비교하여 오차범위 내에서 일치하는지에 대한 확인 절차과정을 거치는 것이 필요하다.

Keywords