Comparison of Calibration Methods of $^{192}\textrm{Ir}$ Sources for High Dose Rate Brachytherapy
고선량률 근접조사치료용 이리듐-192 방사성동위원소의 교정방법 비교연구
Abstract
The activity of Ir-192 sources for high dose rate (HDR) Brachytherapy in Korea were measured by using the well-type chamber and using the calibration Jig with the Farmer-type ionization chamber to compare the manufacturer certificated source strength which is supplied with each new Ir-192 source. The activity of two different source models used in six hospitals were measured. The range of measured activities to the manufacturer's suggested ones was -2.40% to +3.31% for the calibration Jig and -3.12% to 0.00% for the well-type chamber system. The source strength values given by the manufacturer for the 6 sources were within
본 연구는 국내에서 이용되고 있는 고선량률 근접치료기에 사용되는 Ir-192 선원에 대하여 Farmer type전리함과 우물형 전리함을 이용하여 선원의 세기를 측정하여 제조사에서 제공된 값과의 오차를 확인하고 두 증기간의 오차를 비교하였다. 동일한 Ir-192가 장착되어 가동 중인 6개의 병원의 고선량률 근접치료기(microSelctron, Nucletron, Netherlands)를 대상으로 선원의 모델별로 구분하여 상품화된 교정 Jig와 우물형 전리함 시스템을 이용, 선원의 세기를 측정하고 제조사에서 제공된 값과 비교하여 오차를 구하였다. 교정용 Jig를 이용한 선원 세기의 오차범위는 -2.40~+3.12%이고 우물형 전리함의 경우 오차범위는 -3.31~0.00%를 나타내었다. 측정 결과 두 장비 모두 제조사에서 제공된 값과 비교하였을 때