Comparison of Calibration Methods of $^{192}\textrm{Ir}$ Sources for High Dose Rate Brachytherapy (고선량률 근접조사치료용 이리듐-192 방사성동위원소의 교정방법 비교연구)

  • Huh Hyun Do;Park Sung Yong;Lee Rena J;Shin Dong Oh;Kwon Soo Il;Loh John J K;Choi Jinho
    • Progress in Medical Physics
    • /
    • v.15 no.4
    • /
    • pp.192-196
    • /
    • 2004
  • The activity of Ir-192 sources for high dose rate (HDR) Brachytherapy in Korea were measured by using the well-type chamber and using the calibration Jig with the Farmer-type ionization chamber to compare the manufacturer certificated source strength which is supplied with each new Ir-192 source. The activity of two different source models used in six hospitals were measured. The range of measured activities to the manufacturer's suggested ones was -2.40% to +3.31% for the calibration Jig and -3.12% to 0.00% for the well-type chamber system. The source strength values given by the manufacturer for the 6 sources were within ${\pm}5%$ for the two different measuring equipment. Our results demonstrate that well-type chamber as wall as Farmer-type chamber system are appropriate system for the routine source calibration procedures in HDR brachytherapy. Whenever a new source is installed to use in clinics, a source calibration should be carried out.