초록
This study was done to present a criterion in selection of the most proper light sources and materials by measuring metamerism index(MI) of the light curing composite resins with spectrocolorimeter. Metamerism is defined when two objects appear to be the same color in one illuminant but different in another. This is due to the fact that they have different spectral curves that fail to match under the second illuminant. In this study, A1 & A3 shade of five light curing composite resins (Esthet-X, Filteck Z250, Filteck A110. Charisma. Vitalescence) were chosen based on Vita shade. Five samples were made for shade of each product with Teflon mold (diameter: 15mm, thickness: 2mm). Metamerism index of each samples on a Barium sulfate plate($L^{*}=96.54,{\;}a^{*}=0.19,{\;}b^{*}=0.01$) prepared for sample fixation were measured with spectrocolorimeter(Miniscan XE plus. Model 4000s. Hunter Lab. USA) by applying standard light source D^{65}, C. Fcw, TL84 and A. Standardization was done with reference standard(X=80.8, Y=85.7, Z=90.8) and light trap. The results were as follows. 1. Different resins with same Vita shade showed recognizable color difference (${\delta}E^{*}>2$). 2. All composites had MI below accepted value 0.5 between standard illuminant(D$_{65}$, C, & A) and below 1.5 under fluorescent condition (Fcw & TL84). 3. MI value between $D^{65}$ and A showed higher value than MI value between other source of light(p<0.01). 4. All resins except Z250 showed MI value that A3 is higher than A1 between $D^{65}$ and A(p<0.05).except Z250 showed MI value that A3 is higher than A1 between $D^{65}$ and A(p<0.05).