An Experimental Study on Installation Effects of Pipe Elbow on the Electromagnetic Flowmeter Characteristics (Turbulent Flow)

곡관의 하류에 설치된 전자기유량계의 유량신호 특성에 관한 실험적 연구(난류 유동)

  • 임기원 (한국표준과학연구원 유체유동그룹)
  • Published : 2002.11.01

Abstract

An electromagnetic flowmeter(EMF) essentially averages the velocity distribution over the pipe cross- sectional area, and the measured value is dependent on the velocity profiles. In this study, installation effects of 90$^{\circ}$long elbow(KS B 1522, ISO 3419) on the EMF characteristics was investigated. A commercial EMF was adopted and the distribution of magnetic field in the electrodes cross section was measured. In the experiment, the national flow standard system, of which measurement uncertainty was evaluated in accordance with ISO 17025 recommendation, was used fur characterization of EMF. The leading line has 150D long straight pipe to established a fully developed flow before entering into the elbow and the elbow was installed downstream of it. then the flowmeter was tested within 50 D from the elbow. The installation effects of the flowmeter were investigated by varying the mean velocity(Reynolds No.)in pipe section, the locations and the direction of electrodes plane.($\phi$) From the experimental results, we find the optimal conditions to get most accurate measurements. Generally, the deviations from the calibration value were less than 0.5 % in farther than 10D distance from the elbow and the direction of electrode plane. $\phi$ = 90$^{\circ}$yielded the smallest measurement deviation. These characteristics were shown consistently in turbulent region regardless of the mean Reynolds number.

Keywords

References

  1. Shercliff J.A., 1954, 'Relation Between the Velocity Profile and the Sensitivity of Electromagnetic Flowmeters,' J. Appl. Phys., Vol. 25, pp. 817-818 https://doi.org/10.1063/1.1721749
  2. Bevir M.K., 1970, 'The Theory of Induced Voltage Electromagnetic Flowmeters,' J. Fluid Mech., Vol. 43, pp. 577-590 https://doi.org/10.1017/S0022112070002586
  3. Engl, W.L., 1970, 'Der Induktive Druchflussmesser mit Inhomogenem Magnetfeld,' Arch. fur Electrotechn., Vol. 5, No. 6, pp. 126-134
  4. Luntta E. and Halttunen J., 1989, 'The Effect of Velocity Profile on Magnetic Flow Measurement,' Sensors and Actuators, Vol. 16, pp. 335-344 https://doi.org/10.1016/0250-6874(89)85004-8
  5. Kuromori, K., 1993, 'An Electromagnetic Flowmeter with Plural Pairs of Electrodes (in Japanese),' J. of the Society of Instrument and Control Engineers, Vol. 29, pp. 1413-1420 https://doi.org/10.9746/sicetr1965.29.1413
  6. JIS B 7554, 1993, 'Electromagnetic Flowmeters(in Japanese).'
  7. Bevir M. K., O' Sullivan V. T. & Wyatt D. G., 1981, 'Computation of Electromagnetic Flowmeter Characteristics from Magnetic Field Data,' J.Phys. D:Appl. Phys., Vol. 14, pp. 373-388 https://doi.org/10.1088/0022-3727/14/3/007
  8. ISO 4185, 1980, 'Measurement of Liquid Flow in Closed Conduits-Weighing methods.'
  9. ISO, 1993(E), 'Guide to The Expression of Uncertainty in Measurement(1st edition).'
  10. Ki Won Lim, 2001, 'A Study on the Measurement Uncertainty of Flowmeter Calibrator,' J. of KSME(B), Vol. 25, No. 4, pp. 561-571
  11. Ki Won Lim, 1994, 'Standard Calibration Procedure of an Electromagnetic Flowmeter,' KASTO-94-10-408-107
  12. Ki Won Lim, Myung Kyoon Chung, 1999, 'Numerical Investigation on the Installation Effects of an Electromagnetic Flowmeter Downstream of a 90° Elbow-Laminar Flow Case,' Flow Measurement and Instrumentation, Vol. 10, pp. 167-174 https://doi.org/10.1016/S0955-5986(98)00052-1

Cited by

  1. Flow Signal Characteristics of Small Scale Electromagnetic Flowmeter in Low Conductivity Fluid Measurement vol.40, pp.9, 2016, https://doi.org/10.3795/KSME-B.2016.40.9.613