Comparative measurements of Criteria Pollutants Using DOAS and Conventional In-situ Monitoring Technique at Sung Nam city of Korea

  • Kim, Ki-Hyun (Dept. of Earth Sciences, Sejong University) ;
  • Jin, Byong-Bok (Environmental Instrument inspection Team, Environmental Management Corporation)
  • 발행 : 2001.12.01

초록

To test the compatibility of differential optical absorption spectroscopy (DOAS) and conventional in-situ monitoring technique we conducted a comparative analysis of the two systems using hourly measurement data sets of three criteria pollutants including No$_{2}$O$_{3}$, and SO$_{2}$ collected in months between April and June of 2001 at Sung Man city, Kyung Gi Province, Korea. The results of our comparative analysis were useful to evaluate the various aspects of DOAS performance, of particular the level of agreement with the counterpart method through computation of percent differences and correlation analysis. Interpretation of the mixing ratio data for those chemical species was however confined in terms of explaining the differences affected by the changes in environmental conditions because measurements of important meteorological parameters were limited during most of the study period. Nevertheless, the overall results of this study strongly demonstrated that the mixing ratio of major pollutants measured by the two different systems maintain strong compatibility from various respects.

키워드