THE EFFECTS OF FABRICATION AND TIGHTENING METHODS OF GOLD CYLINDER ON THE PRELOAD OF THE IMPLANT ABUTMENT

금 실린더의 제작법과 고정 방법이 임플랜트 지대주의 preload에 미치는 영향

  • Cho, Hye-Won (Department of Prosthodontics, College of Dentistry, Wonkwang University) ;
  • Kim, Sung-Hoon (Wonkwang Dental Research Institute)
  • 조혜원 (원광대학교 치과대학 치과보철학교실) ;
  • 김성훈 (원광치의학연구소)
  • Published : 2000.10.01

Abstract

The purpose of this study was to evaluate the effects of fabrication and tightening methods of gold cyliner on the preload of the standard abutment. Three linear strain gauges (KFR-02N-120-C1-23, Kyowa, Japan) were mounted longitudinally on the 5.5mm Standard abutment (Nobel Biocare, Sweden) and three kinds of gold cylinders such as, as-received gold cylinder, gold cylinder after casting, and plastic cylinder after casting with type IV gold alloy were connected over the top of the standard abutment. Two kinds of tightening methods, such as manual torque with handhold screwdriver and electronic torque using Electronic torque controller were used to generate preload on the abutment. The result were as follows; 1. The preload generated by tightening cast plastic cylinder with handhold screw driver, was the lowest among the six groups. 2. The preload generated by cast plastic cylinder was lower than those by gold cylinders regardless of the tightening methods. 3. The electronic torque controller produced higher torque values than the handheld screwdriver.