A Cosideration on Physical Aspects in Teleradiotherapy Chart QA

원격방사선치료 기록부의 QA 에서 물리적 측면의 고찰

  • 강위생 (서울대학교 의과대학 치료방사선과학교실) ;
  • 허순녕 (서울대학교 의학연구원 방사선의학연구소)
  • Published : 1999.06.01

Abstract

The aims of this report are to classify the incorrect data of patients and the errors of dose and dose distribution observed in QA activities on teleradiotherapy chart, and to analyze their frequency. In our department, radiation physicists check several sheets of patient chart to reduce numeric errors before starting radiation therapy and at least once a week, which include history, port diagram, MU calculation or treatment planning summary and daily treatment sheet. The observed errors are classified as followings. 1) Identity of patient, 2) Omitted or unrecorded history sheet even though not including the item related to dose, 3) Omission of port diagram, or omitted or erroneous data, 4) Erroneous calculation of MU and point dose, and important causes, 5) Loss of summary sheet of treatment planning, and erroneous data of patient in the sheet, 6) Erroneous record of radiation therapy, and errors of daily dose, port setup, MU and accumulated dose in the daily treatment sheet, 7) Errors leading inexact dose or dose distribution, errors not administerd even though its possibility, and simply recorded errors, 8) Omission of sign. Number of errors was counted rather than the number of patients. In radiotherapy chart QA from Jun 17, 1996 to Jul 31, 1999, no error of patient identity had been observed. 431 Errors in 399 patient charts had been observed and there were 405 physical errors, 9 cases of omitted or unrecorded history sheet, and 17 unsigned. There were 23 cases (5.7%) of omitted port diagram, 21 cases (5.2%) of omitted data and 73 cases (18.0 %) of erroneous data in port diagram, 13 cases (3.2 %) treated without MU calculation, 68 cases (16.3 %) of erroneous MU, 8 cases (2.0%) of erroneous point dose, 1 case (0.2 %) of omitted treatment planning summary, 11 cases (2.7%) of erroneous input of patient data, 13 cases (3.2%) of uncorrected record of treatment, 20 cases (4.9%) of discordant daily doses in MU calculation sheet and daily treatment sheet, 33 cases (8.1%) of erroneous setup, 52 cases (12.8%) of MU setting error, 61 cases (15.1%) of erroneous accumulated dose. Cases of error leading inexact dose or dose distribution were 239 (59.0 %), cases of error not administered even though its possibility were 142 (35.1 %), and cases of simply recorded error were 24 (5.9 %). The numeric errors observed in radiotherapy chart ranged over various items. Because errors observed can actually contribute to erroneous dose or dose distribution, or have the possibility to lead such errors, thorough QA activity in physical aspects of radiotherapy charts is required.

원격방사선치료 기록부의 QA 과정에서 관찰된 환자에 대한 부정확한 자료나 방사선치료 기록부에 기록이나 이기의 오류로 인한 선량이나 선량분포의 오류의 종류와 빈도를 분석하는 것이다. 서울대학교병원 치료방사선과에서는 수치 오류을 시정하기 위해 의학물리학자가 치료개시전과 일주일에 일회 이상 방사선치료 기록부의 병록지와 배치도면, MU 계산용지나 치료계획 요약지, 일일 치료기록지를 점검하고 있다. 관찰된 오류를 다음과 같이 분류하였다. 1) 환자신원 확인, 2) 물리적인 요소를 포함하지 않지만 병록지의 누락이나 미기재, 3) 배치도면의 누락이나 setup 에 필요한 자료의 누락과 착오, 4) MU와 점선량 계산용지에서는 MU의 오류의 중요 원인별로 구분 및 점선량의 오류, 5) 치료계획 요약지의 분실여부와 환자자료의 오류, 6) 일일 치료기록지에서는 치료사실의 기재누락, 치료일정 착오, 처방선량 착오, setup 착오, MU 착오, 누적선량착오, 7) 선량이나 선량분포의 부정확성을 초래한 오류와 그런 가능성은 있지만 실제 실현되지 않은 오류, 단순히 기록상의 오류 , 8) 서명의 누락에 대해 검사하였으며, 결과는 환자의 수 대신 오류 건수별로 분석하였다. 1996년 6월 17일부터 1999년 7 월 31 일까지 방사선치료 기록부의 QA 에서 환자신원의 불일치는 한건도 없었으며 ,399명의 환자에 대해 431건의 오류가 관찰되었다. 물리적인 오류는 405건, 병록지의 누락 또는 미기재가 9 건, 서명누락이 17건이었다. 배치도면이 없는 경우 23 건 (5.7%), 자료의 누락 21건 (5.2%), 자료의 오기 73건(18.0%) 이었으며, MU의 계산용지가 없거나 계산 없이 치료가 시행된 경우는 13건 (3.2%), MU 계산의 착오 68 건 (16.3%), 점선량의 계산착오 8 건(2.0%), 방사선치료계획 결과 용지의 분실이 1건 (0.2%), 환자자료의 입력 오류가 11건(2.7%), 치료기록의 누락이나 치료의 누락이 8건(2.0%), 치료일정의 오류 13건 (3.2%), MU 계산이나 치료계획의 처방선량과 일일 치료기록지 처방선량의 불일치 20건(4.9%), 치료 setup 의 착오 33건(8.1%), MU의 설정 착오 52건(12.8%), 누적선량 착오 61건(15.1%) 이었다. 선량이나 선량분포의 부정확성을 초래한 오류는 239건(59.0%) 이었으며, 그런 가능성은 있지만 실제 실현되지 않은 오류 142건(35.1%), 단순히 기록상 오류는 24건 (5.9%) 이었다. 관찰된 수치 오류는 다양한 분야에 걸쳐 있었다. 나타난 대부분의 오류는 선량이나 선량분포의 오류에 직접 기여하거나 기여할 우려가 있기 때문에 방사선치료 기록부에서 물리적인 면의 QA를 철저하게 할 필요가 있다.

Keywords