The Study of Shielding Effect on Ovoids of Three Different Gynecological Applicator Sets in microSelectron-HDR System
microSelectron-HDR System에서 부인암 강내조사에 쓰이는 세 가지 Applicator Set들의 Ovoids에 대한 차폐효과 연구
- Cho, Young-K. (Department of Radiation Oncology, Inha University Medical College, Inha University Hospital) ;
- Park, Sung-Y. (Department of Radiation Oncology, Inha University Medical College, Inha University Hospital) ;
- Choi, Jin-H. (Department of Radiation Oncology, Gachon Medical College, Ghil Medical Center) ;
- Kim, Hung-J. (Department of Radiation Oncology, Inha University Medical College, Inha University Hospital) ;
- Kim, Woo-C. (Department of Radiation Oncology, Inha University Medical College, Inha University Hospital) ;
- Loh, John-J.K. (Department of Radiation Oncology, Inha University Medical College, Inha University Hospital) ;
- Kim, Joo-Y. (Department of Radiation Oncology, Gachon Medical College, Ghil Medical Center)
- 조영갑 (인하대병원 방사선종양학과) ;
- 박성용 (인하대병원 방사선종양학과) ;
- 최진호 (길병원 방사선종양학과) ;
- 김헌정 (인하대병원 방사선종양학과) ;
- 김우철 (인하대병원 방사선종양학과) ;
- 노준규 (인하대병원 방사선종양학과) ;
- 김주영 (길병원 방사선종양학과)
- Published : 1998.12.30
Abstract
There are three different types of gynecological applicator sets available in microSelectron-high dose-rate(HDR) System by Nucletron; standard applicator set(SAS), standard shielded applicator set(SSAS), and Fletcher-Williamson applicator set(FWAS). Shielding effect of a SAS without shielding material was compared with that of a SSAS with shielding material made of stainless steel(density
Nucletron사의 microSelectron-High Dose-Rate(HDR) System에서 부인암의 강내 근접치료에 사용되는 Standard shielded applicator set(SSAS)와 Fletcher-Williamson applicator set(FWAS)는 standard applicator set(SAS)와 비교해서 다른 점이 ovoids 위와 아래 부분에 방광과 직장으로 가는 선량을 줄이기 위해 각각 스테인레스 강철(밀도
Keywords