This paper proposes a high precision measurement technique to obtain the height of gage block with a few millimeter height. The proposed technique is consisted of two steps : In the first step, laser position transducer and electric micrometer are adopted to obtain a coarse value of the height of gage block, and then, in the second step, heterodyne laser interferometry is adopted to acquire the precision value. A new kind of phase detector is constructed in the low cost for the heterodyne interferometer and its linearity with