Determination of Thioglycolic acid in the presence of Copper(II) by Adsorptive Stripping Voltammetry

흡착 벗김 전압전류법에 의한 구리이온(II) 존재하에서 티오글리콜산의 정량

  • 홍미정 (성심여자대학교 화학과) ;
  • 권영순 (성심여자대학교 화학과)
  • Received : 1994.12.05
  • Published : 1995.03.25

Abstract

Determination method of trace thioglycolate has been studied by adsorptive stripping voltammetry. Copper(II)-thioglycolate complex is adsorbed at the hanging mercury drop electrode and stripped during cathodic scan. Electrolyte was used pH 6.5 phosphate and pH 9.5 borate buffer solutions. Optimal conditions were a copper(II) concentration $1{\times}10^{-4}M$, an adsorption accumulation potential -0.2V, an adsorption accumulation time 60 sec and a scan rate 20mV/sec. A detection limit of $1{\times}10^{-9}M$ thioglycolate was obtained. The method was applied to the determination of thioglycolate in cold wave fluids and depilating creams.

구리이온(II)과 티오글리콜산의 착물인 CuRS를 HMDE(Hanging Mercury Drop Electrode)에 흡착시켜 음극으로 주사하는 흡착벗김 전압전류법으로 미량의 티오글리콜산을 정량하는 방법을 고찰하였다. 지지 전해질로는 pH 6.5 인산염 완충용액과 pH 9.5 붕산염 완충용액을 사용하였고, 최적 조건은 구리이온(II) 농도 $1{\times}10^{-4}M$, 흡착 축적 전위 -0.2V, 흡착 축적 시간 60초, 그리고 주사 속도 20mV/sec이다. 이 때 티오글리콜산의 검출 한계는 $1{\times}10^{-9}M$이다. 이 조건하에서 콜드 파마액과 제모제 속에 포함된 티오글리콜산의 함량을 구하였다.

Keywords

Acknowledgement

Supported by : 성심여자대학교