Numerical Study of Effect of counter-pulsation on Hemodynamic Response in the ECLS

체외생명구조장치에서 역박동 방법이 혈류역학 응답에 미치는 영향에 대한 수치적 연구

  • 김인수 (강원대학교 기계메카트로닉스공학과 대학원) ;
  • 임기무 (강원대학교 기계메카트로닉스공학과 대학원) ;
  • 최성욱 (강원대학교 기계메카트로닉스공학과 대학원) ;
  • 전형민 (강원대학교 기계메카트로닉스공학과 대학원) ;
  • 심은보 (강원대학교 기계메카트로닉스공학과)
  • Published : 2008.11.05

Abstract

Extra-corporeal Life Support System (ECLS) is the device used in emergency cases to substitute a extracorporeal circulation in open heart surgery, cardiac arrest or in acute cardiopulmonary failure. To obtain the effect of counter-pulsation on hemodynamic response in the ECLS quantitatively, we developed cardiovascular model which consists of 12 compartment model of heldt et al. and 3 compartment model of Schreiner et al. based on windkessel approximation. We compared coronary perfusion, arterial pulse pressure, cardiac output, and left ventricular pressure-volume diagram according to flow configuration such as counter-pulsation, copulsation, and continous flow. When counter-pulsation was applied, 5% higher coronary perfusion, 26% lower pulse pressure, and 2% higher cardiac output than copulsation condition were calculated. We conclude that counter-pulsation configuration in the ECLS is hemodynamically more stable than copulsation and influences the positive effect to recover ventricles.

Keywords