X-선 스펙트럼 변조 기술 연구

X-Ray Spectrum Modulation for Mammography

  • 김광현 (고려대학교 의공학협동과정) ;
  • 김경락 (고려대학교 전자및정보공학부) ;
  • 오창현 (고려대학교 의공학협동과정)
  • 발행 : 2003.11.21

초록

Energy spectrum modulation of X-ray source in digital mammography has been studied. In this study, we calculated various filtered spectra using the scattering data. Primary spectra were generated by Molybdenum (Mo) and Tungsten (W) targets. The materials of added filters are Molybdenum and Rhodium (Rh) for 40 kVp Mo. primary spectrum, the amounts of photons over whole energy ranges are attenuated to 0.43 with 0.03 mm Mo filter and 0.38 with 0.06 mm Mo filter while the photons of energy ranged from 17 keV to 20 keV. The photons of low energy ranged below 17 keV are considerably attenuated. This effect brings out reducing the scattered radiation and dose to the patient, and enhancing subject contrast in the image. The results show that filtered spectra are not seriously affected by X-ray tube loadability. Because the energy range from 17 keV to 20 keV is directly transmitted although low and high energies are mainly filtered.

키워드