Measurements of Nonlinearity in homodyne interferometer

Homodyne interferometer의 Non I inear ity 측정

  • Published : 2000.11.01

Abstract

Nonlinearity is one of the primary causes of error in precision length measurement using laser interferometer. It arises periodically. The periodical nonlinearity usually ranges from sub-naometre to several namertres. In the homodyne interferometer, it results from a number of factors including polarization mixing, imperfect optical clement, unequal gain of detectors, misalignment of axes between input beam and beam splitter. In this paper, we described a method for measuring and compensating the nonlinearity of homodyne interferometer using the elliptical least-square fitting technique associated with electric method and experimental results in one frequency polarization interferometer.

Keywords