Purpose: There has been recent interest in the radioactivity uptake and image acquisition of radioactivity concentration. The degree of uptake is strongly affected by many factors containing $^{18}F$-FDG injection volume, tumor size and the density of blood glucose. Therefore, we investigated how radioactivity uptake in target influences 2D or 3D image analysis and elucidate radioactivity concentration that mediate this effect. This study will show the relationship between the radioactivity uptake and 2D,3D image acquisition on radioactivity concentration. Materials and Methods: We got image with 2D and 3D using 1994 NEMA PET phantom and GE Discovery(GE, U.S.A) STe 16 PET/CT setting the ratio of background and hot sphere's radioactivity concentration as being a standard of 1:2, 1:4, 1:8, 1:10, 1:20, and 1:30 respectively. And we set 10 minutes for CT attenuation correction and acquisition time. For the reconstruction method, we applied iteration method with twice of the iterative and twenty times subset to both 2D and 3D respectively. For analyzing the images, We set the same ROI at the center of hot sphere and the background radioactivity. We measured the radioactivity count of each part of hot sphere and background, and it was comparative analyzed. Results: The ratio of hot sphere's radioactivity density and the background radioactivity with setting ROI was 1:1.93, 1:3.86, 1:7.79, 1:8.04, 1:18.72, and 1:26.90 in 2D, and 1:1.95, 1:3.71, 1:7.10, 1:7.49, 1:15.10, and 1:23.24 in 3D. The differences of percentage were 3.50%, 3.47%, 8.12%, 8.02%, 10.58%, and 11.06% in 2D, the minimum differentiation was 3.47%, and the maximum one was 11.06%. In 3D, the difference of percentage was 3.66%, 4.80%, 8.38%, 23.92%, 23.86%, and 22.69%. Conclusion: The difference of accumulated concentrations is significantly increased following enhancement of radioactivity concentration. The change of radioactivity density in 2D image is affected by less than 3D. For those reasons, when patient is examined as follow up scan with changing the acquisition mode, scan should be conducted considering those things may affect to the quantitative analysis result and take into account these differences at reading.